GE Medical Systems, LLC

FDA adverse event reports naming GE Medical Systems, LLC as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: GE MEDICAL SYSTEMS, LLC.

Total reports
1,398
Deaths
11
Injuries
659
Malfunctions
656

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Top device categories

Recent recalls

  • 2026-07-13Open, ClassifiedProcess control

    Optima XR200amx, Mobile Digital-Ready Radiographic System

    GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.

  • 2026-07-13Open, ClassifiedProcess control

    Optima XR220amx, Mobile Digital-Ready Radiographic System

    GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.

  • 2026-07-13Open, ClassifiedProcess control

    Optima XR240amx, Mobile Digital Radiographic System

    GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.

  • 2026-06-17Open, ClassifiedComponent design/selection

    Giraffe OmniBed, a combination of an infant incubator and an infant warmer, All Models

    GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury.

  • 2026-06-17Open, ClassifiedComponent design/selection

    Giraffe OmniBed Carestations and Giraffe OmniBed Carestation CS1, a combination of an infant incubator and an infant warmer, Model Numbers 2082844-001-XXX

    GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury.

  • 2026-06-17Open, ClassifiedComponent design/selection

    Giraffe OmniBed Field Replaceable Unit (FRU), Model Numbers 6600-1056-400 and 6600-1461-500

    GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury.

  • 2026-06-05Open, ClassifiedDevice Design

    Allia IGS 3 Pulse angiographic X-ray system

    GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.

  • 2026-06-05Open, ClassifiedDevice Design

    Allia IGS 5 Pulse angiographic X-ray system

    GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.

  • 2026-06-05Open, ClassifiedDevice Design

    Allia IGS 7 Pulse angiographic X-ray system

    GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.

  • 2026-06-05Open, ClassifiedDevice Design

    Allia IGS Pulse angiographic X-ray system

    GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.