Motor, Surgical Instrument, Ac-Powered
FDA adverse event reports and recall history for motor, surgical instrument, ac-powered (product code GEY), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 18,328
- Deaths
- 0
- Injuries
- 124
- Malfunctions
- 18,194
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2023-07-05Open, ClassifiedError in labeling
KLS Martin Battery Pack, REF: KLS BP 001, and KLS BP 040 (bulk case of 40), used in conjunction with the KLS Martin MaxDriver
Battery packs used in surgical procedures where a surgeon uses a powered screwdriver and a right-angle attachment have switched manufacturing and expiration dates on sterile pouch labels, so batteries will appear to be prematurely expired.
- 2017-04-24TerminatedProcess control
GP161R; ELAN 4 1-RING DIAMOND BURR COARSE D2.0; GP162R; ELAN 4 1-RING DIAMOND BURR+ COARSE D2.0; GP163R; ELAN 4 1-RING DIAMOND BURR COARSE D2,3; GP164R; ELAN 4 1-RING DIAMOND BURR+ COARSE D2,3; GP165R; ELAN 4 1-RING DIAMOND BURR COARSE D3,0; GP166R; ELAN 4 1-RING DIAMOND BURR+ COARSE D3,0; GP167R; ELAN 4 1-RING DIAMOND BURR++ COARSE D3,0; GP168R; ELAN 4 1-RING DIAMOND BURR COARSE D4,0; GP169R; ELAN 4 1-RING DIAMOND BURR COARSE D5,0; GP170R; ELAN 4 1-RING DIAMOND BURR COARSE D6,0; GP173R; ELAN 4 1-RING DIAMOND BURR X-COARSE D3,0; GP174R; ELAN 4 1-RING DIAMOND BURR X-COARSE D4,0; GP175R; ELAN 4 1-RING DIAMOND BURR X-COARSE D5,0; GP176R; ELAN 4 1-RING DIAMOND BURR X-COARSE D6,0; GP177R; ELAN 4 1-RING DIAMOND BURR X-COARSE D7,0; GP328R; ELAN 4 2-RING DIAMOND BURR X-COARSE D4.0; GP329R; ELAN 4 2-RING DIAMOND BURR X-COARSE D5.0; GP330R; ELAN 4 2-RING DIAMOND BURR X-COARSE D6.0; Product Usage: The ELAN 4 motor system and the Hi-Line XS handpiece are intended for high speed cutting, sawing, drilling and manipulation of soft tissue and bone in the fields of Spine, ENT, Neuro, and Maxillofacial Surgery.
Diamond particulates may break off the burr and fall into the surgical site. If the particles are not detected and are not removed from the surgical site a foreign body reaction may occur. These particulates could also lead to mechanically induced soft tissue damage.
- 2017-04-24TerminatedProcess control
GP321R; ELAN 4 2-RING DIAMOND BURR COARSE D2.3; GP322R; ELAN 4 2-RING DIAMOND BURR COARSE D3.0; GP323R; ELAN 4 2-RING DIAMOND BURR COARSE D4.0; GP324R; ELAN 4 2-RING DIAMOND BURR COARSE D5.0; GP325R; ELAN 4 2-RING DIAMOND BURR COARSE D6.0; Product Usage: The ELAN 4 motor system and the Hi-Line XS handpiece are intended for high speed cutting, sawing, drilling and manipulation of soft tissue and bone in the fields of Spine, ENT, Neuro, and Maxillofacial Surgery.
Diamond particulates may break off the burr and fall into the surgical site. If the particles are not detected and are not removed from the surgical site a foreign body reaction may occur. These particulates could also lead to mechanically induced soft tissue damage.
- 2017-04-24TerminatedProcess control
GE426R; HI-LINE XS DIAMOND BURR X-CRS.I D4.5MM; GE426SU; HI-LINE XS DISP.DIAM.BURR X-CRS.I D4.5MM; GE526R; HI-LINE XS DIAMOND BURR X-CRS.II D4.5MM; GE526SU; HI-LINE XS DISP.DIAM.BURR X-CRS.II D4.5; GE626R; HI-LINE XS DIAMOND BURR X-CRS.III D4.5MM; GE626SU; HI-LINE XS DISP.DIAM.BURR X-CRS.III D4.5; Product Usage: The ELAN 4 motor system and the Hi-Line XS handpiece are intended for high speed cutting, sawing, drilling and manipulation of soft tissue and bone in the fields of Spine, ENT, Neuro, and Maxillofacial Surgery.
Diamond particulates may break off the burr and fall into the surgical site. If the particles are not detected and are not removed from the surgical site a foreign body reaction may occur. These particulates could also lead to mechanically induced soft tissue damage.
- 2015-09-15TerminatedProcess control
NICO Myriad Console Model number NN-7000 The Myriad System is a surgical system specifically designed for morcellation of soft tissue. It primarily consists of a reusable console with foot pedal and a disposable handpiece. For the morcellation and removal of tissue during pelviscopic, laparoscopic, percutaneous and open surgical procedures whenever access to the surgical site is limited.
Two cables within the console were inadvertently twisted during manufacturing. The individual wires within the cables may break and/or contact one another where the cables terminate into two connectors. When these wires break or contact one another (i.e., short), various types of malfunction may occur with the console. One type of malfunction may result in unexpected aspiration increase.
- 2014-06-25TerminatedProcess design
ANSPACH***REF-98-0056***Custom Device Autolube III modified with a swivel***Rx Only*** This device is intended to be used with Anspach Systems
The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling their Anspach Custom Devices. The units were not manufactured consistently with Quality System requirements.
- 2014-03-04TerminatedLabeling Change Control
Punch Hair Matic A surgical hair transplant device designed for automation of follicular transplantation.
Medicamat is initiating a field correction due to misbranded marketing literature for the Punch Hair Matic SAFER.
- 2014-01-06TerminatedUnknown/Undetermined by firm
Anspach BLACKMAX-N (Pneumatic motor) and Autolube III (Foot Control). Cutting and shaping bone.
Anspach was contacted by a hospital to inform Anspach that a Blackmax motor had been used in a case where the patient was diagnosed with Creutzfeldt-Jacob Disease (CJD).
- 2013-03-06TerminatedError in labeling
D3000 Advantage Drive System, 115v Product Usage: The Advantage Drive Systems function as a powered instrument system consisting of handpieces and accessories to perform cutting of soft tissue and bone
The Membrane Switch Panel (referred to as "front panel" of the component part number A34-003-207, used on both the D3000 and D3000I Advantage Drive System, contained the incorrect symbology per third edition IEC60606-01 Medical Electrical Equipment: General requirements for basic safety and essential performance.
- 2013-03-06TerminatedError in labeling
D3000I Advantage Drive System, 230v, surgical instrument motor. Product Usage: The Advantage Drive Systems function as a powered instrument system consisting of handpieces and accessories to perform cutting of soft tissue and bone
The Membrane Switch Panel (referred to as "front panel" of the component part number A34-003-207, used on both the D3000 and D3000I Advantage Drive System, contained the incorrect symbology per third edition IEC60606-01 Medical Electrical Equipment: General requirements for basic safety and essential performance.
Monitor motor, surgical instrument, ac-powered automatically