Skip to main content

Wright Medical Technology Inc

FDA adverse event reports naming Wright Medical Technology Inc as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: WRIGHT MEDICAL TECHNOLOGY INC.

Total reports
4,751
Deaths
1
Injuries
4,476
Malfunctions
274

Reports per year

96
97
98
05
10
14
19
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 1996–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Top device categories

Recent recalls

  • 2026-10-01Open, ClassifiedProcess control

    CHARLOTTE CLAW Screw, REF: 40122716, intended for use with CLAW II Polyaxial Compression Plating System

    Screws, intended for use with polyaxial compression plating system intended for foot osteotomies, was manufactured with missing threads on the screwheads, which if used would prevent a secure rigid engagement with the plate's internal threads, which may lead to surgery prolongation to retrieve another device, or alternative surgery method if no replacement device is readily available.

  • 2024-04-04Open, ClassifiedProcess change control

    stryker Infinity Resection Adjustment Block, REF 33600030, ankle arthroplasty

    Stryker has identified an issue that impacts specific lots of Infinity" Resection Guide Adjustment Blocks. The parts within these three lots were found to have been missing an internal screw in the finished instrument assembly. The missing screw is used to lockout the Medial/Lateral adjustment on the INFINITY Resection Adjustment Guide.

  • 2023-07-07Open, ClassifiedProcess control

    Stryker Infinity Alignment Frame Distal Sub-Assembly, Catalog #33600020, non-sterile.

    The devices were manufactured with the pin sleeve holes too narrow, which does not allow the pin sleeves to pass through the pin sleeve holes, thereby preventing pin placement through the pin holes.

  • 2023-04-26Open, ClassifiedEmployee error

    stryker ORTHOLOC 3Di Fusion Plate MTP Right, REF 58A220RT, bone stabilization and fixation

    Two lots of Ortholoc 3Di Fusion plates contain incorrect dorsiflexion descriptions laser marked on the parts.

  • 2023-04-26Open, ClassifiedEmployee error

    stryker ORTHOLOC 3Di Fusion Plate MTP Right, REF 58A221RT, bone stabilization and fixation

    Two lots of Ortholoc 3Di Fusion plates contain incorrect dorsiflexion descriptions laser marked on the parts.

  • 2023-02-24Open, ClassifiedEmployee error

    ORTHOLOC 3DSi CLAW II 4 HOLE Plate, Size 25MM, REF 402S0425, Bone fixation device

    Stryker has identified an issue that impacts two specific lots of CLAW" II plates. The plates within these two lots were found to have been comingled, resulting in the incorrect product in packages. Packaging for 402S0425 incorrectly contained 40241425, and vice versa.

  • 2023-02-24Open, ClassifiedEmployee error

    ORTHOLOC 3DSi CLAW II 4 HOLE Plate, Size 25MM, REF 40241425, Bone fixation device

    Stryker has identified an issue that impacts two specific lots of CLAW" II plates. The plates within these two lots were found to have been comingled, resulting in the incorrect product in packages. Packaging for 402S0425 incorrectly contained 40241425, and vice versa.

  • 2022-10-27TerminatedDevice Design

    Stryker INBONE Tibial Tray, Left Plasma Spray Ti6Al4V, CPti, Use w/ 16mm Stem Base, REF 200252903, Size 3

    The tibial tray lock detail is oversized (larger than specification).

  • 2022-10-27TerminatedDevice Design

    Stryker INBONE Tibial Tray, Left Plasma Spray Ti6Al4V, CPti, Use w/ 18mm Stem Base, REF 200252904, Size 4

    The tibial tray lock detail is oversized (larger than specification).

  • 2022-10-04Open, ClassifiedUnder Investigation by firm

    stryker EVOLVE, Proline Stem, CoCr, Size: +2, O: 705mm, REF 496S275, Wright s EVOLVE Proline Plus Radial Head and Repair System is a set configuration designed to address radial head fractures. It combines parts of two systems: EVOLVE Proline and EVOLVE TRIAD.

    The outer label does not match the device inside the package. The affected lot of EVOLVE Proline Stem 7.5mm +2 contains a size 9.5mm +4 stem, while the affected lot of EVOLVE Proline Stem 9.5mm +4 contains a size 7.5mm +2 stem.

Monitor Wright Medical Technology Inc automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.