Skip to main content

Zimmer Biomet Spine Inc.

FDA adverse event reports naming Zimmer Biomet Spine Inc. as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: ZIMMER BIOMET SPINE INC..

Total reports
3,941
Deaths
1
Injuries
1,427
Malfunctions
2,513

Reports per year

15
17
18
19
20
21
22
23

Source: FDA adverse event reports by year received, 20152023. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.

Top device categories

Recent recalls

  • 2024-01-04Open, ClassifiedNonconforming Material/Component

    Vitality Polyaxial Screw, 7.5mmx 50mm, REF: 07.02000.118, used with Vitality and Vital Spinal Fixation System

    Spinal fixation system 7.5mmx 50mm screws are packaged with the correct labeling, but the screws themselves were laser marked with an incorrect size description of 7.5 x 40; use of incorrect screws could lead to screw perforation and indention of vascular or other structures.

  • 2019-02-22TerminatedNonconforming Material/Component

    Vitality Spinal Fixation System. Thoracolumbosacral pedicle screw system (non-sterile). Pedicle screw fixation is indicated for skeletally mature patients and for adolescent patients. For use in Orthopedic medical specialty area. Product Usage: The Vitality Shear-off Set Screw is not threading properly with the mating tulip head of the Vitality screws, hooks and connectors due to a manufacturing issue. These devices are indicated as an adjunct to fusion for various indications

    The Vitality Shear-off Set Screw is not threading properly with the mating tulip head of the Vitality screws, hooks and connectors due to a manufacturing issue.These devices are indicated as an adjunct to fusion for various indications

  • 2019-02-22TerminatedNonconforming Material/Component

    Vitality Spinal Fixation System. Spinal Interlaminal, Appliance, Fixation, (non-sterile). Pedicle screw fixation is indicated for skeletally mature patients and for adolescent patients. For use in Orthopedic medical specialty area. Product Usage: The Vitality Shear-off Set Screw is not threading properly with the mating tulip head of the Vitality screws, hooks and connectors due to a manufacturing issue. These devices are indicated as an adjunct to fusion for various indications.

    The Vitality Shear-off Set Screw is not threading properly with the mating tulip head of the Vitality screws, hooks and connectors due to a manufacturing issue.These devices are indicated as an adjunct to fusion for various indications

  • 2019-02-22TerminatedNonconforming Material/Component

    Vitality Spinal Fixation System.Spinal Intervertebral Body, Appliance, Fixation (non-sterile). Pedicle screw fixation is indicated for skeletally mature patients and for adolescent patients. For use in Orthopedic medical specialty area. Product Usage: The Vitality Shear-off Set Screw is not threading properly with the mating tulip head of the Vitality screws, hooks and connectors due to a manufacturing issue. These devices are indicated as an adjunct to fusion for various indications

    The Vitality Shear-off Set Screw is not threading properly with the mating tulip head of the Vitality screws, hooks and connectors due to a manufacturing issue.These devices are indicated as an adjunct to fusion for various indications

  • 2017-05-15TerminatedDevice Design

    Vitality Spinal Fixation System, Vitality T27 Final Drivers and Vitality Torque Limiting Handle 90 in-lbs Model Number(s): 07.02066.001, 07.02053.001, 730M0017

    Zimmer Biomet is conducting a medical device recall for Vitality T27 Final Drivers and Vitality Torque Limiting Handle 90 in-lbs because when applying torque to tighten the closure top the driver has the potential to create a torsional spring which can cause damage and the potential for the values of torque to go out of specification.

  • TerminatedDevice Design

    Altius M-INI OCT Posterior Spinal Fixation System, Model Number 1200-9216 The AltiusTM OCT and AltiusTM M-INITM OCT Systems when intended to promote fusion of the cervical spine and occipitocervico-thoracic junction (Occiput-T3), these systems are indicated for: DDD (neck pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies); spondylolisthesis; spinal stenosis; fracture/dislocation; atlantoaxial fracture with instability; occipitocervical dislocation; revision of previous cervical spine surgery; and tumors. The occipital bone screws are limited to occipital fixation only

    Zimmer Biomet is recalling the Altius M-INI OCT Posterior Spinal Fixation System because the driver is unable to mate with the screw head tulip.

  • TerminatedUnder Investigation by firm

    Zyston Transform Instrument Case. Model Number PCR8200-2101. Device labeled sterile.

    Zimmer Bioment Spine, Inc. announces a voluntary field action for the Zyston Transform Instrument Case and Zyston Transform Implant Kit because the locking mechanism on the implant is not functioning as intended.

  • TerminatedUnder Investigation by firm

    Zyston Transform Implant Kit. Model Number PCR8200-3101. Device labeled sterile.

    Zimmer Bioment Spine, Inc. announces a voluntary field action for the Zyston Transform Instrument Case and Zyston Transform Implant Kit because the locking mechanism on the implant is not functioning as intended.

Monitor Zimmer Biomet Spine Inc. automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.