Prosthesis, Vascular Graft, Of 6mm And Greater Diameter

FDA adverse event reports and recall history for prosthesis, vascular graft, of 6mm and greater diameter (product code DSY), from the public MAUDE/AEMS corpus and the live FDA recall database.

Total reports
5,150
Deaths
374
Injuries
3,287
Malfunctions
1,022

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 20122026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.

Recent recalls

  • 2026-03-04Open, ClassifiedError in labeling

    Gelsoft Plus, REF: 635024PE, 631514PE, 631222PE, 633008PE, 633010PE, 633007PE, 636010PE, 632522PE, 633012PE, 635018PE, 631516PE, 636016PE, 631220PE, 631608PE, 631206PE, 631407P50E, 631508PE, 636008PE, 636006PE, 632518PE, 634008PE, 634006PE, 631218PE, 632211PE, 634007PE, 633006PE, 631506PE, 632524PE, 636007PE, 631512PE, 636012PE, 636014PE, 632211P50E, 633016PE, 631407PE, 631809PE, 632010PE, 631608P50E, 632520PE, 632412PE, 631510PE, 635022PE, 631224PE, 632412P50E, 631507PE, 633014PE, 632010P50E, 635020PE, 631809P50E. Gelweave, REF: 736038E, 736036E, 736034E, 736032ST/10DE, 736032E, 736030ST/10DE, 736030E, 736028ST/10DE, 736028E, 736026ST/10DE, 736026E, 736024ST/10DE, 736024E, 736022ST/10DE, 736022E, 736020E, 736018E, 736016E, 736014E, 736012E, 736010E, 736008E, 736006E, 735034ST/8AE, 735034/8RME, 735034/10SEE, 735034/10RME, 735032STE, 735032ST/8AE, 735032/8SE, 735032/8RME, 735032/10SEE, 735032/10RME, 735030STE, 735030ST/8AE, 735030/8SE, 735030/8RME, 735030/10SEE, 735030/10RME, 735028STE, 735028ST/8AE, 735028E, 735028/8SE, 735028/8RME, 735028/10RME, 735026STE, 735026ST/8AE, 735026E, 735026/8SE, 735026/8RME, 735026/10RME, 735024STE, 735024ST/8AE, 735024E, 735024/8SE, 735024/8RME, 735024/10SEE, 735024/10RME, 735022STE, 735022E, 735022/8SE, 735022/8RME, 735020E, 735020/8RME, 735018E, 734038E, 734036E, 734034E, 734034CX4RMSE, 734034/8E, 734034/10E, 734032E, 734032CX4RMSE, 734032/8E, 734032/10E, 734030E, 734030CX4RMSE, 734030/8E, 734030/10E, 734028CX4RMSE, 734028/8E, 734028/10E, 734026CX4RMSE, 734026/8E, 734026/10E, 734024CX4RMSE, 734024/8E, 734024/10E, 734022CX4RMSE, 734022/8E, 734022/10E, 734020CX4RMSE, 734020/8E, 734020/10E, 734018/8E, 73341088E, 73341088/8RME, 73341088/8E, 73341088/10RME, 73341088/10E, 73321488/8ARME, 7332148/10RMFE, 7332148/10RMEE, 733212810/10STDE, 7332128/10RMFE, 7332128/10RMEE, 73321088E, 73321088/8SDE, 73321088/8RME, 73321088/8E, 73321088/10RME, 73321088/10E, 7330888/8E, 7330668/6E, 733038E, 733036E, 733034E, 733032E, 733030E, 733028E, 733026E, 733024E, 733024/6AE, 733022E, 733022/6AE, 733020E, 733020/6AE, 733018E, 733018/6AE, 733016E, 733014E, 73301488/8ARME, 7330148/10RMFE, 7330148/10RMEE, 73301412/10ETE, 733012E, 733012810/10STDE, 7330128/10RMFE, 7330128/10RMEE, 733010E, 73301088E, 73301088/8SDE, 73301088/8RME, 73301088/8E, 73301088/10RME, 73301088/10E, 733010108/8S4E, 733010108/8RME, 733010/10RME, 733008E, 733006E, 7328888/8E, 7328668/6E, 73281488/8ARME, 7328148/10RMEE, 73281412/10ETE, 73281288/8E, 732812810/10STDE, 7328128/10RMFE, 7328128/10RMEE, 73281088E, 73281088/8SDE, 73281088/8RME, 73281088/8E, 73281088/10RME, 73281088/10E, 732810108/8S4E, 7326888/8E, 7326668/6E, 73261488/8ARME, 7326148/10RMEE, 732612810/10STDE, 7326128/10RMFE, 7326128/10RMEE, 73261088E, 73261088/8SDE, 73261088/8RME, 73261088/8E, 73261088/10RME, 73261088/10E, 732610108/8S4E, 732610/10RME, 732528E, 732526E, 732524E, 732522E, 732520E, 732518E, 7324888/8E, 7324668/6E, 73241488/8ARME, 7324148/10RMEE, 732412E, 732412810/10STDE, 7324128/10RMEE, 732412/6E, 73241088E, 73241088/8SDE, 73241088/8RME, 73241088/8E, 73241088/10RME, 73241088/10E, 732410108/8S4E, 73221488/8ARME, 732214/10/10CRME, 732212810/10STDE, 732212/10/8CRME, 732212/10/8ARME, 732211E, 73221088/8RME, 73221088/8E, 73221088/10E, 732210108/8S4E, 7320888/8E, 7320668/6E, 732038E, 732034E, 732034CX4RME, 732034/8E, 732034/10E, 732032E, 732032CX4RME, 732032/8E, 732032/10E, 732030E, 732030CX4RME, 732030/8E, 732030/10E, 732028CX4RME, 732028/8E, 732028/10E, 732026CX4RME, 732026/8E, 732026/10E, 732024CX4RME, 732024/8E, 732024/10E, 732022/10E, 732014/8X2E, 732014/8X2AE, 732014/10X2E, 732014/10X2AE, 732012810/10STDE, 732012/8X2E, 732012/8X2AE, 732012/10X2E, 732010E, 73201088/8RME, 73201088/8E, 732010108/8S4E, 732010/6E, 731809E, 731809/6E, 731608XL55E, 731608E, 731608/8/10RME, 731608/6E, 731608/10/10RME, 731516E, 731514E, 731512E, 731510E, 731508E, 731506E, 7314108/10RME, 731407XL55E,

    Gelatin sealed woven polyester prostheses devices, designed for vascular repair, were incorrectly labeled with an additional month of shelf life. Device expiry should have been calculated from gel impregnation (date of manufacture) but it was incorrectly calculated from device packaging.

  • 2025-10-02Open, ClassifiedUnknown/Undetermined by firm

    GORE ACUSEAL Vascular Graft, REF: ECH050020J, ECH050020W, ECH050050J, ECH050050W, ECH060010A, ECH060020A, ECH060020J, ECH060020W, ECH060040, ECH060040A, ECH060040W, ECH060050A, ECH060050J, ECH460045A, ECH460045J, ECH470045, ECH470045A; used with Instructions for Use.

    Reports of vascular graft delamination, which may lead to reduced access flow; cannulation difficulties; bleeding or bruising; and harms related to reintervention procedures, so Instructions for Use will be updated to modify existing warning to 1) further clarify techniques that may contribute to the risk of delamination, 2) Addition of delamination to device-related adverse event section.

  • 2024-06-04Open, ClassifiedEquipment maintenance

    ADVANTA VXT, 6X40, 1GDS, NH, STR-SW. Single-Ended Slider. Vascular graft

    Complaints of reported separation of the Slider GDS Swivel Rod from the Swivel Core and of a notable gap between the two pieces.

  • 2024-06-04Open, ClassifiedEquipment maintenance

    ADVANTA VXT, 8X40, 1GDS, NH, STR-SW. Single-Ended Slider. Vascular graft

    Complaints of reported separation of the Slider GDS Swivel Rod from the Swivel Core and of a notable gap between the two pieces.

  • 2024-06-04Open, ClassifiedEquipment maintenance

    ADVANTA VXT, 6X50, 1GDS, NH, STR-SW. Single-Ended Slider. Vascular graft

    Complaints of reported separation of the Slider GDS Swivel Rod from the Swivel Core and of a notable gap between the two pieces.

  • 2024-06-04Open, ClassifiedEquipment maintenance

    ADVANTA VXT, 7X50, 1GDS, NH, STR-SW. Single-Ended Slider. Vascular graft

    Complaints of reported separation of the Slider GDS Swivel Rod from the Swivel Core and of a notable gap between the two pieces.

  • 2024-06-04Open, ClassifiedEquipment maintenance

    ADVANTA VXT, 8X50, 1GDS, NH, STR-SW. Single-Ended Slider. Vascular graft

    Complaints of reported separation of the Slider GDS Swivel Rod from the Swivel Core and of a notable gap between the two pieces.

  • 2024-06-04Open, ClassifiedEquipment maintenance

    ADVANTA VXT, 6X70, 1GDS, NH, STR-SW. Single-Ended Slider. Vascular graft

    Complaints of reported separation of the Slider GDS Swivel Rod from the Swivel Core and of a notable gap between the two pieces.

  • 2024-06-04Open, ClassifiedEquipment maintenance

    ADVANTA VXT, 7X70, 1GDS, NH, STR-SW. Single-Ended Slider. Vascular graft

    Complaints of reported separation of the Slider GDS Swivel Rod from the Swivel Core and of a notable gap between the two pieces.

  • 2024-06-04Open, ClassifiedEquipment maintenance

    ADVANTA VXT, 8X70, 1GDS, NH, STR-SW. Single-Ended Slider. Vascular graft

    Complaints of reported separation of the Slider GDS Swivel Rod from the Swivel Core and of a notable gap between the two pieces.