LFRClass 2

Glucose Dehydrogenase, Glucose

FDA adverse event reports and recall history for glucose dehydrogenase, glucose (product code LFR), from the public MAUDE corpus and the live FDA recall database.

Total reports
74,380
Deaths
33
Injuries
5,147
Malfunctions
69,144

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2018-05-09TerminatedEquipment maintenance

    Accu-Chek Aviva Plus Test Strip, Blood glucose monitoring system, Catalog nos. 06908217001, 06908268001, 06908349001,06908217001 Product Usage: The Accu-Chek Aviva Plus test strips are for use with the Accu-Chek Aviva Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm.

    The recalling firm identified four US lots of test strips out of specification. Affected strips show increased potential for strip errors prior to applying blood, strips not detected once inserted, or biased results, which may not be detectable. In the case of a biased result, incorrect action could be taken by the user in response to the displayed blood glucose result.

  • 2018-05-09TerminatedEquipment maintenance

    Accu-Chek Performa Test Strip, Blood glucose monitoring system, Catalog no. 07299702001 Product Usage: The Accu-Chek Performa test strips are for use with the Accu-Chek Performa meter to quantitatively measure glucose (sugar) in venous whole blood or fresh capillary whole blood samples from the fingertips.

    The recalling firm has identified two lots of test strips that are out of specification. Affected test strips show an increased potential for: strip errors prior to applying a blood sample; strips not detected once inserted; or a biased result may occur, which may not be detectable. In the case of a biased result, incorrect action could be taken by the user in response to the displayed blood glucose result as compared to the actual blood glucose value.

  • 2018-05-09TerminatedEquipment maintenance

    Accu-Chek Inform II Test Strip, Whole blood glucose test system, Catalog no. 05942861001 Product Usage: The Accu-Chek Inform II test strips are for use with the Accu-Chek Inform II meter to quantitatively measure glucose (sugar) in venous whole blood, arterial whole blood, neonatal heel stick, or fresh capillary whole blood samples drawn from the fingertips as an aid in monitoring the effectiveness of glucose control. The system is not for use in diagnosis or screening of diabetes mellitus, nor for testing neonate cord blood samples.

    The recalling firm has identified two lots of test strips that are out of specification. Affected test strips show an increased potential for: strip errors prior to applying a blood sample; strips not detected once inserted; or a biased result may occur, which may not be detectable. In the case of a biased result, incorrect action could be taken by the user in response to the displayed blood glucose result as compared to the actual blood glucose value.

  • 2016-03-14TerminatedSoftware Design Change

    Accu-Chek Inform II Base Unit as a part of the Accu-Chek Inform II Blood Glucose Monitoring System Accu-Chek inform II Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings. This system should only be use with single-use, auto-disabling lancing devices.

    Accu-Chek Inform II Base Unit might produce physical transmission errors in the form of data loss in the communication between the meter and the Data Management Systems (DMS). The issue can lead to the data loss or in the worst case to an erroneous assignment of the patient data (patient mismatch). The issue will only occur at sites using POTCT1-A communication via USB.

  • 2012-04-02TerminatedSoftware design

    OneTouch Verio IQ Blood Glucose Monitoring System Kit; One Touch Verio IQ Blood Glucose IQ Blood Glucose Monitoring Starter Kit; One Touch Verio IQ Blood Glucose Monitoring Warranty Kit. Distributed by : LifeScan, Inc. Milpitas, CA 95035 for: LifeScan Europe Division of Cilag GmBH International 6300 Zug Switzerland Meter made in China. Intended to be used for the quantitative measurement of glucose in fresh capillary whole blood.

    An error was discovered in the OneTouch IQ meter software so that the meter turns itself off when a user attempts to view results in the "Results Log" when the log has 256 or a multiple of 256 items to display.

  • 2011-09-19TerminatedPackaging

    Contour, Contour TS Blood Glucose Test Strips and Bonus Packs Intended for self-testing by people with diabetes and healthcare professionals to monitor glucose concentrations in whole blood

    All 10- or 25- count glucose test strip vials, either loose vials or found in patient starter kits of Contour, Contour TS and Contour" USB blood glucose meters are being recalled a packaging issue which, under certain conditions, can cause some test strips packaged in some small-count vials to operate outside of US FDA-cleared performance specifications and produce a low-bias reading during blood

  • 2009-08-11TerminatedOther

    Roche ACCU-Chek Active Meter Kit, Roche Diagnostics, Indianapolis, IN; REF 03184501001 includes the Roche ACCU-CHEK Softclix Lancing Device, Distributed by Roche Diagnostics, Indianapolis, IN. Model 3144844001 (10 lancets). The device is used to prick the finger and to obtain a drop of blood in preparation for diabetes/insulin level testing.

    The lancet cap may be missing, which could result in an unintended lancet stick to the user.

  • 2009-08-11TerminatedOther

    Roche ACCU-Chek Voicemate Meter, Roche Diagnostics, Indianapolis, IN; REF 12030802001 includes the Roche ACCU-CHEK Softclix Lancing Device, Distributed by Roche Diagnostics, Indianapolis, IN. Model 3144844001 (10 lancets). The device is used to prick the finger and to obtain a drop of blood in preparation for diabetes/insulin level testing.

    The lancet cap may be missing, which could result in an unintended lancet stick to the user.

  • 2009-08-11TerminatedOther

    Roche ACCU-Chek Aviva Meter Kit, Roche Diagnostics, Indianapolis, IN; REF 04893247001 includes the Roche ACCU-CHEK Softclix Lancing Device, Distributed by Roche Diagnostics, Indianapolis, IN. Model 3144844001 (10 lancets). The device is used to prick the finger and to obtain a drop of blood in preparation for diabetes/insulin level testing.

    The lancet cap may be missing, which could result in an unintended lancet stick to the user.

  • 2009-08-11TerminatedOther

    Roche ACCU-CHEK Compact Plus Meter Kit, Roche Diagnostics, Indianapolis, IN; REFS 03149137001, 05177294001 and 05079241001 include the Roche ACCU-CHEK Softclix Lancing Device, Distributed by Roche Diagnostics, Indianapolis, IN. Model 3583031002 (17 lancets). The device is used to prick the finger and to obtain a drop of blood in preparation for diabetes/insulin level testing.

    The lancet cap may be missing, which could result in an unintended lancet stick to the user.