Enzyme Immunoassay, Tracrolimus
FDA adverse event reports and recall history for enzyme immunoassay, tracrolimus (product code MLM), from the public MAUDE/AEMS corpus and the live FDA recall database.
- Total reports
- 327
- Deaths
- 0
- Injuries
- 12
- Malfunctions
- 311
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.
Recent recalls
- 2026-06-10Open, ClassifiedUnder Investigation by firm
Siemens Dimension Tacrolimus Flex Reagent Cartridge (TAC), REF DF207, SMN 10700795; enzyme immunoassay
A potential positive bias in quality control (QC) and patient sample results in three lots of the Dimension Tacrolimus (TAC) Flex reagent cartridge.
- 2023-03-15Open, ClassifiedUnder Investigation by firm
Dimension Tacrolimus (TAC) Flex reagent cartridge-In vitro diagnostic test for the quantitative measurement of tacrolimus in human whole blood on the Dimension clinical chemistry system SMN: 10700795 (DF207)
Siemens Healthineers has received customer complaints and confirmed imprecision for Quality Control (QC) and patient samples with Dimension Tacrolimus (TAC) lots GA2286, GA3047 and GA3171. The health consequences of erroneous patient results are major, since in a worst case scenario, a patient s tacrolimus dosing could be altered, resulting in potential organ failure and rejection (due to too low or no tacrolimus being administered) or toxicity (due to too much tacrolimus being administered). Siemens internal investigation of the Dimension TAC assay showed the worst case imprecision with patient samples at the low end of the Analytical Measurement Range (AMR). A patient sample at 2.0 ng/mL (2.6 nmol/L) recovered as 0.0 ng/mL (0.0 nmol/L) (100% negative bias), another patient sample at 2.1 ng/mL (2.7 nmol/L) recovered at 4.3 ng/mL (5.6 nmol/L) (103% positive bias).
- 2015-04-20TerminatedUnder Investigation by firm
DiaSorin, PRO-Trac II Tacrolimus ELISA kit, REF 32400 and 32400-CN. Product Usage: The PRO-Trac II Tacrolimus ELISA kit is an in vitro reagent system intended for the quantitative determination of tacrolimus (Prografl, FK506) and some metabolites in human EDT A or heparinized whole blood as an aid in the management of liver transplant patients receiving tacrolimus therapy.
Instability in the PRO-Trac II Tacrolimus ELISA kit which may cause falsely elevated patient results.
- 2013-12-16TerminatedComponent design/selection
Dimension(R) TACR Flex(R) reagent cartridge (DF107) The TACR method is an in vitro diagnostic test intended to quantitatively measure Tacrolimus in human whole blood on the Dimension(R) clinical chemistry system. Measurements of Tacrolimus are used as an aid in the management of Tacrolimus therapy in kidney and liver transplant patients
Siemens has confirmed that the TACR method may demonstrate reduced on-board stability which may result in imprecise and inaccurate QC and patient results.
- 2013-08-07TerminatedProcess control
Dimension Tacrolimus Flex reagent cartridge (DF107) and Dimension TACR CAL (DC107) Product Usage: The TACR method is an in vitro diagnostic test intended to quantitatively measure Tacrolimus in human whole blood on the Dimension clinical chemistry system. Measurements of Tacrolimus are used as an aid in the management of tacrolimus therapy in kidney and liver transplant patients.
Siemens initiated a recall due to confirmed complaints of low patient sample recovery with Dimension TACR Flex Reagent cartridge (DF107) lot BB4087 and its linked calibrator (DC 107) lot 3BD029. Falsely low Tacrolimus results across the assay range may lead to alterations in dosing.
- 2013-03-15TerminatedProcess control
Siemens Dimension(R) Tacrolimus Flex(R) reagent cartridge (DF107). The TACR method is used as an in vitro diagnostic test.
Siemens has confirmed customer complaints of low patient sample recovery with Dimension(R) TACR Flex(R) lot FA3085. Quality Control (QC) materials have not exhibited low recovery of the same magnitude. Internal method regression testing versus LC/MS has shown an average 25% low bias compared to LC/MS across the assay range. Individual patients, particularly samples with lower TACR values
- 2010-10-15TerminatedOther
Architect Tacrolimus Reagent Kit; an in-vitro diagnostic chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of tacrolimus in human whole blood; Abbott Laboratories, Abbott Diagnostics Division, Abbott Park, IL 60064 USA; list 1L77-25 The Architect Tacrolimus assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of tacrolimus in human whole blood as an aid in the management of liver and kidney allograft patients receiving tacrolimus therapy.
There are reagent barcode read errors when using reagent lot 86599M500 due to the print quality of the barcode.
- 2010-09-13TerminatedOther
Dimension Tacrolimus Flex Reagent Cartridge, manufactured by Siemens Healthcare Diagnostics Inc., Newark, Delaware.
Low patient sample recovery results in all affected lots, with possibility of inappropriate increase in tacrolimus dosage based on false low result.
- 2005-04-13TerminatedOther
Version 12 software program written for the Bio-Tek ELx800 automated microtiter plate reader with use of PRO-Trac Tacrolimus ELISA kit
The software program written for use in the automated microtiter plate reader (Bio-Tek ELx800) for use with the Pro-Trac II Tacrolimus ELISA kit was found to be faulty. Assay software program did not include the Calibrator Zero OD specification (STDO>=1.500) as specified in the product insert.
- 2003-12-17TerminatedOther
IMx Tacrolimus II Assay, list 3C10-20; Abbott Diagnostics, a Division of Abbott Laboratories, Abbott Park, IL 60064
Inadequate instructions for use resulting in calibration errors and the protein interference data in the package insert is not representative of assay performances.
Monitor enzyme immunoassay, tracrolimus automatically
DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.