FDA adverse event reports naming THE Anspach Effort, Inc. as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: THE ANSPACH EFFORT, INC..

Total reports
3,941
Deaths
0
Injuries
17
Malfunctions
3,909

Reports per year

06
07
08
09
10
11
12
13
14
15
16
18
19
20
24

Source: FDA adverse event reports by year received, 2006–2024. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Top device categories

Recent recalls

  • 2023-01-07Open, ClassifiedUnder Investigation by firm

    6.5 cm Adult Rotating Craniotome Attachment. Use with XMax, microMax, and eMax Systems. Ref: CRANI-A Intended for cutting and shaping bone including the spine and cranium

    Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury

  • 2023-01-07Open, ClassifiedUnder Investigation by firm

    6.5 cm Adult Craniotome, Thin Foot Plate Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-A-01 intended for cutting and shaping bone including the spine and cranium by trained medical/surgical personnel

    Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury

  • 2023-01-07Open, ClassifiedUnder Investigation by firm

    Adult Craniotome Ref: CRANI-A-G1 Intended for cutting and shaping bone including the spine and cranium

    Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury

  • 2023-01-07Open, ClassifiedUnder Investigation by firm

    6.5 cm Adult Rotating Craniotome Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-A-R intended for cutting and shaping bone including the spine and cranium

    Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury

  • 2023-01-07Open, ClassifiedUnder Investigation by firm

    7.5 cm Large Craniotome Attachment Use with XMax, microMax, and eMax Systems. Ref: CRANI-L Intended for cutting and shaping bone including the spine and cranium

    Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury

  • 2023-01-07Open, ClassifiedUnder Investigation by firm

    Adult Craniotome, Large Ref: CRANI-L-G1 Intended for cutting and shaping bone including the spine and cranium

    Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury

  • 2023-01-07Open, ClassifiedUnder Investigation by firm

    7.5 cm Large Rotating Craniotome Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-L-R Intended for cutting and shaping bone including the spine and cranium

    Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury

  • 2023-01-07Open, ClassifiedUnder Investigation by firm

    6.5 cm Pediatric Craniotome Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-P Intended for cutting and shaping bone including the spine and cranium

    Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury

  • 2023-01-07Open, ClassifiedUnder Investigation by firm

    Pediatric Craniotome Ref:CRANI-P-G1

    Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury

  • 2018-01-11TerminatedUnder Investigation by firm

    OSCILLATING SAW ATTACHMENT Surgical instrument motors and accessories/attachments are intended for use during surgical procedures to cut hard tissue or bone and soft tissue.

    This voluntary recall was initiated because a veterinary unit was physically shipped in error to a non-veterinary customer account.

Monitor THE Anspach Effort, Inc. automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.