HBBClass 2

Motor, Drill, Pneumatic

FDA adverse event reports and recall history for motor, drill, pneumatic (product code HBB), from the public MAUDE/AEMS corpus and the live FDA recall database.

Total reports
9,322
Deaths
4
Injuries
265
Malfunctions
9,004

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 20122026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.

Recent recalls

  • 2025-10-02Open, ClassifiedProcess control

    AESCULAP ELAN 4 FIXED DURAGUARD STANDARD. Model Number: GB942R.

    Mislabeling. The ELAN 4 FIXED DURAGUARD STANDARD was labeled as "Long" and the ELAN 4 FIXED DURAGUARD LONG was labeled as "Standard."

  • 2025-10-02Open, ClassifiedProcess control

    AESCULAP ELAN 4 FIXED DURAGUARD LONG. Model Number: GB943R.

    Mislabeling. The ELAN 4 FIXED DURAGUARD STANDARD was labeled as "Long" and the ELAN 4 FIXED DURAGUARD LONG was labeled as "Standard."

  • 2018-02-15TerminatedNo Marketing Application

    The ELAN 4 Air System is intended for high speed cutting, sawing and drilling of bone in fields of Spine, ENT, Neuro, and Maxillofacial Surgery.

    The ELAN 4 Air System was released to the market place prior to an authorized FDA 510K clearance. This issue does not affected sales outside the United States (OUS).

  • 2016-06-09TerminatedComponent design/selection

    Maestro Air Motor Pneumatic cranial drill motor. The Maestro Air Pneumatic System is a pneumatically operated surgical instrument system. The pneumatic motor provides power to operate removable rotating surgical cutting tools and their accessories.

    Due to the buildup of grease within the hose and motor, the product may require multiple attempts to start. The grease may also migrate into the attachment end of the motor. There is a negligible possibility that the orange grease may leak from the attachment end of the motor and contaminate the surgical wound resulting in soft tissue inflammation.

  • 2016-05-04TerminatedLabeling mix-ups

    Medtronic Midas Rex Dissecting Tool Legend 10 cm 4mm BA DIAM. Intended for use in neurosurgery.

    Incorrect color coding and attachment reference were included on the labeling of 10BA40D dissecting tools with lot number H5237248.

  • 2014-06-25TerminatedProcess design

    ANSPACH***REF 98-0044***Custom Device Speed Reducer with Bumper***Rx Only*** This device is intended to be used with Anspach Systems

    The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling their Anspach Custom Devices. The units were not manufactured consistently with Quality System requirements.

  • 2014-06-25TerminatedProcess design

    ANSPACH***REF 98-0048***Custom Circular ePLUS-FP***Rx Only*** This device is intended to be used with Anspach Systems

    The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling their Anspach Custom Devices. The units were not manufactured consistently with Quality System requirements.

  • 2014-06-25TerminatedProcess design

    ANSPACH***1.85mm x 16 mm Fluted Router for MRI***Rx Only*** This device is intended to be used with Anspach Systems

    The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling their Anspach Custom Devices. The units were not manufactured consistently with Quality System requirements.

  • 2014-06-25TerminatedProcess design

    ANSPACH***REF SP-3014-00***4mm Fluted Ball for MRI***Rx Only*** This device is intended to be used with Anspach Systems

    The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling their Anspach Custom Devices. The units were not manufactured consistently with Quality System requirements.

  • 2014-01-09TerminatedComponent design/selection

    ***REF IMRI-MOTOR***iMRI Handpiece. Use with iMRI System Only***Rx Only***Manufacturer: The Anspach Effort, Inc. Intended for cutting and shaping bone.

    The firm Anspach Effort of Palm Beach Gardens, FL initiated a voluntary removal of the Pneumatic Drill System, including the motor hand-piece, Auto Lube foot control, without a pressure relief valve (PRV) due to the hand piece's outer hose can rupture without warning if the air flow is blocked during use..

Monitor motor, drill, pneumatic automatically