HXXClass 1

Screwdriver

FDA adverse event reports and recall history for screwdriver (product code HXX), from the public MAUDE corpus and the live FDA recall database.

Total reports
23,582
Deaths
4
Injuries
1,569
Malfunctions
21,868

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2024-05-28Open, ClassifiedProcess change control

    STARDRIVE SCRWDRVR SHFT/T4 50MM/SELF-RETAINING/HXC-Intended for tightening, loosening, or removing of screw implants during orthopaedic surgery. Part Number: 03.130.010

    Stardrive Screwdrivers manufactured without a 5-degree relief cut around the tip and will not engage with their respective 1.3mm and 1.5mm screws, may result in surgical delay

  • 2016-06-01TerminatedDevice Design

    Firebird NXG Spinal Fixation System Set Screw Driver used to insert a set screw into a spinal fixation construct. Rx Only Provided non Sterile.

    There is a possibility that the subject set screw driver could disengage from the torque limiting handle when the counter torque wrench is allowed to contact the counter torque handle.

  • 2014-06-17TerminatedNonconforming Material/Component

    Biomet Spine, Cypher MIS Screw Inserter, Catalog number 14-501669, instrument for Spinal Screw Fixation System.

    The Cypher MIS Screw Inserter may exhibit an increased rate of instrument tip failure.

  • 2014-01-30TerminatedReprocessing Controls

    SET SCREW RETAINING DRIVER, REF 9010000820, QTY: 1 EA, Medtronic Sofamor Danek, USA, Inc. Screwdriver - orthopedic manual surgical instrument.

    The retaining tabs component of the Set Screw Retaining Drive may be oversized, which will not allow proper mating between the Set Screw Retaining Drive and the set screw.

  • 2013-11-27TerminatedNonconforming Material/Component

    Shortened Plug Driver; Low Volume Specialty Device; Qty:1; Rx Only; Non Sterile; Biomet Spine, Parsippany, NJ 07054 The Polaris Spinal Systems is a non-cervical spinal fixation device intended for immobilization and stabilization as an adjunct to fusion as a pedical screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterior or anterolateral fixation system. Pedicle screw fixation is limited to skeletal mature patients for use with autograft and/or allograft. The device is indicated for all the following indications:degenerative disc disease (defined as discogeneic back pain with degeneration of the disk confirmed by history and radiographic studies), spondylolisthesis, and or lordosis) tumor, stenosis, pseudoarthrosis, or failed previous fusion

    Biomet received a complaint on 5/20/2013 reporting that the tip of the instrument sheared off during a removal case.

  • 2013-09-14TerminatedNonconforming Material/Component

    T8 REMOVAL HEAD, REF 8910641, QTY: 1 EA, Product Usage: The non-sterile T8 Removal Diriver is used with the Spire and Spire Z Spinal System in conjunction with the set screwdriver to remove the implant during removal surgery. The set screw driver is attached to the end of the removal head and used to disengage the remaining portion of the set screw within the device.

    One lot of the T8 Removal Head had incorrect hardness specification, which could result in the T8 Removal head failing during removal of the CD Horizon Spire and Spire Z Spinal System implant. There is a potential for tissue damage to occur and/or minor surgical delay to remove the remaining fragments of the tip.

  • 2013-08-13TerminatedDevice Design

    Invue Solid Driver, Tipped, AO. Part Number: 11-70083. Allows the user to retain the screw when changing axis of insertion during fixation of spinal intervertebral.

    The nipple of the tipped driver broke off when surgeons were attempting to insert the screw.

  • 2013-08-08TerminatedDevice Design

    Rev E PedFuse Return Screw Inserter, Part Number: 11-80030 Screw inserters are instruments which capture, hold and drive pedicle screw assemblies. There is a capturing feature that engages and holds the screw in place. There is also a feature used to lock the screw to the inserter before insertion of the screw. They can also be used to remove screws or back screws out. Screw inserters are used during the implantation of spinal pedicle screws intended to provide immobilization and stabilization of spinal segments in skeletally mature patients. Used in conjunction with Pedicle Screw Spinal system marketed as KRD1 PedFuse Pedicle Screw System .

    Inserters have a set screw that fixes a collar to the inserter. The screw set can loosen and fall out of the inserter which allows the collar to slide off the proximal end of the inserter. When the collar is not in place the locking sleeve can also slide off the inserter.

  • 2013-08-08TerminatedDevice Design

    Invue Split Driver, Assembly with Handle, Model SI70006 Rev G. Used in conjunction with the Invue Anterior Cervical Plate System under 501(k) clearance number K121060. The Invue Driver is an instrument consisting of a Radel handle, a shaft and hexalobe-like tip with a 4.0mm split, which allows the driver to collapse to fit into a screw and expand to retain the screw.

    Device design.

  • 2013-07-05TerminatedDevice Design

    InSpan Inserter, Right Assembly; InSpan Inserter, Left Assembly Product Usage: Inserters are stainless steel instruments that have an inner shaft. The left and right inserters are used in combination with each other to place the implants into the body. Inserters are used to install the InSpan Spinous Plate System for the purpose of achieving supplemental fixation.

    Potential for set screw to be come deformed.