Siemens Healthcare Diagnostics Inc
FDA adverse event reports naming Siemens Healthcare Diagnostics Inc as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: SIEMENS HEALTHCARE DIAGNOSTICS INC.
- Total reports
- 2,857
- Deaths
- 1
- Injuries
- 10
- Malfunctions
- 2,706
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Top device categories
- Analyzer, Chemistry (Photometric, Discrete), For Clinical UseJJE1,724
- Immunoassay Method, Troponin SubunitMMI416
- Electrode, Ion Specific, SodiumJGS151
- Enzyme Immunoassay, TracrolimusMLM114
- System, Test, Human Chorionic GonadotropinDHA58
- Cresolphthalein Complexone, CalciumCIC42
- Automated Urinalysis SystemKQO20
- Enzyme Immunoassay, DigoxinKXT16
- Test, Hepatitis B (B Antigen, B Surface Antigen, Be Antigen)LOM14
- Equipment, Laboratory, General Purpose, Labeled Or Promoted For A Specific Medical UseLXG14
Recent recalls
- 2026-08-17Open, ClassifiedUnknown/Undetermined by firm
Syva EMIT 2000 Quinidine Assay REF: 4Q019UL Contents: 1x28 mL, 1x14mL
Due to unacceptable precision performance exhibited in certain lots of the Emit 2000 Quinidine Assay, which may lead to erroneously elevated quinidine measurements, creating risk of sub-optimal patient treatment, including potential life-threatening cardiac arrhythmias or inadequate antimalarial therapy.
- 2026-08-17Open, ClassifiedUnknown/Undetermined by firm
Beckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mL
Due to unacceptable precision performance exhibited in certain lots of the Emit 2000 Quinidine Assay, which may lead to erroneously elevated quinidine measurements, creating risk of sub-optimal patient treatment, including potential life-threatening cardiac arrhythmias or inadequate antimalarial therapy.
- 2026-08-05Open, ClassifiedUnder Investigation by firm
Atellica CH Urine Albumin REF: 11537225
processing Atellica CH Urine Albumin (UAlb) tests on Atellica CH and Atellica CI analyzers may lead to the following issues: Falsely depressed Atellica CH Urea Nitrogen (UN_c) results; Falsely depressed Atellica CH Enzymatic Hemoglobin A1c (A1c_E) results; and Increased rate of reaction cuvette QC failures and Delayed results due to QC failures.
- 2026-08-04Open, ClassifiedUnder Investigation by firm
Syva Emit Disopyramide Assay Enzyme Immunoassay, Lidocaine SMN 10445370 Catalog No 6N319UL
Out-of-specification sodium azide concentration in Syva Emit Disopyramide Enzyme Immunoassay Calibrator Lot U1. While the assay results were not impacted, the overall health risk to lab technicians due to this issue is low, as exposure is limited by small volumes and a 0.2% post-reconstitution sodium azide concentration, with the worst outcome being mild skin or eye irritation and no serious harm anticipated.
- 2026-08-04Open, ClassifiedUnder Investigation by firm
Syva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219UL
out-of-specification sodium azide concentration in Syva Emit Tox Serum Tricyclic Antidepressant Calibrator/Controls Lot U2. While the assay results were not impacted, the overall health risk to lab technicians due to this issue is low, as exposure is limited by small volumes and a 0.2% post-reconstitution sodium azide concentration, with the worst outcome being mild skin or eye irritation and no serious harm anticipated.
- 2026-07-08Open, ClassifiedUnder Investigation by firm
Atellica CH Enzymatic Hemoglobin A1c (A1c_E). Material Number: 11097536.
The potential for intermittent well-to-well bias on the A1c_E assay on the Atellica CH and Atellica CI analyzer which could cause falsely depressed or falsely elevated HbA1c patient results.
- 2026-06-10Open, ClassifiedUnder Investigation by firm
Siemens Dimension Tacrolimus Flex Reagent Cartridge (TAC), REF DF207, SMN 10700795; enzyme immunoassay
A potential positive bias in quality control (QC) and patient sample results in three lots of the Dimension Tacrolimus (TAC) Flex reagent cartridge.
- 2026-04-29Open, ClassifiedUnder Investigation by firm
epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;
Use of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary patient samples. An erroneously low pH or mTCO2 reading can lead to inappropriate treatment of nonexistent acidosis (respiratory of metabolic) or missed diagnosis of alkalosis (a true high pH: respiratory or metabolic alkalosis), potentially causing harm through unnecessary interventions with the former or failure to deliver appropriate treatment with the latter.
- 2026-04-28Open, ClassifiedUnder Investigation by firm
Dimension Creatinine Flex reagent cartridge. Material Number: 10872079.
Siemens Healthcare Diagnostics is recalling certain lots of Dimension Creatinine Flex reagent cartridge (Dimension CRE2) because a potential imprecision in the Dimension CRE2 quality control (QC) and patient sample results using lot numbers GA6307 and BA7005 on the Dimension System. This imprecision may lead to erroneously decreased or increased creatinine patient results
- 2026-03-27Open, ClassifiedProcess control
Atellica CH Enzymatic Hemoglobin A1c (A1c_E); Siemens Material Number (SMN): 11097536;
A potential for depressed results for Atellica CH A1c_E when processing Atellica CH RCRP on the same analyzer. Customers processing A1c_E and RCRP on separate analyzers are not impacted by the recall. The frequency and magnitude of the negative bias and imprecision increase over time and are not consistent with the assay Instructions For Use (IFU). When this issue is present, Quality Control (QC) results demonstrate a negative trend and increased imprecision. The clinical impact of a negative (A1c_E) bias may lead to a delay in diagnosis for diabetes or lead to less-than-optimal treatment for existing diabetes.
Monitor Siemens Healthcare Diagnostics Inc automatically
DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.