Ear, Nose, And Throat Stereotaxic Instrument
FDA adverse event reports and recall history for ear, nose, and throat stereotaxic instrument (product code PGW), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 1,756
- Deaths
- 1
- Injuries
- 61
- Malfunctions
- 1,694
Reports per year
Source: FDA adverse event reports by year received, 2014–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2025-11-04Open, ClassifiedDevice Design
TruDi Navigation System - Multi Instrument Adapter; - Manufacturer's Product Number (Catalog Number): EM-2025-000F Rev 00 or Rev 01;
It was determined that when the navigation system is configured with specific revision combinations of the Multi Instrument Adapter together with the Patient Tracker the system may not meet its specified accuracy for visual verification of device location within the patient s anatomy.
- 2024-08-05Open, ClassifiedProcess control
TruDi NAV Suction, 0 Degrees; Model/Catalog Number: TDNS000Z. TruDi" NAV Suction Instruments are intended for use with the TruDi" Navigation System during surgical procedures in ENT medicine and skull base surgery
Specific lots of TruDi NAV Suction Instruments 0 Degrees were incorrectly calibrated which when used with the TruDi Navigation System, may result in a discrepancy between the actual tip location and the visualized location displayed on the system.
- 2024-05-23Open, ClassifiedSoftware Manufacturing/Software Deployment
8000-021-002, Scopis ENT Software with TGS with Sofware Version: NOVA 3.6.0 RC16. Scopis is a next-generation solution for navigating functional endoscopic sinus surgery (FESS) offering surgeons highly advanced image guidance and visualization capabilities in a single system. The Stryker Electromagnetic Navigation Unit (8000-010-003) powered by Scopis ENT Software with TGS (8000-021-002), allows the analysis and identification of sinus cells in the complex patient anatomy and planning of the natural drainage pathways through the sinus cavity. During surgery, the planned pathways are overlaid in real-time onto the endoscopic image, providing a unique Scopis augmented reality technology. Guidance of endoscopic instruments may help perform a minimally invasive, accurate and selective surgery.
Unreleased software was installed on customer systems resulting in the visual feedback on the screen to show the points to be off from the physical reference point of the pointer or suction.
- 2023-10-20Open, ClassifiedSoftware Design Change
TruDi Navigation System, Model: FG-2000-00, Catalog: ENS022B, when used with TruDi Curette, Models: TDC0005Z and TDC0005
When using affected curette and software, there is a discrepancy between the actual curette tip location and the location displayed on navigation systems intended for use during surgical procedures of the Ear Nose and Throat (ENT) and ENT skull base surgery, which may cause delayed/prolonged surgery, cerebrospinal fluid leak, visual impairment, or skull base structural damage.
- 2017-04-19TerminatedStorage
Stryker PROFESS Registration Kit. The Stryker PROFESS Registration Kit is a member of the Navigation portfolio and is used as an intraoperative guidance system to enable intranasal and sinus computer-assisted surgery. The system is comprised of PROFESS Registration Stickers, a PROFESS Patient Tracker, and a PROFESS Adaptor Cable.
Product shipped proximate to or past the expiration date listed on the product label.
- Open, ClassifiedProcess control
TruDi NAV Suction Instruments: 0-Degree, REF: TDNS000Z; 70-Degree, REF: TDNS070Z; and 90-Degree, REF: TDNS090Z
Suction Instruments were incorrectly calibrated so they may not meet the system accuracy specification for visual verification of device location in the patient anatomy. If the surgeon does not detect the accuracy error, this could lead to an intraprocedural delay, a potential for reversible tissue damage or damage to the carotid artery, CSF leak, or damage to optic nerve.
- TerminatedDevice Design
TruDi NAV Suction, TDNS000Z, 0 Degree Tip Angle, (01)10705031245877
The firm became aware that three (3) lots of the NAV Suction Instruments contain calibration issues causing incorrect tip orientation display in the system.
- TerminatedDevice Design
TruDi NAV Suction, TDNS070Z, 70 Degree Tip Angle (01) 10705031245884
The firm became aware that three (3) lots of the NAV Suction Instruments contain calibration issues causing incorrect tip orientation display in the system.
- TerminatedDevice Design
TruDi NAV Suction, TDNS090Z, 90 Degree Tip Angle, (01) 10705031245891
The firm became aware that three (3) lots of the NAV Suction Instruments contain calibration issues causing incorrect tip orientation display in the system.
- TerminatedDevice Design
Medtronic Straight Suction 9733449 EM ENT, reusable, Rx. The firm name on the label is Medtronic Navigation, Inc., Louisville, CO. The suction devices are part of a family of reusable ENT instruments that are used with the StealthStationTM AxiEM surgical navigation system. The AxiEM system is an electromagnetic (EM) tracking solution for surgical tools. The system measures an induced voltage created by local magnetic fields from a set of coils in the transmit coil array (TCA). Each surgical tool has a set of receiving coils to sense the magnetic field strength. Using a mathematical algorithm, the position and orientation of the instrument can be computed and overlaid/registered to imagery on a computer monitor, resulting in the ability to track the instrument tip in relation to the patients anatomy. AxiEM ENT Suction devices are used as suction and pointing devices. The suction devices are intended to have an ENT Instrument Tracker mounted on it
The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.
Monitor ear, nose, and throat stereotaxic instrument automatically
DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.