W. L. Gore & Associates, Inc.
FDA adverse event reports naming W. L. Gore & Associates, Inc. as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: W. L. GORE & ASSOCIATES, INC..
- Total reports
- 14,704
- Deaths
- 615
- Injuries
- 12,123
- Malfunctions
- 1,962
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Top device categories
- System, Endovascular Graft, Aortic Aneurysm TreatmentMIH8,450
- Iliac Covered Stent, ArterialPRL1,264
- Stent, Superficial Femoral ArteryNIP1,168
- Introducer, CatheterDYB1,097
- Mesh, Surgical, PolymericFTL812
- Transcatheter Septal OccluderMLV629
- Prosthesis, Vascular Graft, Of 6mm And Greater DiameterDSY413
- System, Endovascular Graft, Arteriovenous (Av) Dialysis Access Circuit Stenosis TreatmentPFV161
- Endovascular System For Treatment Of Thoracoabdominal And Pararenal Aortic LesionsQZK146
- Catheter, PercutaneousDQY141
Recent recalls
- 2026-10-14Open, ClassifiedUnder Investigation by firm
Generic Name: GORE VIABIL Short Wire Biliary Endoprosthesis; Model/Catalog Number: VSWVH0808 and VSWVN1006 Product Description: The GORE VIABIL Short Wire Biliary Endoprosthesis is a flexible, self-expanding endoprosthesis designed to maintain luminal patency of the common bile duct which may be obstructed. The endoprosthesis is radially compressed and secured onto the distal end of a catheter-based delivery system for deployment. The catheter provides a means for accessing and implanting the GORE VIABIL Short Wire Biliary Endoprosthesis at the target site in the biliary tract.
The potential for the devices to be mislabeled with the incorrect length.
- 2026-08-07Open, ClassifiedProcess control
GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VSWVN1010. Flexible, self-expanding endoprosthesis designed to maintain luminal patency of the common bile duct which may be obstructed.
10 mm diameter x 7 cm with transmural drainage holes (VSWVH1006) incorrectly labeled as 10 mm diameter x 10 cm length devices without transmural drainage holes (VSWVN1010).
- 2026-08-07Open, ClassifiedProcess control
GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VSWVN1006. Flexible, self-expanding endoprosthesis designed to maintain luminal patency of the common bile duct which may be obstructed.
10 mm diameter x 7 cm with transmural drainage holes (VSWVH1006) incorrectly labeled as 10 mm diameter x 10 cm length devices without transmural drainage holes (VSWVN1010).
- 2026-07-02Open, ClassifiedUnder Investigation by firm
GORE TAG Thoracic Branch Endoprosthesis Side Branch Component Catalog Numbers (by Region) United States (includes Puerto Rico) and Canada: TSB080806A, TSB081006A, TSB081206A, TSB081506A, TSB081706A, TSB121506A, TSB121706A, TSB122006A EMEA Countries, Australia, New Zealand: TSB080806E, TSB081006E, TSB081206E, TSB081506E, TSB081706E, TSB121506E, TSB121706E, TSB122006E Japan: TSB080806J, TSB081006J, TSB081206J, TSB081506J, TSB081706J, TSB121506J, TSB121706J, TSB122006J China and Taiwan: TSB080806W, TSB081006W, TSB081206W, TSB081506W, TSB081706W, TSB121506W, TSB121706W, TSB122006W The GORE TAG Thoracic Branch Endoprosthesis is indicated for endovascular repair of lesions of the aortic arch and descending thoracic aorta, while maintaining flow into a single aortic arch branch vessel The GORE TAG Thoracic Branch Endoprosthesis is a modular device consisting of the Aortic Component (AC), the Side Branch (SB) Component, and an optional Aortic Extender (AE), and their respective delivery systems. These components may be used together as a stand-alone device or in conjunction with the GORE TAG Conformable Thoracic Endoprosthesis to accommodate the intended treatment site.
Due to catheter separation
- 2025-10-02Open, ClassifiedUnknown/Undetermined by firm
GORE ACUSEAL Vascular Graft, REF: ECH050020J, ECH050020W, ECH050050J, ECH050050W, ECH060010A, ECH060020A, ECH060020J, ECH060020W, ECH060040, ECH060040A, ECH060040W, ECH060050A, ECH060050J, ECH460045A, ECH460045J, ECH470045, ECH470045A; used with Instructions for Use.
Reports of vascular graft delamination, which may lead to reduced access flow; cannulation difficulties; bleeding or bruising; and harms related to reintervention procedures, so Instructions for Use will be updated to modify existing warning to 1) further clarify techniques that may contribute to the risk of delamination, 2) Addition of delamination to device-related adverse event section.
- 2024-08-30Open, ClassifiedPackage design/selection
Heparin, Gore Viabahn VBX, Balloon Expandable Endoprosthesis, for the following Reference numbers: United States: BXB051501A BXB051502A BXB051901A BXB051902A BXB052901A BXB052902A BXB053901A BXB053902A BXB055901A BXB055902A BXB057901A BXB057902A BXB061501A BXB061502A BXB061901A BXB061902A BXB062901A BXB062902A BXB063901A BXB063902A BXB065901A BXB065902A BXB067901A BXB067902A BXB071501A BXB071502A BXB071901A BXB071902A BXB072901A BXB072902A BXB073901A BXB073902A BXB075901A BXB075902A BXB077901A BXB077902A BXB082901A BXB082902A BXB083901A BXB083902A BXB085901A BXB085902A BXB087901A BXB087902A BXBL082901A BXBL082902A BXBL083901A BXBL083902A BXBL085901A BXBL085902A BXBL087901A BXBL087902A BXB092901A BXB092902A BXB093901A BXB093902A BXB095901A BXB095902A BXB097901A BXB097902A Japan: BXB073901J BXB073902J BXB075901J BXB075902J BXB077901J BXB077902J
Due to packaging defects, sterility assurance and heparin activity may be compromised.
- 2024-08-30Open, ClassifiedPackage design/selection
Heparin, Gore Viabahn VBX, Balloon Expandable Endoprosthesis, for the following Reference numbers: United States: BXB055902C, BXB063902C, BXB065902C, BXB067902C, BXB073902C, BXB075902C, BXB077902C, BXB083902C, BXB085902C, BXB087902C, BXB093902C, BXB095902C, BXB097902C *For Clinical use.
Due to packaging defects, sterility assurance and heparin activity may be compromised.
- 2023-02-01Open, ClassifiedEmployee error
GORE CARDIOFORM Septal Occluder, REF: GSX0030A
Septal Occluders released with incomplete quality test documentation that assesses the release of the device from the delivery system during the index procedure. A potential failure of this quality test could potentially result could result in occluder embolization, anoxic brain injury, central venous or arterial perforation or rupture.
- 2022-10-26Open, ClassifiedEmployee error
GORE CARDIOFORM ASD Occluder. cardiovascular implant.
Due to manufacturing records (Release Test Results) indicating "Failed".
- 2020-05-20TerminatedLabeling Change Control
Conformable Thoracic Stent Graft, Rx only, Sterile EO - Product Usage: This device is intended for endovascular repair of lesions of the descending thoracic aorta.
The incorrect label was placed on both the primary and secondary package.