Tray, Surgical
FDA adverse event reports and recall history for tray, surgical (product code LRP), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 497
- Deaths
- 1
- Injuries
- 53
- Malfunctions
- 375
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2026-10-02Open, ClassifiedNonconforming Material/Component
Presource Packs containing Medtronic Microstream Components: Intubation Kit Catalog Numbers AN23366084 AN23366085 AN2336608H AN2336608J Intubating Bag Kit Catalog Number OMSVM4166A Intubation UH Kit Catalog Number AN2336608G
Kits contain Microstream components. Medtronic issued an update to the Instructions for Use of these components due reports of difficulty in disconnecting the adapter from the patient's endotracheal tube to perform a procedure. The issue can result in a delay to treatment and unintended extubation. The potential exists for respiratory failure, hypoxia, aspiration, or low oxygen saturation.
- 2026-09-22Open, ClassifiedNonconforming Material/Component
Presource Convenience Kits: 1) KIT, PRE OP , Catalog Number: AMN548616A; 2) KIT, NICU ADMIT NEW HAN , Catalog Number: BA1DH0116B; 3) KIT, PRE PROCEDURE , Catalog Number: IVSJV8302E; 4) KIT, IV SUPPLIES , Catalog Number: IVS3223531; 5) KIT, IV SUPPLIES , Catalog Number: IVS3223532; 6) KIT, IV SUPPLIES , Catalog Number: IVS3223533; 7) KIT, IV SUPPLIES , Catalog Number: IVS3223534; 8) KIT, IV SUPPLIES , Catalog Number: IVS3223535; 9) KIT, IV SUPPLIES , Catalog Number: IVS322358M; 10) KIT, IV SUPPLIES , Catalog Number: IVS322358N; 11) KIT, IV SUPPLIES , Catalog Number: IVS322358O; 12) KIT, PATIENT ADMISSIONS , Catalog Number: AMN333032A; 13) KIT, C SECTION , Catalog Number: MAT42267AL; 14) KIT, C SECTION , Catalog Number: MAT4226725; 15) KIT, VAGINAL DELIVERY , Catalog Number: MAT422677G; 16) KIT, VAGINAL DELIVERY , Catalog Number: MAT4226792; 17) KIT, VAGINAL DELIVERY , Catalog Number: MAT4226793; 18) KIT, VAGINAL DELIVERY , Catalog Number: MAT4226794; 19) KIT, VAGINAL DELIVERY , Catalog Number: MAT4226795
The kits contain applicators that may exhibit an open or incomplete seal on the packaging of the applicator. Use of a potentially contaminated applicator could result in no patient harm or may lead to localized skin or soft tissue infection. Less commonly, systemic infection or other infectious complications may occur.
- 2026-09-03Open, ClassifiedProcess change control
Presource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays: 1) KIT, LUNG TX CIRCULATOR , Catalog Number: OVC3507231
The Urinary Catheter Trays contain CliniClean Pericare Wipes which may potentially be contaminated with a lubricant used during the manufacturing process. If a contaminated wipe is used on a patient, localized skin effects may occur. These may include irritation, erythema, or mild inflammation.
- 2026-08-31Open, ClassifiedNonconforming Material/Component
Medline Convenience Kits: 1) ARTERIAL LINE INSERTION KIT, Model Number: ART1295
The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach of sterility in the packaging, due to wrinkles in the paper lidding which may extend to the seal area. If the sterile barrier is compromised, patients and users may be exposed to microbial contamination. Potential health consequences may include no injury, localized infection, or, less commonly, more serious infection requiring medical treatment, antimicrobial therapy, procedural intervention, or hospitalization. In rare cases, particularly in critically ill or immunocompromised patients, severe infection could become life-threatening.
- 2026-06-29Open, ClassifiedNonconforming Material/Component
Medline Convenience Kits: 1) SUTURE TRAY, Model Number: SUT12260; 2) CHEST TUBE INSERTION TRAY, Model Number: MNS3390; 3) CONTRACEPTIVE IMPLANT INSERTION KIT, Model Number: MNS9715; 4) 20GX3" RADIAL ART LINE KIT, Model Number: ART351; 5) MYELOGRAM PACK, Model Number: P155615A; 6) CHEST TUBE INSERTION TRAY, Model Number: MNS3390; 7) LACERATION TRAY, Model Number: SUT18660A; 8) GENERAL BIOPSY TRAY, Model Number: MNS5985; 9) LACERATION TRAY, Model Number: DYNDL1351; 10) 20GX3" RADIAL ART LINE KIT, Model Number: ART351; 11) ARTERIAL LINE TRAY, Model Number: ART600; 12) SUTURE TRAY, Model Number: SUT12260; 13) ER SUTURING TRAY, Model Number: SUT9710C; 14) INCISION & DRAINAGE TRAY, Model Number: ID2060; 15) LACERATION TRAY - SNAG FREE, Model Number: SUT12265; 16) BIOPSY TRAY, Model Number: DYNDH1466; 17) GENERAL BIOPSY TRAY, Model Number: MNS5985; 18) SUTURE SET, Model Number: SUT19665; 19) MYELOGRAM-LF SAFE PACK, Model Number: P124863; 20) 20GX3" RADIAL ART LINE KIT, Model Number: ART351; 21) LACERATION TRAY, Model Number: SUT18660A; 22) 20GX6" FEMORAL ART LINE KIT, Model Number: ART352; 23) BIOPSY PACK, Model Number: DYNDH1699; 24) SUTURE TRAY, Model Number: SUT12260; 25) CHEST TUBE INSERTION TRAY, Model Number: MNS3390; 26) ARTHROGRAM TRAY, Model Number: SPEC0070B; 27) SUTURING SET, Model Number: SUT4155; 28) SUTURE TRAY, Model Number: SUT16080; 30) MYELOGRAM TRAY CHOICE, Model Number: DYNDH1213; 31) MYELOGRAM PACK, Model Number: P155615A; 32) LACERATION TRAY - SNAG FREE, Model Number: SUT12265
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
- 2026-06-12Open, ClassifiedProcess control
Medline EBSI Kit EBSI1534
Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
- 2026-05-07Open, ClassifiedProcess control
Medline Convenience Kits: 1) CVC PACK, Model Number: CVI4500A; 2) TOTE KAMAL OSC HAND REPAIR, Model Number: DYKM1730A; 3) TOTE CHANGE OSC HAND REPAIR, Model Number: DYKM1751A; 4) TOTE SB NGUYEN CAPSULOTOMY, Model Number: DYKM1756A; 5) TOTE MOMENI SB CAPSULOTOMY REC, Model Number: DYKM1765A; 6) TOTE SAFRAN OSC HIP ARTHRSCPY, Model Number: DYKM1767B; 7) TOTE SEN OSC HAND REPAIR, Model Number: DYKM1776; 8) TOTE CURTIN OSC HAND REPAIR, Model Number: DYKM1781A; 9) TOTE FOX OSC HAND REPAIR, Model Number: DYKM1783A; 10) TOTE YAO OSC HAND REPAIR, Model Number: DYKM1796A; 11) TOTE ABRAMS KNEE ARTHROSCOPY, Model Number: DYKM1801B; 12) TOTE YAO OSC WRIST ARTHROSCOPY, Model Number: DYKM1807; 13) TOTE CARROLL BLOCK BLOOD PATCH, Model Number: DYKM1864; 14) TOTE SB STONE JWIRE LUMPECTOMY/SEN NODE, Model Number: DYKM1925; 15) TOTE SB NGUYEN LIPOSUCTION, Model Number: DYKM1927A; 16) OSC SHERMAN KNEE ARTHRO BASIC, Model Number: DYKM1977B; 17) CHEST TUBE SUPPLIES, Model Number: DYKM2131
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
- 2026-04-30Open, ClassifiedUnder Investigation by firm
Centurion medical procedure kits, containing Medline Neuro Sponges, labeled as follows: SURGICAL ENT PREP KIT, Medline Kit Number/SKU MNS11715
During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
- 2026-04-30Open, ClassifiedUnder Investigation by firm
Centurion medical convenience kits, containing Webcol Large Alcohol Prep Pads, labeled as: LATEX FREE GENERAL PURPOSE TRAY, Medline Kit SKU 75670
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Cardinal Health Webcol Large Alcohol Prep Pads. Cardinal Health has deemed Webcol Large Alcohol Prep Pads to be non-sterile following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.
- 2026-02-17Open, ClassifiedUnder Investigation by firm
Medline Convenience Kits: 1) TOTE ABRAMS KNEE ARTHROSCOPY, Model Number: DYKM1801B; 2) TOTE SEN OSC HAND REPAIR, Model Number: DYKM1776; 3) TOTE MOMENI SB CAPSULOTOMY REC, Model Number: DYKM1765A; 4) TOTE YAO OSC HAND REPAIR, Model Number: DYKM1796A; 5) TOTE SB NGUYEN CAPSULOTOMY, Model Number: DYKM1756A; 6) TOTE FOX OSC HAND REPAIR, Model Number: DYKM1783A; 7) TOTE KAMAL OSC HAND REPAIR, Model Number: DYKM1730A; 8) TOTE SB NGUYEN LIPOSUCTION, Model Number: DYKM1927A; 9) TOTE YAO OSC WRIST ARTHROSCOPY, Model Number: DYKM1807; 10) TOTE CURTIN OSC HAND REPAIR, Model Number: DYKM1781A; 11) OSC SHERMAN KNEE ARTHRO BASIC, Model Number: DYKM1977B; 12) TOTE CHANGE OSC HAND REPAIR, Model Number: DYKM1751A; 13) TOTE CARROLL BLOCK BLOOD PATCH, Model Number: DYKM1864; 14) TOTE SB STONE JWIRE LUMPECTOMY/SEN NODE, Model Number: DYKM1925; 15) TOTE SAFRAN OSC HIP ARTHRSCPY, Model Number: DYKM1767B; 16) CVC PACK, Model Number: CVI4500A
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Monitor tray, surgical automatically
DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.