Neurological Stereotaxic Instrument

FDA adverse event reports and recall history for neurological stereotaxic instrument (product code HAW), from the public MAUDE corpus and the live FDA recall database.

Total reports
42,962
Deaths
104
Injuries
2,965
Malfunctions
39,630

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2026-04-29Open, ClassifiedDevice Design

    Brand Name: Leksell Vantage Arc System REF: 1053958

    Locking mechanisms may not properly secure to the Arc system, resulting in shifting of coordinates during procedures.

  • 2025-12-19Open, ClassifiedUnder Investigation by firm

    The Stryker CranialMask Tracker is a single-use device consisting of a flexible printed circuit board (PCB) with 31 infrared (IR) light-emitting diodes (LEDs) as well as an electronics control unit for the LEDs which includes its own battery power supply. The measured LED positions can be used in one of three different modes of operation: Patient tracking, Registration using surface matching, and Automatic intra-operative mask registration.

    When the device is activated during surgery, the device software issues an error message "Instrument Face Mask (6001385000000) is not valid for this application" resulting in an inability to use the device.

  • 2025-11-05Open, ClassifiedProcess change control

    Brand Name: Leksell Stereotactic System Product Name: Leksell Vantage Stereotactic System Software Version: n/a Product Description: The Intended Purpose of Leksell Vantage" Stereotactic System is target localization and fixation of the patient head in a coordinate system in order to perform stereotactic neurosurgical procedures, for example; deep brain stimulation, lesioning, biopsies, targeted injections, aspirations and minimal invasive tumor treatments.

    Due to increased risk that paint flakes from the engravings may separate from the device and there is the possibility that the flake may inadvertently enter the wound

  • 2024-11-08Open, ClassifiedProcess change control

    microTargeting Insertion Tube Set, P/N 66-IT-05P, Lot 244517. The intended use is to accurately guide a microelectrode or instrument in the brain during stereotactic functional neurosurgical procedures.

    A complaint was received (MDR 3002250546-2024-00003) prompting an investigation. The investigation revealed a schematic for a spacer tube which is too long relative to the size of other components was included in the build order for Lot 244517 of P/N 66-IT-06P. Use of the product with a spacer tube which is too long could result in intercranial hemorrhaging and brain damage.

  • 2024-11-06Open, ClassifiedUnder Investigation by firm

    Elekta Disposable Biopsy Needle Kit for Leksell Stereotactic System, REF 911933, Diameter / 14 Gauge, containing 6 inner boxes labeled REF 911761, Disposable Biopsy Needle Kit for Leksell Stereotactic System. Each inner box of REF 911761 contains a Tyvek bag labeled REF 912465, Disposable Biopsy Needle and Aspirator Tube. The system is used for for localization and diagnosis of intracranial disorders and their surgical treatment including radiosurgery and stereotactic radiation therapy.

    Elekta has become aware that Disposable Biopsy Needles can contain some microscopic debris on the inside of the biopsy needle. The material in the debris is stainless steel, same material as the biopsy needle. No debris has been found on the outer parts of the biopsy needles.

  • 2024-05-09Open, ClassifiedSoftware design

    Stealth S8 Clinical Software Application, REF: 9735762, CE0344, Rx Only. Used in the StealthStation System as an aid for locating anatomical structures in either open or percutaneous neurosurgical procedures.

    Due to a software issue, there is the potential for a missing digit, letter, or decimal in either the "distance to target" or "tip stop point" display during navigation.

  • 2024-04-16Open, ClassifiedUnder Investigation by firm

    Elekta Disposable Biopsy Needle Kit for Leksell Stereotactic System, REF 911933, 2.1mm Diameter / 14 Gauge, containing 6 inner boxes labeled REF 911761, Disposable Biopsy Needle Kit for Leksell Stereotactic System. Each inner box of REF 911761 contains a Tyvek bag labeled REF 912465, Disposable Biopsy Needle and Aspirator Tube.

    Elekta has become aware that Disposable Biopsy Needles (911933) from one batch (837838839) can contain some microscopic debris on the inside of the biopsy needle. The material in the debris is stainless steel, same material as the biopsy needle. No debris has been found on the outer parts of the biopsy needles. The sterility of the biopsy needles has not been affected. This issue has been reported from one site.

  • 2024-04-04Open, ClassifiedProcess control

    NaviNetics Skull Anchor Key and Drill Kit, REF NN1215

    Defective Skull Anchor Key. After attachment of the single use Skull Anchor Key to the patient, the user was unable to attach the CT Localizer to perform stereotactic Imaging. Attachment of the localizer is typically accomplished by fully screwing in the four thumbscrews of the localizer into the threaded holes found on the top surface of the Skull Anchor Key.

  • 2024-02-06Open, ClassifiedPackage design/selection

    Globus Medical Cranial Drill with Stop, 2.40mm - Instruments for access, preparation, and placement of cranial devices with the ExcelsiusGPS. Ref : 6143.1834S

    Sterile-packed Cranial Drills may not be sterile due to a breach in the packaging and/or may have discoloration near the drill tip.

  • 2024-02-06Open, ClassifiedPackage design/selection

    Globus Medical Cranial Drill with Stop, 3.20mm - instruments for access, preparation, and placement of cranial devices with the ExcelsiusGPS. Ref: 6143.1842S

    Sterile-packed Cranial Drills may not be sterile due to a breach in the packaging and/or may have discoloration near the drill tip.