Prosthesis, Elbow, Hemi-, Radial, Polymer
FDA adverse event reports and recall history for prosthesis, elbow, hemi-, radial, polymer (product code KWI), from the public MAUDE/AEMS corpus and the live FDA recall database.
- Total reports
- 1,792
- Deaths
- 3
- Injuries
- 1,556
- Malfunctions
- 210
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.
Recent recalls
- 2022-10-04Open, ClassifiedUnder Investigation by firm
stryker EVOLVE, Proline Stem, CoCr, Size: +2, O: 705mm, REF 496S275, Wright s EVOLVE Proline Plus Radial Head and Repair System is a set configuration designed to address radial head fractures. It combines parts of two systems: EVOLVE Proline and EVOLVE TRIAD.
The outer label does not match the device inside the package. The affected lot of EVOLVE Proline Stem 7.5mm +2 contains a size 9.5mm +4 stem, while the affected lot of EVOLVE Proline Stem 9.5mm +4 contains a size 7.5mm +2 stem.
- 2022-10-04Open, ClassifiedUnder Investigation by firm
stryker EVOLVE, Proline Stem, CoCr, Size: +4, OD: 9.5mm, REF 496S275, Wright s EVOLVE Proline Plus Radial Head and Repair System is a set configuration designed to address radial head fractures. It combines parts of two systems: EVOLVE Proline and EVOLVE TRIAD.
The outer label does not match the device inside the package. The affected lot of EVOLVE Proline Stem 7.5mm +2 contains a size 9.5mm +4 stem, while the affected lot of EVOLVE Proline Stem 9.5mm +4 contains a size 7.5mm +2 stem.
- 2014-12-12TerminatedDevice Design
Align Radial Stems, Prosthesis, Elbow, Hemi Replacement of the radial head for degenerative or post-traumatic disabilities presenting pain, crepitation, and decreased motion at the radio-humeral and/or proximal radio-ulnar joint with: Joint destruction and/or subluxation and resistance to conservative treatment. Intended use is also for primary replacement after fracture of the radial head and symptomatic sequelae after radial head resection.
Report received where the Align Radial Stem fractured.
- 2014-05-13TerminatedProcess control
ExploR 7x26mm Modular Radial Stem, Catalog number 11-210062. Elbow prosthesis orthopedic implant.
This lot of the ExploR Modular Radial Stem may be missing the wedge ramp threads. If the wedge ramp threads are missing from the stem, the likelihood of the set screw loosening from the radial stem intraoperatively is increased. If the set screw loosens, a revision surgery may be necessary.
- 2013-10-29TerminatedPackaging change control
Small Bone Innovations (SBi) rHead Radial Stem, rHead Recon Radial System, rHead Radial Stem Lateral Assembly, rHead Radial Stem 6mm Collar, rHead Recon Radial Stem 6mm Collar, rHead Radial Stem Lateral Assembly 6mm Collar, rHead Standard Extended Stem, rHead Radial Stem Plasma Coated, rHead Recon Radial Stem Plasma Coated, rHead Radial Stem 6mm Collar Plasma Coated, rHead Recon Radial Stem 6mm Collar Plasma Coated, uHead Ulnar Stem Standard, uHead Ulnar Stem 20mm Collar packaged in a sterile pouch inside a shrink-wrapped box. The recall was expanded on December 20, 2013 to include the following devices: rHead lateral assembly, radial head implant, sizes 2 through 4, Uni-elbow lateral assembly, radial implant assembly, sozes 2 through 4, ReMotion Left, large distal radial component, x-small, small, medium, and large, ReMotion Right, distal radial component, x-small, small, medium, and large.
The recall was initiated due to an increased risk of implants breaching the sterile pouches inside the shrink-wrapped box.
- 2013-05-03TerminatedDevice Design
Synthes Radial Head Prosthesis: Trial Radial Head Elbow joint prosthesis: Synthes Radial Head Prosthesis System is a two-piece modular system comprised of titanium alloy stem and cobalt chrome components with an integral screw and side-loading application to allow in-situ assembly.
The Trial Head may come loose from the implant stem during manipulation of the arm during surgery. To date, no adverse events have been reported related to this issue and this recall is not being initiated as a result of adverse events.
- TerminatedPackaging change control
ExploR Radial Head Replacement System. PN: 11-210061 LN: 243610. The stem is placed inside a polyurethane bag. The bagged component will then be placed in an inner PETG tray. The screw will be placed inside a separate polybag, and placed in the PETG tray with foam pieces added to the top and bottom of the tray for protection. The inner tray and the contents within will then have a Tyvek lid heat sealed to the tray. The sealed inner tray will be fit into an outer PETG tray and a Tyvek lid heat sealed to the tray. The entire sealed unit is placed in a SBS carton that will be shrink wrapped for protection.
Zimmer Biomet is conducting a medical device recall for the ExploR 6x24mm Implantable Stem with Screw due to a complaint alleging that the Modular Radial HD lock screw was not included in the package for the stem.
- Open, ClassifiedLabeling Change Control
Radial Head Replacement System Implant is a one-piece, stemmed radial head replacement. The radial head is available in diameters of 19, 22 and 25mm. The radial stems range from 5.5 to 8.5mm in diameter and from 21 24mm in length.
DePuy Synthes Radial Head Replacement System Contraindications added to Electronic Instructions for Use (e-IFU): Absolute contraindications include: Infection, sepsis, osteomyelitis.
- TerminatedProcess control
WRIGHT EVOLVE Proline Head, Size: +2, OD: 20mm, REF 496H220
Two lots of EVOLVE Modular Radial Head could have the wrong size product in the package. Both the label and laser marking are incorrect on the affected product.
- TerminatedProcess control
WRIGHT EVOLVE Proline Head, Size: +2, OD: 22mm, REF 496H222
Two lots of EVOLVE Modular Radial Head could have the wrong size product in the package. Both the label and laser marking are incorrect on the affected product.
Monitor prosthesis, elbow, hemi-, radial, polymer automatically
DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.