Skip to main content

Bayer

FDA adverse event reports naming Bayer as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: BAYER.

Total reports
7,816
Deaths
28
Injuries
7,679
Malfunctions
42

Reports per year

05
06
08
10
11
13
14
15
16
17
18
20
21
23
25

Source: FDA adverse event reports by year received, 2005–2025. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Top device categories

Recent recalls

  • 2026-09-11Open, ClassifiedEmployee error

    MEDRAD Stellant FLEX Syringe Kit used with Stellant FLEX CT Injection System; Catalog number: FLEXS-150-QFT;

    Some devices may not have been properly sterilized due to a manufacturing sterilization issue.

  • 2026-06-03Open, ClassifiedProcess control

    Avanta Multi-Patient Administration Tube Sets (AVA 500 MPAT (Part 2 of 2)) used with the Medrad Avanta Injection System; Catalog number: AVA 500 MPAT (Part 2 of 2); Affected material numbers: 87629007, 60729458, 86566648;

    Supplier changes resulted in a fit issue with the Contrast Flow Control Valve (CFCV) interface on the Avanta Injector System. Multiple customer reports indicate that the stopcock cannot be reliably engaged with the CFCV snap interface. In instances where engagement is achieved through applied force, the connection is unstable, and the stopcock may disengage before or during contrast injection, leading to procedure interruption.

  • 2023-01-06Open, ClassifiedUnder Investigation by firm

    Source Administration Set (SAS) used with the Medrad Intego PET Infusion System, Sterile Catalog number: INT CSS Affected material number: 86566621 Intended to deliver accurate doses of 18F Fluorodeoxyglucose (FDG) or 18F Sodium Fluoride (NaF) radiopharmaceuticals and commonly used flushing solutions to patients during molecular imaging (PET Positron Emission Tomography - nuclear medicine) diagnostic procedures.

    Black specks may be present on the packaged needles

  • 2016-11-11TerminatedProcess control

    PET Infusion Pump

    The Source Administration Sets (SAS) used with the Medrad Intego PET Infusion System do not meet pressure testing specifications in an occlusion situation when the product is more than one year past manufacturing date.

  • 2016-01-11TerminatedNonconforming Material/Component

    Medrad Source Administration Sets, Catalog number INT CSS. For use with MEDRAD Intego PET Infusion pump Systems, in nuclear medicine diagnostic procedures.

    The firm's investigation of the Medrad Intego indicates that the inner diameter of the saline tubing is out of specification and may result in the system failing to position the dose correctly during extraction, potentially leading to a low measurement of the extracted dose or to a higher dose being extracted. The occluded SAS tubing may impair the priming function and prompt an RP prime fail

  • 2015-09-16TerminatedSoftware design

    Angiographic Injector. Medrad Mark 7 Arterion Injection System, which utilizes Software Version SW 005.006_SH; to be used specifically for the purposes of injecting contrast medium and common flushing solutions into humans for angiographic studies.

    Software Version SW 005.006_SH, resulted in the removal of purge enforcement from traditional New-Case, Power Up and Syringe Change use cases while the injector head is in the upright position.

  • 2015-08-29TerminatedDevice Design

    Contour Next Control Solution for Contour Next Blood Glucose Test System. Liquid quality control solution. Plastic bottle, 2.5mL. Indicated for use with fresh fingertip capillary whole blood samples.

    If the control solution bottle is not thoroughly shaken prior to use, the control test result may be higher than the control range and may not be recognized and marked as a control test by the blood glucose meter. Also, the control solution, used as a quality control test in blood glucose meter systems, may read outside the specified range too high or too low.

  • 2015-05-28TerminatedSoftware design

    MEDRAD Stellant CT Injector System with Certegra Workstation; intended for the specific purpose of injecting intravenous contrast media into humans for diagnostic studies in computed tomography (CT) applications.

    The root cause investigation showed that the under-volume hazard can occur when the unit has been configured to prime with contrast, but instead primes with saline due to the configuration data not being sent to the injector head. This may create a non-optimal priming condition that could lead to under-infusion.

  • 2014-05-30TerminatedComponent design/selection

    Medrad Continuum MR Infusion system- non-wireless system The MEDRAD Continuum MR Infusion System is designed for patients who require medications and other fluids during an MR procedure. It is intended to provide infusion therapy directly prior to, during and immediately after the MR procedure functioning while either stationary or mobile. It is not intended to provide long-term patient care outside the MR environment. The system is to be used by trained medical staff, primarily critical care, emergency room and radiology nursing staff.

    There is a potential safety risk associated with potentially damaged locking pins within the bracket of non-wireless versions of Continuum systems shipped prior to March 2008.

  • 2013-11-02TerminatedNonconforming Material/Component

    Bayer brand A1CNow+ MULTI-TEST A1C SYSTEM TEST KITS (PART NUMBER 3021/08842610); Hemoglobin A1c Monitor; Glycosylated Hemoglobin Assay; Product is manufactured and distributed by Bayer HealthCare LLC, Sunnyvale, CA The A1CNow+ test provides quantitative measurement of the percent of glycated hemoglobin (%HbA1c or %A1C) levels in capillary (fingerstick) or venous whole blood samples. The test is for professional use to monitor glycemic control in people with diabetes.

    Test results from these lots may produce a high-bias reading (by up to 20%) that is outside of limits defined for stability and product accuracy when product is stored at room temperature four (4) to five (5) months after release.

Monitor Bayer automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.