Skip to main content

Globus Medical, Inc.

FDA adverse event reports naming Globus Medical, Inc. as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: GLOBUS MEDICAL, INC..

Total reports
2,402
Deaths
7
Injuries
769
Malfunctions
1,603

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Top device categories

Recent recalls

  • 2025-04-23Open, ClassifiedNonconforming Material/Component

    MAGEC 2 Rod, 5.0mm 90mm Standard; Version/Model: MC2-5090S;

    Devices for the affected lot were not assembled according to product specifications. The distraction rod outer diameter was assembled with a 5.5mm rod versus the required 5.0mm.

  • 2024-02-07Open, ClassifiedDevice Design

    ExcelsiusGPS Cannula A, 9mm x 250mm, Cannula used with ball tip probes, Model No. 6143.2844

    Devices may not be rendered sterile due to an internal gap that may be present on some devices causing increased difficulty cleaning prior to steam sterilization. There is an increased risk for infection, inflammation, local toxicity, systemic toxicity, sensitivity, and cross-contamination.

  • 2024-02-06Open, ClassifiedPackage design/selection

    Globus Medical Cranial Drill with Stop, 2.40mm - Instruments for access, preparation, and placement of cranial devices with the ExcelsiusGPS. Ref : 6143.1834S

    Sterile-packed Cranial Drills may not be sterile due to a breach in the packaging and/or may have discoloration near the drill tip.

  • 2024-02-06Open, ClassifiedPackage design/selection

    Globus Medical Cranial Drill with Stop, 3.20mm - instruments for access, preparation, and placement of cranial devices with the ExcelsiusGPS. Ref: 6143.1842S

    Sterile-packed Cranial Drills may not be sterile due to a breach in the packaging and/or may have discoloration near the drill tip.

  • 2024-02-06Open, ClassifiedPackage design/selection

    Globus Medical Cranial Drill, 6.00mm - instruments for access, preparation, and placement of cranial devices with the ExcelsiusGPS. Ref: 6143.1870S

    Sterile-packed Cranial Drills may not be sterile due to a breach in the packaging and/or may have discoloration near the drill tip.

  • 2024-01-18Open, ClassifiedDevice Design

    ExcelsiusGPS Flat Panel Fluoroscopy Fixture Kit, Part number 6203.2100

    A calibration error associated with the Flat Panel Fluoroscopy Fixture when used with ExcelsiusGPS for fluoroscopy workflow may lead to inaccurate navigation, which may result in device misplacement.

  • 2020-08-22TerminatedNonconforming Material/Component

    ALTERA Spacer ALTERA Spacer, 10x31, 8-12mm, 8 (sterile)

    Specific lots of ALTERA implants have internal components that may have been manufactured using stainless steel rather than the specified cobalt chromium molybdenum alloy.

  • 2020-08-22TerminatedNonconforming Material/Component

    ALTERA Spacer ALTERA Spacer, 10x26, 8-12mm, 8

    Specific lots of ALTERA implants have internal components that may have been manufactured using stainless steel rather than the specified cobalt chromium molybdenum alloy.

  • 2020-08-22TerminatedNonconforming Material/Component

    ALTERA Spacer ALTERA Spacer, 10x26, 9-13mm, 8

    Specific lots of ALTERA implants have internal components that may have been manufactured using stainless steel rather than the specified cobalt chromium molybdenum alloy.

  • 2020-08-22TerminatedNonconforming Material/Component

    ALTERA Spacer ALTERA Spacer, 10x26, 10-14mm, 8

    Specific lots of ALTERA implants have internal components that may have been manufactured using stainless steel rather than the specified cobalt chromium molybdenum alloy.

Monitor Globus Medical, Inc. automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.