Skip to main content
KWPClass 2

Appliance, Fixation, Spinal Interlaminal

FDA adverse event reports and recall history for appliance, fixation, spinal interlaminal (product code KWP), from the public MAUDE corpus and the live FDA recall database.

Total reports
14,672
Deaths
40
Injuries
7,707
Malfunctions
6,786

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2023-04-19Open, ClassifiedUnder Investigation by firm

    INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 24mm, REF 3604024; for immobilization and stabilization of spinal segment as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical unction, the cervical spine (C1 to C7), and the thoracic spine from T1-T3.

    Incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws

  • 2023-04-19Open, ClassifiedUnder Investigation by firm

    INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 32mm, REF 3604532; for immobilization and stabilization of spinal segment as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical unction, the cervical spine (C1 to C7), and the thoracic spine from T1-T3.

    Incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws

  • 2023-04-19Open, ClassifiedUnder Investigation by firm

    INFINITY Occipitocervical Upper Thoracic System, PARTIALLY THREADED MULTI AXIAL SCREW, Size: 4.0mm x 22mm, REF 360PT4022; for immobilization and stabilization of spinal segment as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical unction, the cervical spine (C1 to C7), and the thoracic spine from T1-T3.

    Incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws

  • 2023-04-19Open, ClassifiedUnder Investigation by firm

    INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 26mm, REF 3604026; for immobilization and stabilization of spinal segment as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical unction, the cervical spine (C1 to C7), and the thoracic spine from T1-T3.

    Incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws

  • 2023-04-19Open, ClassifiedUnder Investigation by firm

    INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 34mm, REF 3604034; for immobilization and stabilization of spinal segment as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical unction, the cervical spine (C1 to C7), and the thoracic spine from T1-T3.

    Incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws

  • 2019-02-22TerminatedNonconforming Material/Component

    Vitality Spinal Fixation System. Spinal Interlaminal, Appliance, Fixation, (non-sterile). Pedicle screw fixation is indicated for skeletally mature patients and for adolescent patients. For use in Orthopedic medical specialty area. Product Usage: The Vitality Shear-off Set Screw is not threading properly with the mating tulip head of the Vitality screws, hooks and connectors due to a manufacturing issue. These devices are indicated as an adjunct to fusion for various indications.

    The Vitality Shear-off Set Screw is not threading properly with the mating tulip head of the Vitality screws, hooks and connectors due to a manufacturing issue.These devices are indicated as an adjunct to fusion for various indications

  • 2017-01-13TerminatedPackaging

    Medacta International SA, M.U.S.T. Rod, REF 03.50.433, Size 5.5mm, Length 480mm, Type Straight, Mat Co-Cr-Mo The rods act as a connector between the different pedicle screws to create a stable construct for the stabilization and fusion of the lumbar and thoracic spine. Indications for Use: The M.U.S.T. pedicle screw system is intended for posterior non-cervical pedicle fixation (T1-S2/ilium) or anterolateral fixation (T8-L5). The pedicle screws are connected by 5.5mm Titanium alloy or CoCrMo rods to create a stable construct. The construct is secured using a set screw made of CoCrMo.

    Damaged packaging, lack of sterility assurance

  • 2016-10-20TerminatedProcess control

    RTI Surgical Streamline OCT Occipito-Cervico-Thoracic Spinal Rod, CoCr, 3.2mm (D) x 240mm (L) Product Usage: When intended to promote fusion of the cervical spine and occipitocervico-thoracic junction (occiput-T3), the Streamline OCT Occipito-Cervico-Thoracic System is intended for: degenerative disc disease (as defined by neck or back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture/ dislocation, atlanto/axial fracture with instability, occipito-cervical dislocation, deformities or curvature, tumors, pseudoarthrosis, and revision of previous cervical and upper thoracic spine surgery. The occipital bone screws are limited to occipital fixation only. The use of the pedicle screws (standard and high angle) is limited to placement in the upper thoracic spine (T1-T3) in treating thoracic conditions only. The pedicle screws are not intended for use in the cervical spine. The hooks, connectors, and rods are also intended to provide stabilization to promote fusion following reduction of fracture/ dislocation or trauma in the cervical/ upper thoracic (C1-T3) spine.

    RTI Surgical Inc. (RTI) is conducting a voluntary recall on two specific lots of the Streamline OCT Straight Rod, CoCr. 3.2mm (D) x 240mm (L) due to the potential for small cracks to be present on the rods. The small cracks on the rods used in implants could cause reduced strength of these affected implants, and lead to patient injury.

  • 2016-05-26TerminatedEquipment maintenance

    Various trauma and sports medicine instruments and implants. Spinal interlaminal fixation orthosis.

    Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is used or implanted, it may lead to an infection.

  • 2016-05-26TerminatedProcess design

    Streamline OCT Occipito-Cervico-Thoracic System, Crosslink Counter Torque Sleeve. Spinal fixation device instrument.

    Potential for a product defect where the cannula may be obstructed. This obstruction could prevent a driver from going through, which increases the potential for the driver to slip, resulting in possible patient injury.

Monitor appliance, fixation, spinal interlaminal automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.