Immunoassay Method, Troponin Subunit
FDA adverse event reports and recall history for immunoassay method, troponin subunit (product code MMI), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 18,369
- Deaths
- 13
- Injuries
- 1,559
- Malfunctions
- 16,086
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2024-12-20Open, ClassifiedComponent design/selection
Access hsTnI Reagent, REF B52699, Immunoassay Method, Troponin Subunit. For the quantitative determination of cardiac troponin I (cTnI) levels in human serum and plasma.
Beckman Coulter, Inc. is recalling their Access hsTnI assay, a medical device by correction. The reason for the recall is: high troponin antigen (>55,000 pg/mL) can contaminate an analyzers probe and cause inter-assay carryover due to insufficient washing. The contamination can cause a falsely elevated Access hsTnI result in subsequent sample(s) which can impact patient care (potentially but not limited to unnecessary coronary imaging or diagnostic catheterization) if the result is near the medical decision points.
- 2023-08-08Open, ClassifiedProcess control
VITROS Immunodiagnostic Products Troponin I ES Reagent Pack
Reagent packs could potentially contain incorrect wells, which will generate incorrect, and potentially false negative, results.
- 2023-08-08Open, ClassifiedProcess control
VITROS Immunodiagnostic Products Troponin I ES Calibrators
Reagent packs could potentially contain incorrect wells, which will generate incorrect, and potentially false negative, results.
- 2023-07-07Open, ClassifiedNonconforming Material/Component
Quidel Triage Cardiac Panel: REF: 97000HS, 97021HS (Troponin I), 97022HS (CK-MB and Troponin I), 97000HSEU, 97000HSEUJP, 97000HZ01, 97000QIL. Lots T13664 - T14045.
Raw material issue may cause cardiac panel products to have an average of 30% troponin signal reduction across measurement range 0.05 ng/mL to 30 ng/mL, QC testing won't detect anomaly, so an alternate method is recommended, if unavailable, use recommendations to mitigate potential patient impact, which could be missed/delayed myocardial infarction diagnosis.
- 2022-07-06TerminatedUnder Investigation by firm
Dimension LTNI -In vitro diagnostic test intended to quantitively measure cardiac troponin-I levels in human serum and heparinized plasma to aid in the diagnosis of myocardial infarction. Siemens Material Number (SMN)/REF (Catalog Number): 10444896/RF521
Positive bias with Lithium Heparin plasma samples from individuals that are expected to be below the 99th percentile of 0.07 ng/mL [0.07 g/L], may lead to inappropriate intervention for myocardial infarction
- 2022-07-06TerminatedUnder Investigation by firm
Dimension CTNI-In vitro diagnostic test intended to quantitively measure cardiac troponin-I levels in human serum and heparinized plasma to aid in the diagnosis of myocardial infarction. Siemens Material Number (SMN)/REF (Catalog Number): 10444905/RF421C
Positive bias with Lithium Heparin plasma samples from individuals that are expected to be below the 99th percentile of 0.07 ng/mL [0.07 g/L], may lead to inappropriate intervention for myocardial infarction
- 2018-05-04TerminatedError in labeling
Dimension Vista Cardiac Troponin I (CTNI) Flex reagent cartridge
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.
- 2017-05-05TerminatedUnder Investigation by firm
Dimension Troponin 1 (TNI) Flex reagent cartridge
The concentrations for Biotin listed in the Non-Interfering Substances section of the current Dimension and Dimension Vistat Instructions for Use (IFU) for Dimension TNI, Dimension Vista DIGXN, E2, FERR, PRL, and TSH have incorrect units, and incorrectly state the level at which biotin does not interfere.
- 2016-01-07TerminatedLabeling design
Siemens Acute Care CTNI TestPak-in vitro diagnostic test for the measurement of cardiac troponin I in heparinized plasma Catalog Number: CCTNI SMN: 10445071
Stratus CS Cartridge Barcode labels may not scan on the Stratus CS systems
- 2015-07-23TerminatedUnder Investigation by firm
ADVIA Centaur Tni-Ultra Assay; SMN 10317708 - 100 tests and SMN 10317709 -500 tests Product Usage: For in vitro diagnostic use in the quantitative determination of cardiac troponin I (cTnl) in serum, heparinized plasma, or EDTA plasma using the ADVIA Centaur/ADVIA Centaur XP/ADVIA Centaur XPT/ADVIA Centaur CP systems. Cardiac troponin I determinations aid in the diagnosis of acute myocardial infarction and in the risk stratification of patients with non-ST segment elevation acute coronary syndromes with respect to relative risk of mortality, myocardial infarction (MI) or increased probability of ischemic events requiring urgent revascularization procedures.
System-to-system bias between the Tni-Ultra assay on the ADVIA Centaur CP and ADVIA Centaur/ADVIA Centaur XP/ADVIA Centaur XPT systems.
Monitor immunoassay method, troponin subunit automatically
DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.