Skip to main content
OWBClass 2

Interventional Fluoroscopic X-Ray System

FDA adverse event reports and recall history for interventional fluoroscopic x-ray system (product code OWB), from the public MAUDE corpus and the live FDA recall database.

Total reports
50,680
Deaths
108
Injuries
608
Malfunctions
49,897

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2026-10-07Open, ClassifiedProcess control

    Alphenix INFX-8000C/B5

    Loosening of the bolts securing the longitudinal movement mechanism may cause the tabletop to shift, resulting in [positional/imaging/patient] misalignment and a longitudinal motion error

  • 2026-08-25Open, ClassifiedProcess control

    Alphenix Models INFX-8000C INFX-8000F INFX-8000V This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

    Due to the fixing screws of ceiling movement gear in ceiling suspended arm that are part of the X-ray interventional systems may become loose.

  • 2026-08-10Open, ClassifiedSoftware design

    Allura Xper FD. Model Numbers: 722026, 722027, 722028, 722029, 722035, 722038, 722039, 722058. Interventional fluoroscopic x-ray system.

    Firmware issue potentially leading to loss of geometry movements.

  • 2026-08-10Open, ClassifiedSoftware design

    Philips Azurion Rystem, Model Numbers: 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222,722223, 722224, 722225, 722226, 722227, 722228, 722280*, 722281*, and 722282* (*These models are distributed only in China), with Software Version Number R2.2.10 and R2.1.10* (*This software version is distributed only in China); Interventional Fluoroscopic X-Ray System

    Two software related issues affecting Azurion R2.2.10 Systems: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate treatment and 2) Potential loss of geometry movements that may lead to delay of therapy and procedural complications

  • 2026-08-10Open, ClassifiedSoftware design

    Philips Azurion, Model Numbers: 722229, 722230, 722231, 722232, 722233, 722234, 722235, and 722236, with Software Version Numbers: R3.0 and R3.1; Interventional fluoroscopic x-ray system

    Two software related issues affecting Philips Azurion R3.0 and R3.1 systems: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate treatment and 2) Potential loss of geometry movements that may lead to delay of therapy and procedural complications.

  • 2026-08-05Open, ClassifiedUnder Investigation by firm

    Philips Allura under the following System Descriptions and corresponding model numbers: 1. Allura Xper FD10; Model Numbers: 722003, 722010, 722026. 2. Allura Xper FD10 OR Table; Model Number: 722033. 3. Allura Xper FD10/10; Model Numbers: 722005, 722011, 722027. 4. Allura Xper FD10C; Model Number: 722001. 5. Allura Xper FD20; Model Numbers: 722006, 722012, 722028. 6. Allura Xper FD20 Biplane; Model Numbers: 722008, 722013. 7. Allura Xper FD20 Biplane OR Table; Model Number: 722025. 8. Allura Xper FD20 OR Table; Model Numbers: 722015, 722023, 722035. 9. Allura Xper FD20/10; Model Number: 722029. 10. Allura Xper FD20/15; Model Number: 722058. 11. Allura Xper FD20/15 OR Table; Model Number: 722059. 12. Allura Xper FD20/20; Model Number: 722038. 13. Allura Xper FD20/20 OR Table; Model Number: 722039. 14. Allura Xper R9 7 M12; Model Number: 722065. 15. Allura Xper R9 7 M20; Model Number: 722066. Interventional Fluoroscopic X-Ray System.

    Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.

  • 2026-08-05Open, ClassifiedUnder Investigation by firm

    Philips Azurion under the following System Descriptions and corresponding model numbers: 1. Azurion 3 M12; Model Numbers: 722063, 722221. 2. Azurion 3 M15; Model Numbers: 722064, 722222, 722280. 3. Azurion 5 M12; Model Number: 722227. 4. Azurion 5 M20; Model Numbers: 722228, 722281. 5. Azurion 7 B12; Model Numbers: 722067, 722225. 6. Azurion 7 B20; Model Numbers: 722068, 722226. 7. Azurion 7 M12; Model Numbers: 722078, 722223. 8. Azurion 7 M20; Model Numbers: 722079, 722224, 722282. Interventional Fluoroscopic X-Ray System.

    Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.

  • 2026-08-05Open, ClassifiedUnder Investigation by firm

    Philips Allura, Cardio Vascular-Allura Centron. Model Number: 722400. Interventional Fluoroscopic X-Ray System.

    Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.

  • 2026-08-03Open, ClassifiedProcess change control

    Brand Name: OXOS Medical, Inc Product Name: MC2 Portable X-Ray System Model/Catalog Number: MC2, MC2.1 Software Version: 5.7.0 Product Description: The MC2 Portable X-Ray System ( MC2 System or MC2 ) is a portable and handheld x-ray system designed to aid clinicians with point of care visualization through diagnostic x-rays of extremities and shoulders. The device allows a clinician to select desired technique factors best suited for their patient anatomy. The MC2 consists of two major system components: the emitter and the cassette. The MC2 emitter and cassette are battery-powered and can be charged with either a wired or wireless charger. The system is intended to interface wirelessly to an external tablet or display when used with the OXOS Device App. The MC2 utilizes an infrared Tracking System and Viewfinder that allows the operator to position the emitter within a safe source-to-image distance (SID) and allows the operator to visualize the anatomy and x-ray field projection prior to initiating an x-ray acquisition. The MC2 also utilizes a LIDAR system to ensure patient safety by maintaining a safe source-to-skin distance. Component: The MC2 Portable X-Ray System is not a component.

    Due to a defective circuit board that overheats

  • 2026-07-01Open, ClassifiedSoftware design

    ARTIS icono biplane. Model Number: 11327600.

    Potential for loss of movement capability after adjusting the detector lift height to its maximum or minimum limits.

Monitor interventional fluoroscopic x-ray system automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.