Device, Thermal Ablation, Endometrial
FDA adverse event reports and recall history for device, thermal ablation, endometrial (product code MNB), from the public MAUDE/AEMS corpus and the live FDA recall database.
- Total reports
- 6,301
- Deaths
- 22
- Injuries
- 4,204
- Malfunctions
- 1,936
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.
Recent recalls
- 2025-04-16Open, ClassifiedUnder Investigation by firm
Brand Name: Cerene Cryotherapy Device Product Name: DEVICE, THERMAL ABLATION, ENDOMETRIAL Model/Catalog Number: FGS-7000 Product Description: The Cerene Cryotherapy Device is an endometrial ablation device that uses nitrous oxide (N2O) to freeze and ablate the endometrium to reduce future menstrual bleeding. The device is intended for use by healthcare professionals who have received appropriate training and are familiar with the principles, clinical applications, complications, side effects, and hazards commonly associated with endometrial ablation. The Cerene Device is single use, disposable, and provided sterile. The average procedure time is 7 minutes, with 2.5 minutes of active cryoablation. Ablation is achieved throughout the uterine cavity through the use of cryothermic energy. The cryothermic energy is provided by a liquid-to-gas phase change of N2O. During the 2.5-minute treatment cycle, liquid N2O (originating from a small Cylinder located in the device handle) flows through a delivery line and into an inflow line with multiple jets. This liquid N2O is infused into an ultra-thin polyurethane Liner, where it converts into gas. The gaseous N2O is exhausted through the Exhaust Hose exiting the bottom of the handle. Component: N/A
Due to evaporator gasket not sealing which can cause liquid nitrous oxide to drip/leak from the bottom of the device handle during treatment or after treatment during venting of the device
- 2024-05-23Open, ClassifiedDevice Design
Brand Name: Cerene Cryotherapy Device Product Name: DEVICE, THERMAL ABLATION, ENDOMETRIAL Model/Catalog Number: FGS-7000 Product Description: The Cerene Cryotherapy Device is an endometrial ablation device that uses nitrous oxide (N2O) to freeze and ablate the endometrium to reduce future menstrual bleeding. The device is intended for use by healthcare professionals who have received appropriate training and are familiar with the principles, clinical applications, complications, side effects, and hazards commonly associated with endometrial ablation. The Cerene Device is single use, disposable, and provided sterile. The average procedure time is 7 minutes, with 2.5 minutes of active cryoablation. Ablation is achieved throughout the uterine cavity through the use of cryothermic energy. The cryothermic energy is provided by a liquid-to-gas phase change of N2O. During the 2.5-minute treatment cycle, liquid N2O (originating from a small Cylinder located in the device handle) flows through a delivery line and into an inflow line with multiple jets. This liquid N2O is infused into an ultra-thin polyurethane Liner, where it converts into gas. The gaseous N2O is exhausted through the Exhaust Hose exiting the bottom of the handle. Component: N/A
Error code 003 was not listed in certain rows of the Device Instructions for Use Troubleshooting section (i.e., Table 22. Summary of LCD Messages and Error Codes).
- 2014-02-26TerminatedComponent design/selection
NovaSure Impedance Controlled Endometrial Ablation System Radiofrequency Controller, Model Number: 71978-001 Catalog Number: RFC2009 Product Usage: The NovaSure Impedance Controlled Endometrial Ablation System is intended to ablate the endometrial lining of the uterus in premenopausal women with mehorrhagia (excessive bleeding) due to benign causes for whom childbearing is complete.
NovaSure Radiofrequency Control Units may not meet a requirement of the electrical standard, IEC 60601-1,
- 2009-09-22TerminatedProcess design
Boston Scientific HydroThermAblator Endometrial Ablation System- HTA Procedure Set (5/bx) Material/UPN/Catalog Number: M006550161 Boston Scientific HydroThermAblator Endometrial Ablation System-HTA Procedure Set (ea) Material/UPN/Catalog Number: M006550160 Intended to ablate the endometrial lining of the uterus in premenopausal women with menorrhagia due to benign causes for whom childbearing is complete.
Cracked procedure sheaths, incorrect care/use of device, and console malfunctions may result in fluid leaks and loss of a cervical seal that are responsible for burns.
- 2009-09-22TerminatedProcess design
Boston Scientific HydroThermAblator Endometrial Ablation System- HTA System Procedure Set with Tenaculum Stabilizer and Disposable Heater Canister (5/bx) Material/UPN/Catalog Number: M006560201 System-HTA System Procedure Set with Tenaculum Stabilizer and Disposable Heater Canister (ea) Material/UPN/Catalog Number: M006560200 Intended to ablate the endometrial lining of the uterus in premenopausal women with menorrhagia due to benign causes for whom childbearing is complete.
Cracked procedure sheaths, incorrect care/use of device, and console malfunctions may result in fluid leaks and loss of a cervical seal that are responsible for burns.
- 2007-03-31TerminatedOther
HTA Console System, 110 Volt, Product Number: 56000 (HTA system is a software-controlled hysteroscopic thermal endometrial ablation system)
Defective integrated circuit board could result in loss of the system pump and patient injury (hot fluid 90 degree C into uterus) if in use during the recirculation/heating phase of treatment
- 2007-03-31TerminatedOther
HTA Console System, 230 Volt, English, Product Number: 560010 (HTA system is a software-controlled hysteroscopic thermal endometrial ablation system)
Defective integrated circuit board could result in loss of the system pump and patient injury (hot fluid 90 degree C into uterus) if in use during the recirculation/heating phase of treatment
- 2007-03-31TerminatedOther
HTA Console System, 230 Volt, Product Number: 56001 DO- English Demonstration Evaluation Unit (HTA system is a software-controlled hysteroscopic thermal endometrial ablation system)
Defective integrated circuit board could result in loss of the system pump and patient injury (hot fluid 90 degree C into uterus) if in use during the recirculation/heating phase of treatment
- 2007-03-31TerminatedOther
HTA Console System, 230 Volt , French, Product Number: 56001FO (HTA system is a software-controlled hysteroscopic thermal endometrial ablation system)
Defective integrated circuit board could result in loss of the system pump and patient injury (hot fluid 90 degree C into uterus) if in use during the recirculation/heating phase of treatment
- 2007-03-31TerminatedOther
HTA Console System, 230 Volt , Italian, Product Number: 56001 IO (HTA system is a software-controlled hysteroscopic thermal endometrial ablation system)
Defective integrated circuit board could result in loss of the system pump and patient injury (hot fluid 90 degree C into uterus) if in use during the recirculation/heating phase of treatment
Monitor device, thermal ablation, endometrial automatically