Stent, Superficial Femoral Artery

FDA adverse event reports and recall history for stent, superficial femoral artery (product code NIP), from the public MAUDE corpus and the live FDA recall database.

Total reports
10,730
Deaths
186
Injuries
5,036
Malfunctions
5,494

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2024-01-02CompletedPackaging change control

    Protege GPS Self-Expanding Peripheral Stent System, product number SERB65-09-60-80, Sterile.

    Seal defects could compromise the ability of the product packaging to maintain sterility.

  • 2018-02-22Open, ClassifiedNonconforming Material/Component

    Innova Self-Expanding Stent System, 200 mm. Labeled with Material Number and description as follows: Product Usage: The Innova Vascular Self-Expanding Stent System is indicated to improve luminal diameter in the treatment of symptomatic de-novo or restenotic lesions in the native superficial femoral artery (SFA) and/or proximal popliteal artery (PPA) with reference vessel diameters from 4.0 mm to 7.0 mm and lesion lengths up to 190 mm. a. H74939293052030 (Material Description Innova 5 x 200 x 130); b. H74939293052070 (Material Description Innova 5 x 200 x 75); c. H74939293062030 (Material Description Innova 6 x 200 x 130); d. H74939293062070 (Material Description Innova 6 x 200 x 75); e. H74939293072030 (Material Description Innova 7 x 200 x 130); f. H74939293072070 (Material Description Innova 7 x 200 x 75); g. H74939293082030 (Material Description Innova 8 x 200 x 130); h. H74939293082070 (Material Description Innova 8 x 200 x 75)

    Stent possibly unable to be fully released from the delivery system.

  • 2018-02-22Open, ClassifiedNonconforming Material/Component

    Innova Self-Expanding Stent System, 180mm. Labeled with Material Number and description as follows: Product Usage: The Innova Vascular Self-Expanding Stent System is indicated to improve luminal diameter in the treatment of symptomatic de-novo or restenotic lesions in the native superficial femoral artery (SFA) and/or proximal popliteal artery (PPA) with reference vessel diameters from 4.0 mm to 7.0 mm and lesion lengths up to 190 mm. a. H74939180051870 (Material Description Innova-EU 5 x 180 x 75); b. H74939180061870 (Material Description Innova-EU 6 x 180 x 75); c. H74939180071870 (Material Description Innova-EU 7 x 180 x 75); d. H74939180071870 (Material Description Innova-EU 8 x 180 x 75); e. H74939181051830 (Material Description Innova-EU 5 x 180 x 130); f. H74939181061830 (Material Description Innova-EU 6 x 180 x 130); g. H74939181071830 (Material Description Innova-EU 7 x 180 x 130); h. H74939181081830 (Material Description Innova-EU 8 x 180 x 130); i. H74939259051830 (Material Description Innova-China 5 x 180 x 130); j. H74939259061830 (Material Description Innova-China 6 x 180 x 130); k. H74939259071830 (Material Description Innova-China 7 x 180 x 130); l. H74939259081830 (Material Description Innova-China 8 x 180 x 130); m. H74939259081870 (Material Description Innova-China 8 x 180 x 75); n.H74939293051830 (Material Description Innova 5 x 180 x 130); o. H74939293051870 (Material Description Innova 5 x 180 x 75); p. H74939293061830 (Material Description Innova 6 x 180 x 130); q. H74939293061870 (Material Description Innova 6 x 180 x 75); r. H74939293071830 (Material Description Innova 7 x 180 x 130); s. H74939293071870 (Material Description Innova 7 x 180 x 75); t. H74939293081830 (Material Description Innova 8 x 180 x 130); u. H74939293081870 (Material Description Innova 8 x 180 x 75);

    Stent possibly unable to be fully released from the delivery system.

  • 2013-10-10TerminatedDevice Design

    Bard LifeStent Solo Vascular Stent, Item Numbers: EX062001CL, EX072001CL, EX062003CL, EX072003CL. The LifeStent Solo Vascular Stent System is intended to improve luminal diameter in the treatment of symptomatic de-novo or restenotic lesions up to 240 mm in length in the native superficial femoral artery (SFA) and in the proximal popliteal artery with reference vessel diameters ranging from 4.0 6.5 mm.

    Bard Peripheral Vascular is recalling specific lots of Bard Lifestent Solo Vascular Stent because they have determined a higher incidence rate of partial stent deployment associated with their use.

  • 2013-06-06TerminatedPackage design/selection

    FLEXSTENT Femoropopliteal Self-Expanding Stent System. Product Usage: The FSS FLEXSTENT Femoropopliteal SES clinical product is indicated for treatment of atherosclerotic de novo native superficial femoral artery lesions and proximal popliteal lesions.

    Holes and/or tears can be introduced in the FLEXSTENT inner pouch under some conditions, and to a lesser degree, in the outer pouch.

  • 2013-06-06TerminatedPackage design/selection

    FLEXSTENT Femoropopliteal Self-Expanding Stent System. Product Usage: The FSS FLEXSTENT Femoropopliteal SES clinical product is indicated for treatment of atherosclerotic de novo native superficial femoral artery lesions and proximal popliteal lesions.

    Holes and/or tears can be introduced in the FLEXSTENT inner pouch under some conditions, and to a lesser degree, in the outer pouch.

  • 2013-06-06TerminatedPackage design/selection

    FLEXSTENT Femoropopliteal Self-Expanding Stent System. Product Usage: The FSS FLEXSTENT Femoropopliteal SES clinical product is indicated for treatment of atherosclerotic de novo native superficial femoral artery lesions and proximal popliteal lesions.

    Holes and/or tears can be introduced in the FLEXSTENT inner pouch under some conditions, and to a lesser degree, in the outer pouch.

  • 2011-06-16TerminatedDevice Design

    Boston Scientific Innova Over-the-Wire Self-Expanding Stent System, Catalog No. 39180-05107. 5mm x 100 mm x 75 cm (stent diameter x stent length x delivery system length). Made in USA, Two Scimed Place, Maple Grove, MN 55311 USA. The Innova Self-Expanding Stent System is comprised of two components: the implantable endoprosthesis and the stent delivery system. The stent is a laser cut self-expanding stent composed of a nickel titanium alloy (Nitinol). On both the proximal and distal ends of the stent, radiopaque markers made of tantalum increase visibility of the stent to aid in placement. The stent is constrained within a 6F (2.1 mm maximum OD) delivery system. The delivery system is a triaxial design with an outer shaft to stabilize the stent delivery system, a middle shaft to protect and constrain the stent, and an inner shaft to provide a guidewire lumen. The delivery system is compatible with 0.035 in (0.89 mm) guidewires. When ready to be implanted, the stent is deployed by retracting the middle shaft of the delivery system. A radiopaque marker at the distal end of the delivery system aids in visibility during deployment. As the stent is exposed to body temperature, it expands to appose the vessel wall. The Innova Self-Expanding Stent is available in a variety of diameters and lengths. The delivery system is also offered in two working lengths (75 cm and 130 cm). The Innova Self-Expanding Stent System is indicated for the treatment of peripheral vascular lesions

    Boston Scientific Corporation is conducting a recall of the Innova Self-Expanding Stent System because they have received 6 complaints involving no deployment/partial deployment of the Innova Stent. Potential health hazard events resulting from this type of failure include increased procedure time, vessel wall injury, and/or emergency surgery to remove the partially deployed stent. This recall d

  • 2011-06-16TerminatedDevice Design

    Boston Scientific Innova Over-the-Wire Self-Expanding Stent System, Catalog No. 39180-05127. 5 mm x 120 mm x 75 cm (stent diameter x stent length x delivery system length). Made in USA, Two Scimed Place, Maple Grove, MN 55311 USA. The Innova Self-Expanding Stent System is comprised of two components: the implantable endoprosthesis and the stent delivery system. The stent is a laser cut self-expanding stent composed of a nickel titanium alloy (Nitinol). On both the proximal and distal ends of the stent, radiopaque markers made of tantalum increase visibility of the stent to aid in placement. The stent is constrained within a 6F (2.1 mm maximum OD) delivery system. The delivery system is a triaxial design with an outer shaft to stabilize the stent delivery system, a middle shaft to protect and constrain the stent, and an inner shaft to provide a guidewire lumen. The delivery system is compatible with 0.035 in (0.89 mm) guidewires. When ready to be implanted, the stent is deployed by retracting the middle shaft of the delivery system. A radiopaque marker at the distal end of the delivery system aids in visibility during deployment. As the stent is exposed to body temperature, it expands to appose the vessel wall. The Innova Self-Expanding Stent is available in a variety of diameters and lengths. The delivery system is also offered in two working lengths (75 cm and 130 cm). The Innova Self-Expanding Stent System is indicated for the treatment of peripheral vascular lesions

    Boston Scientific Corporation is conducting a recall of the Innova Self-Expanding Stent System because they have received 6 complaints involving no deployment/partial deployment of the Innova Stent. Potential health hazard events resulting from this type of failure include increased procedure time, vessel wall injury, and/or emergency surgery to remove the partially deployed stent. This recall d

  • 2011-06-16TerminatedDevice Design

    Boston Scientific Innova Over-the-Wire Self-Expanding Stent System, Catalog No. 39180-05157. 5 mm x 150 mm x 75 cm (stent diameter x stent length x delivery system length). Made in USA, Two Scimed Place, Maple Grove, MN 55311 USA. The Innova Self-Expanding Stent System is comprised of two components: the implantable endoprosthesis and the stent delivery system. The stent is a laser cut self-expanding stent composed of a nickel titanium alloy (Nitinol). On both the proximal and distal ends of the stent, radiopaque markers made of tantalum increase visibility of the stent to aid in placement. The stent is constrained within a 6F (2.1 mm maximum OD) delivery system. The delivery system is a triaxial design with an outer shaft to stabilize the stent delivery system, a middle shaft to protect and constrain the stent, and an inner shaft to provide a guidewire lumen. The delivery system is compatible with 0.035 in (0.89 mm) guidewires. When ready to be implanted, the stent is deployed by retracting the middle shaft of the delivery system. A radiopaque marker at the distal end of the delivery system aids in visibility during deployment. As the stent is exposed to body temperature, it expands to appose the vessel wall. The Innova Self-Expanding Stent is available in a variety of diameters and lengths. The delivery system is also offered in two working lengths (75 cm and 130 cm). The Innova Self-Expanding Stent System is indicated for the treatment of peripheral vascular lesions

    Boston Scientific Corporation is conducting a recall of the Innova Self-Expanding Stent System because they have received 6 complaints involving no deployment/partial deployment of the Innova Stent. Potential health hazard events resulting from this type of failure include increased procedure time, vessel wall injury, and/or emergency surgery to remove the partially deployed stent. This recall d

Monitor stent, superficial femoral artery automatically