Skip to main content
FMKClass 2

Single Use Only Blood Lancet With An Integral Sharps Injury Prevention Feature

FDA adverse event reports and recall history for single use only blood lancet with an integral sharps injury prevention feature (product code FMK), from the public MAUDE/AEMS corpus and the live FDA recall database.

Total reports
14,101
Deaths
2
Injuries
830
Malfunctions
13,184

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 20122026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.

Recent recalls

  • 2013-09-11TerminatedComponent change control

    BD Microtainer Quikheel Lancet 1.0 mm incision Depth 2.5 mm Incision Length REF 368101 Rx Only Becton, Dickinson and Company, Franklin Lakes, NJ 07417 USA Made in USA. Used to perform heel sticks for the purpose of collecting skin puncture blood.

    The Lancet will not retract after activation thus creating a potential for exposure to a used sharp.

  • 2013-08-28TerminatedEquipment maintenance

    ACCU-CHECK FastClix Lancing Device. It is intended for use in conjunction with a blood glucose monitoring system.

    Internal Roche inspections have revealed that after the lancet is used for a finger stick, the needle may not fully retract back within the lancet drum or back into the cap. This allows the needle to protrude inside the cap or outside of the cap of the ACCU-CHEK FastClix lancing device.

  • 2013-07-13TerminatedProcess control

    Safety Lancets - Odyssey, Button Activated; Tri- Bevel Tip for Virtually Painless Sampling; Needle automatically retracts, No lancing device required. 26G - Single Use Only, Made In Korea Distributed by Hummingbird Medical, Savannah, GA 100 count Boxes; 40 boxes per case Blood collection by penetrating skin with needle.

    Lancets were not sterilized within the dose range established by sterilization validation.

  • 2009-08-11TerminatedOther

    Roche ACCU-CHEK Softclix Lancing Device, Distributed by Roche Diagnostics, Indianapolis, IN. Models 3144844001 (10 lancets) and 3583031002 (17 lancets). The device is used to prick the finger and to obtain a drop of blood in preparation for diabetes/insulin level testing.

    The lancet cap may be missing, which could result in an unintended lancet stick to the user.

  • 2008-10-27TerminatedError in labeling

    Premier Value Lancing Device. UPC Code: 840986016707, The Package Includes: 1 Lancing Device, 1 Clear Cap and 5 Premier Value Ultra Thin Lancets, Distributed by: Chain Drug Consortium, LLC, Boca Raton, FL 33431. Lancing Device intended to be used by diabetes patients that require a drop of capillary blood for testing.

    The Alternate Site Cap was not included in the package as indicated on the labeling.

  • 2008-10-14TerminatedProcess control

    Safety Lancet for capillary blood sampling, 1.6 mm x 28G, Mini, Sarstedt. (Blood lancet)

    The product from this lot may not retract the lancet into the safe position inside the lancet body after the trigger is actuated.

  • 2007-10-18TerminatedProcess design

    Microtainer Safety Flow Lancets; Green Sterile, Use Once and Discard; 1.4 mm Blade Length; 1.0mm Blade Width; Becton Dickinson Vacutainer Systems; Becton Dickinson & Company; Franklin Lakes, NJ 07417

    Potential for the blade and holder assembly to become loose and then separate from the main body of the lancet during the process of disposal.

  • 2007-08-16TerminatedNonconforming Material/Component

    One Touch SureSoft Lancing Device in carton. Part Number 021-140-01, aka OT SureSoft Hospital, Gentle 200

    Lancet tip may not fully retract and cause needle stick and exposure to used lancets.

  • 2007-06-09TerminatedOther

    NeatNick Sweeping Action Heel Lancet -Full-Term- Item Number: 1030085 (box of 100) and Item Number: 1030086, case of 1,000 (10 boxes of 100)

    Lancet may misfire and an inadequate blood sample may result; additionally, the blade may not retract after use and result in a needle stick to the user/patient

  • 2007-06-09TerminatedOther

    NeatNick Sweeping Action Heel Lancet - Preemie- Item Number: 1030083 (box of 100) and Item Number: 1030084, case of 1,000 (10 boxes of 100)

    Lancet may misfire and an inadequate blood sample may result; additionally, the blade may not retract after use and result in a needle stick to the user/patient

Monitor single use only blood lancet with an integral sharps injury prevention feature automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.