FDA adverse event reports and recall history for staple, fixation, bone (product code JDR), from the public MAUDE corpus and the live FDA recall database.

Total reports
2,692
Deaths
0
Injuries
1,385
Malfunctions
1,296

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2026-09-17Open, ClassifiedNo Marketing Application

    Reign Medical Clench Staple System (Compression), Description/REF: 9mm x 7mm/152-30907; 11mm x 10mm/152-31110; 15mm x 12mm/152-31512; 18mm x 18mm/152-31818; 20mm x 20mm/152-32020; 9mm x 7mm/152-40907; 11mm x 10mm/152-41110; 15mm x 12mm/152-41512; 18mm x 18mm/152-41818; 20mm x 20mm/152-42020

    Bone fixation staple requires a new 510(k); the performance characteristics of this device have not been adequately established.

  • 2026-06-08Open, ClassifiedProcess control

    NeoSpan SuperElastic Compression Staple w/Instruments Catalog Number and Product Name T50 SN010 NeoSpan SuperElastic Compression Staple w/Instruments, Size 10x10x10 UDI-DI Code: 00817906020083 T50 SN012 NeoSpan SuperElastic Compression Staple w/Instruments, Size 12x12x12 UDI-DI Code: 00817906020090 T50 SN020 NeoSpan SuperElastic Compression Staple w/Instruments, Size 20x20x20 UDI-DI Code: 00817906020151 T50 SN110 NeoSpan SuperElastic Compression Staple w/Instruments, Size 10x15x13 UDI-DI Code: 00817906020113 T50 SN115 NeoSpan SuperElastic Compression Staple w/Instruments, Size 15Wx15x15 UDI-DI Code: 00817906020137 T50 SN118 NeoSpan SuperElastic Compression Staple w/Instruments, Size 18x14x14 UDI-DI Code: 00810021861364 Single/multiple component metallic bone fixation appliances and accessories The In2Bones USA NeoSpan" Compression Staple Implant w/instruments is indicated for hand and foot bone fragments, osteotomy fixation and joint arthrodesis.

    Due to the Magnetic Resonance Conditional Symbol missing on the carton, inner tray, and patient label.

  • 2022-10-18Open, ClassifiedError in labeling

    SPEEDSHIFT20x20x20 Offset 10mm Implant Drill Kit, Product No. SE-2020-10. Used with DePuy Synthes SPEEDSHIFT Continuous Compression Staple Implants for bone fixation and management of fracture and reconstructive surgery.

    The subject products are being recalled because they contain the incorrect size drill guide and will not prepare the bone correctly for the size of the implant included in the sterile kit. If the incorrect drill guide is used, there is potential for bone damage due to incorrectly placed drill holes for the implant included in the kit.

  • 2022-10-18Open, ClassifiedError in labeling

    SPEED 15X12X12mm Continuous Compression Implant, Product No. SE-1512

    The subject products are being recalled because they contain the incorrect size drill guide and will not prepare the bone correctly for the size of the implant included in the sterile kit. If the incorrect drill guide is used, there is potential for bone damage due to incorrectly placed drill holes for the implant included in the kit.

  • 2022-10-18Open, ClassifiedError in labeling

    SPEED 9X10X10mm Continuous Compression Implant, Product No. SE-0910

    The subject products are being recalled because they contain the incorrect size drill guide and will not prepare the bone correctly for the size of the implant included in the sterile kit. If the incorrect drill guide is used, there is potential for bone damage due to incorrectly placed drill holes for the implant included in the kit.

  • 2022-10-18Open, ClassifiedError in labeling

    DePuy Synthes Hammertoe Continuous Compression Implants, Product No. 46.239.001

    The subject products are being recalled because they contain the incorrect size drill guide and will not prepare the bone correctly for the size of the implant included in the sterile kit. If the incorrect drill guide is used, there is potential for bone damage due to incorrectly placed drill holes for the implant included in the kit.

  • 2022-10-06Open, ClassifiedEmployee error

    Hammertoe Continuous Compression Implant (CCI) Kit Part Number: 46.239.001

    Drill Pin (drill bit) is not compatible with the Drill Template included in the kit due to a packaging error, result in a surgical delay and may cause soft tissue damage if a larger incision is required to accommodate an alternative form of fixation. Additionally, there is a potential for adverse tissue reaction if the surgeon attempts to use the guide regardless of tight fit and plastic debris is created and ends up in the wound

  • 2017-03-20TerminatedLabeling mix-ups

    MotoCLIP(TM) Super Elastic Fusion System, Size 18mm x 14mm x 14mm, REF 1442-1814, Rx Only, STERILE R.

    Seven 18mm x 18mm x 18 mm implants had been placed in kits, packed and labeled as 18mm x 14mm x 14mm.

  • 2015-04-02TerminatedEmployee error

    ToggleLoc; fastener, fixation, non degradable, soft tissue; Artificial Ligament Fixation Device #7 PE Ziploop Extended Toggleloc Product Usage: The Toggleloc System is a non-resorbable system intended to aid in arthroscopic and orthopedic reconstructive procedures requiring soft tissue fixation, due to injury or degenerative disease.

    The firm is recalling ToggleLocs and Ziploops manufactured in May and July of 2014 as they were not processed following a standard work process, which resulted in complaints of sterile packaging not being sealed.

  • 2015-04-02TerminatedEmployee error

    Ziploop Button Product Usage: The Ziploop Button is a non-resorbable system intended to aid in ACL reconstructive procedure due to injury or degenerative disease

    The firm is recalling ToggleLocs and Ziploops manufactured in May and July of 2014 as they were not processed following a standard work process, which resulted in complaints of sterile packaging not being sealed.