Transcatheter Septal Occluder
FDA adverse event reports and recall history for transcatheter septal occluder (product code MLV), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 7,010
- Deaths
- 201
- Injuries
- 4,058
- Malfunctions
- 2,664
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2023-02-01Open, ClassifiedEmployee error
GORE CARDIOFORM Septal Occluder, REF: GSX0030A
Septal Occluders released with incomplete quality test documentation that assesses the release of the device from the delivery system during the index procedure. A potential failure of this quality test could potentially result could result in occluder embolization, anoxic brain injury, central venous or arterial perforation or rupture.
- 2022-10-26Open, ClassifiedEmployee error
GORE CARDIOFORM ASD Occluder. cardiovascular implant.
Due to manufacturing records (Release Test Results) indicating "Failed".
- 2015-07-24TerminatedOther
Medical Device Exchange 10MM VASCULAR; Pouch, Tyvek/Mylar, Sterile, Single Use Only; OEM Catalog number 9-AVP2-010; EXP (repackager), Aga Medical (OEM) Transcatheter septal occluder
EXP did not register as a medical device establishment, list the devices being recalled, or establish its own quality system and instead relied on the fact that its vendors were registered and had their own quality systems.
- 2015-07-24TerminatedOther
Medical Device Exchange 12MM VASCULAR; Pouch, Tyvek/Mylar, Sterile, Single Use Only; OEM Catalog number 9-AVP2-012; EXP (repackager), Aga Medical (OEM) Transcatheter septal occluder
EXP did not register as a medical device establishment, list the devices being recalled, or establish its own quality system and instead relied on the fact that its vendors were registered and had their own quality systems.
- 2006-10-20TerminatedOther
AMPLATZER PFO Occluder. Order No. 9-PFO-018. Device is not PMA approved in the US. 18mm Sterile EO. The AMPLATZER PFO Occluder is a self-expandable, double disc device made from a Nitinol wire mesh. AGA Medical Corporation, 682 Mendelssohn Avenue, Golden Valley, MN 55427 USA.
AMPLATZER PFO Occluders were mislabeled with incorrect device sizes. The lot M06B01-58 contains 18mm PFO Occluders but is labeled as containing 25mm devices. Lot M06B01-52 contains 25mm PFO Occluders but is labeled as containing 18mm devices. This device was not distributed within the United States and does not affect U.S. consignees.
- 2006-10-20TerminatedOther
AMPLATZER PFO Occluder. Order No. 9-PFO-025. Device is not PMA approved in the US. 25mm Sterile EO. The AMPLATZER PFO Occluder is a self-expandable, double disc device made from a Nitinol wire mesh. AGA Medical Corporation, 682 Mendelssohn Avenue, Golden Valley, MN 55427 USA.
AMPLATZER PFO Occluders were mislabeled with incorrect device sizes. The lot M06B01-58 contains 18mm PFO Occluders but is labeled as containing 25mm devices. Lot M06B01-52 contains 25mm PFO Occluders but is labeled as containing 18mm devices. The affected product was not distributed within the United States and does not affect U.S. consignees.
- Open, ClassifiedSoftware change control
REF/Catalogue Number GSX0030A, GORE CARDIOFORM SEPTAL OCCLUDER 30 mm Diameter, Catheter Working Length 75 cm, Delivery Profile 10 Fr, STERILE EO, Rx Only, UDI: 00733132631032
Septal Occluders contain the incorrect expiration date. Product is labeled with a 3 yr. expiration date instead of 2 yrs.
- Open, ClassifiedSoftware change control
REF/Catalogue Number GSX0020A, GORE CARDIOFORM SEPTAL OCCLUDER 20 mm Diameter, Catheter Working Length 75 cm, Delivery Profile 10 Fr, STERILE EO, Rx Only, UDI: 00733132631018
Septal Occluders contain the incorrect expiration date. Product is labeled with a 3 yr. expiration date instead of 2 yrs.
- Open, ClassifiedSoftware change control
REF/Catalogue Number GSX0025A, GORE CARDIOFORM SEPTAL OCCLUDER 25 mm Diameter, Catheter Working Length 75 cm, Delivery Profile 10 Fr, STERILE EO, Rx Only, UDI: 00733132631025 *Not distributed within the US
Septal Occluders contain the incorrect expiration date. Product is labeled with a 3 yr. expiration date instead of 2 yrs.
- Open, ClassifiedSoftware change control
REF/Catalogue Number GSXE0020, GORE CARDIOFORM SEPTAL OCCLUDER 20 mm Diameter, Catheter Working Length 75 cm, Delivery Profile 10 Fr, STERILE EO, Rx Only, UDI: 00733132617630 *Not distributed within the US
Septal Occluders contain the incorrect expiration date. Product is labeled with a 3 yr. expiration date instead of 2 yrs.
Monitor transcatheter septal occluder automatically
DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.