System, X-Ray, Mobile
FDA adverse event reports and recall history for system, x-ray, mobile (product code IZL), from the public MAUDE/AEMS corpus and the live FDA recall database.
- Total reports
- 922
- Deaths
- 8
- Injuries
- 158
- Malfunctions
- 701
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.
Recent recalls
- 2026-07-13Open, ClassifiedProcess control
Optima XR200amx, Mobile Digital-Ready Radiographic System
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.
- 2026-07-13Open, ClassifiedProcess control
Optima XR220amx, Mobile Digital-Ready Radiographic System
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.
- 2026-07-13Open, ClassifiedProcess control
Optima XR240amx, Mobile Digital Radiographic System
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.
- 2026-04-22Open, ClassifiedComponent design/selection
Rover Mobile X-ray System Model/Catalog Number: MXU-RV35 and MXU-RV71
X-Ray generator may malfunction resulting in the X-Ray being inoperable.
- 2025-11-07Open, ClassifiedDevice Design
GM85 Digital Mobile X-ray imaging System; Model Number: GM85;
The four (4) M5 bolts that secure the Tube Head Unit to the end of the arm may loosen or fracture due to metal fatigue.
- 2025-10-23Open, ClassifiedDevice Design
Sedecal Mobile Digital Diagnostic X-Ray Systems; SM-40HF-B-D-KM (AeroDR X30)
Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
- 2025-10-23Open, ClassifiedDevice Design
MobileDiagnost wDR 2.2; Model Number REF 9890-010-89522, Mobile X-ray system
Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
- 2025-10-23Open, ClassifiedDevice Design
Model Number 40KWFX, Mobile X-ray system
Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
- 2025-10-23Open, ClassifiedDevice Design
Model Number 40KW-FX.899, Mobile X-ray system
Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
- 2025-10-23Open, ClassifiedDevice Design
Model Number 40KWFXPLUS, Mobile X-ray system
Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
Monitor system, x-ray, mobile automatically