KQOClass 1

Automated Urinalysis System

FDA adverse event reports and recall history for automated urinalysis system (product code KQO), from the public MAUDE/AEMS corpus and the live FDA recall database.

Total reports
1,085
Deaths
1
Injuries
24
Malfunctions
902

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 20122026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.

Recent recalls

  • 2023-05-17Open, ClassifiedUnder Investigation by firm

    iQ200 Series Urine Microscopy Analyzer: a) iQ 200 Select, Part Numbers C10684 and 700-3345; b) iQ 200 Elite, Part Numbers C10683 and 700-3375; in-vitro diagnostic device, urine analysis, c) iQ200 Select 1500, Part Number 700-3347, d) iQ200 Sprint, Part Number 700-3325

    Beckman Coulter has become aware of an intermittent issue in which the optional flag Previous Sample Had Sperm was enabled but not displayed so that a carryover event could have been investigated.

  • 2023-05-17Open, ClassifiedUnder Investigation by firm

    DxU 850m Iris Urine Microscopy Analyzer Catalog Number C49513, in-vitro diagnostic device, urine analysis

    Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have been investigated.

  • 2023-05-17Open, ClassifiedUnder Investigation by firm

    DxU 840m Iris Urine Microscopy Analyzer, Catalog Number C76947, in-vitro diagnostic device, urine analysis

    Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have been investigated.

  • 2023-05-17Open, ClassifiedUnder Investigation by firm

    UA WIN 10 Kit, Catalog Number C52900, in-vitro diagnostic device, urine analysis

    Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have been investigated.

  • 2020-05-09TerminatedNonconforming Material/Component

    iChemVELOCITY Urine Chemistry System-

    A defect in the SATA power adapter cable within the instruments computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computers metal enclosure, smoke emanating from PC6, flames contained within PC6, or an inability to use the devices CD-ROM.

  • 2020-05-09TerminatedNonconforming Material/Component

    iQ200 Series Urine Microscopy Analyzer- All part numbers

    A defect in the SATA power adapter cable within the instrument s computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computer s metal enclosure, smoke emanating from PC6, flames contained within PC6, or an inability to use the device s CD-ROM.

  • 2018-04-27TerminatedProcess control

    iChem VELOCITY Automated Urine Chemistry System, Catalog #700-7176-001 and iChem VELOCITY Automated Urine Chemistry System Computerless, Catalog #700-7177-001. The iChemVELOCITY automated urine chemistry system is an in-vitro diagnostic device used to automate the urine chemistry analysis profile using iChemVELOCITY Urine Chemistry Strips. The iChemVELOCITY can be used as a stand-alone system, as well as in a iQ200 Series system, a configuration given the proprietary name iRICELL as it is designed to be hardware and software compatible with iQ200 Series systems.

    The Strip Provider Modules (SPM) on some of the instruments had not been inspected during manufacture for correct seal placement and integrity.

  • 2017-03-01TerminatedDevice Design

    iChemVELOCITY Urine Chemistry System, Catalog No. 800-3530, 800-7167, 800-7719, 800-7720, 800-3564, 800-7100, 800-7103, 800-7106, 800-7162, 800-7163, 800-7166, 800-3061. The iChem Velocity automated urine chemistry system is used to test for the following analytes in urine: ketones, blood, protein, glucose, bilirubin, urobilinogen, nitrite, leukocyte esterase, specific gravity, pH, ascorbic acid, color, and clarity. The iChem Velocity automated urine chemistry analyzer is intended to be used for in vitro measurement of various analytes in urine; it is intended for use only with the iChem Velocity urine chemistry strips

    The probe misalignment or bending has the potential to lead to delay in reporting results or incorrect patient results.

  • 2015-08-21TerminatedNonconforming Material/Component

    iChemVELOCITY Automated Urine Chemistry Catalog No. 800-3049, 800-3050, 800-3079, 800-3080, 800-3530, 800-7167, 800-7190, 800-7713,800-7714, 800-7719, 800-7720. An in vitro diagnostic device used to automate the urine chemistry analysis profile using iChemVELOCITY Urine Chemistry Strips.

    Iris International is recalling the iChemVELOCITY systems because they do not maintain the on-board strip stability claim of 5 days at 18-28 degrees celsius at 20-80% relative humidity (RH) as stated in the iChemVELOCITY product labeling.

  • 2014-05-04TerminatedReprocessing Controls

    CLINITEK Status+ Analyzer (portable urine chemistry analyzer). Catalog number 10490946.

    Device with Sample Interference Notes (SIN) enabled not cleared for US marketing.