FDA adverse event reports naming Medtronic Navigation, Inc (littleton) as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: MEDTRONIC NAVIGATION, INC (LITTLETON).

Total reports
8,654
Deaths
7
Injuries
212
Malfunctions
8,435

Reports per year

18
19
20
21
22
23
24
25

Source: FDA adverse event reports by year received, 2018–2025. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Top device categories

Recent recalls

  • 2024-06-06Open, ClassifiedDevice Design

    O-arm O2 Imaging System-Mobile x-ray system designed for 2D fluoroscopic and 3D imaging for adult and pediatric patients. Product Number: BI70002000

    Potential for an electrical component of the O-arm" O2 Imaging System to intermittently shut off,failure can cause the potential for surgical delay, additional imaging, cancellation and rescheduling of surgery, or completion of surgery without the benefit of image guidance.

  • 2023-08-24Open, ClassifiedUnder Investigation by firm

    The O-arm O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging for adult and pediatric patients weighing 60 lbs. or greater and having an abdominal thickness greater than 16 cm and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects.

    Ground cable installed incorrectly.

  • 2019-12-31TerminatedDevice Design

    O-arm" O2 Imaging System, Product Number/CFN: BI-700-02000

    Due to a manufacturing issue affecting particular O-arm" O2 Imaging Systems, Medtronic has determined that the gantry tractor motor drive belt may loosen over an extended number of uses at a higher frequency in these systems. If the belt loosens, it may result in rotation of 3D images about the gantry isocenter. While the 3D image is anatomically accurate within the image itself and may be used to confirm therapy, its electronic registered location may be rotated relative to the actual physical position of the patient.

  • 2016-10-11TerminatedSoftware design

    Medtronic Navigation O-arm O2 Surgical Imaging System Product Catalog Number: BI-700-02000 Product Usage: The O-arm Surgical Imaging System is a multi-dimensional surgical imaging platform that is designed for use in spine, orthopaedic, and trauma-related surgeries. It provides real-time, intra-operative imaging of a patients anatomy with high quality images and a large field-of-view in both two and three dimensions

    O-arm O2 Surgical Imaging System Spatial calibration may be erroneous in Stealth Station navigated images

  • TerminatedSoftware design

    O-arm 1000 2nd Edition Imaging System, OARM ASSY Bl70000027 SYSTEM PRODUCT, Model Number Bl70000027 Product Usage: The O-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-arm Imaging System is compatible with certain Image Guided Surgery Systems.

    Software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version 3.1.7 will be installed on all 2nd edition imaging systems.

  • TerminatedSoftware design

    O-arm 1000 2nd Edition Imaging System, OARM ASSY Bl70000027R SYS PRODUCT RWK, Model Number Bl70000027R Product Usage: The O-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-arm Imaging System is compatible with certain Image Guided Surgery Systems.

    Software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version 3.1.7 will be installed on all 2nd edition imaging systems.

  • TerminatedSoftware design

    O-arm 1000 2nd Edition Imaging System, BASE OARM Bl70000027100 SYSTEM lOOV, Model Number Bl70000027100 Product Usage: The O-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-arm Imaging System is compatible with certain Image Guided Surgery Systems.

    Software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version 3.1.7 will be installed on all 2nd edition imaging systems.

  • TerminatedSoftware design

    O-arm 1000 2nd Edition Imaging System, BASE OARM Bl70000027100R SYSTEM lOOV RWK, Model Number Bl70000027100R Product Usage: The O-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-arm Imaging System is compatible with certain Image Guided Surgery Systems.

    Software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version 3.1.7 will be installed on all 2nd edition imaging systems.

  • TerminatedSoftware design

    O-arm 1000 2nd Edition Imaging System, BASE OARM Bl70000027120 SYSTEM 120V, Model Number Bl70000027120 Product Usage: The O-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-arm Imaging System is compatible with certain Image Guided Surgery Systems.

    Software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version 3.1.7 will be installed on all 2nd edition imaging systems.

  • TerminatedSoftware design

    O-arm 1000 2nd Edition Imaging System, BASE BASE OARM Bl70000027120R SYSTEM 120V RWK, Model Number Bl70000027120R Product Usage: The O-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-arm Imaging System is compatible with certain Image Guided Surgery Systems.

    Software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version 3.1.7 will be installed on all 2nd edition imaging systems.