Skip to main content

BD

FDA adverse event reports naming BD as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: BD.

Total reports
2,083
Deaths
11
Injuries
760
Malfunctions
1,124

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Top device categories

Recent recalls

  • 2026-09-15Open, ClassifiedDevice Design

    BD Alaris Pump Infusion Sets 10561554* 10885403232565 50885403232563 BD Alaris Pump Infusion Set, 15 Micron Filter, Back Check Valve, 3 SmartSite Y-sites 10015861A 10885403232626 50885403232624 BD Alaris Pump Infusion Set, SmartSite Bag Access Non-Vented, PE Lined Tubing, Back Check Valve, 3 SmartSite Y-sites 11171447* 7613203021234 37613203021242 BD Alaris Pump Infusion Set, 2 Back Check Valves, 3 SmartSite Y-sites 11426965* 10885403232558 50885403232556 BD Alaris Pump Infusion Set, Back Check Valve, 5 SmartSite Y-sites 2260-0500 7613203012591 37613203012608 BD Alaris Pump Infusion Set, PE Lined Tubing 24600-0007 10885403241970 50885403241978 BD Alaris Pump Infusion Set, PE Lined Tubing, Back Check Valve, 3 SmartSite Y-sites 2278-0500 10885403223266 50885403223264 BD Alaris Pump Infusion Blood Set, 200 Micron Filter 2477-0007 7613203019460 37613203019478 BD Alaris Pump Infusion Blood Set, 180 Micron Filter, SmartSite Y-site 22603-B007T 10885403240942 50885403240940 BD Alaris Pump Infusion Set, SmartSite Bag Access Non-Vented, PE Lined Tubing, Bonded Texium Closed Male Luer with Priming Cap 24010-0007T 10885403271021 50885403271029 BD Alaris Pump Infusion Set, Bonded Texium, Closed Male Luer with Priming Cap, Back Check Valve, 3 SmartSite Y-sites 24301-0007T 10885403223198 50885403223196 BD Alaris Pump Infusion Set, SmartSite Bag Access Non-Vented, 0.2 Micron Filter, PE Lined Tubing, Bonded Texium Closed Male Luer with Priming Cap, SmartSite Y-site 24601-B007T 10885403240959 50885403240957 BD Alaris Pump Infusion Set, SmartSite Bag Access Non-Vented, PE Lined Tubing, Bonded Texium Closed Male Luer with Priming Cap, Back Check Valve, 2 SmartSite Y- sites 22602-B007T 10885403240935 50885403240933 BD Alaris Pump Infusion Set, SmartSite Bag Access Non-Vented, PE Lined Tubing, Bonded Texium Closed Male Luer with Priming Cap 11426964 10885403232367 50885403232365 BD AlarisLVP 20D LOW SORB 2204-0007* 10885403199363 50885403199361 BD Alaris Pump Infusion Set 10015862 10885403221866 50885403221864 BD Alaris Pump Infusion Set, SmartSite Bag Access Non-Vented, PE Lined Tubing 10010453 7613203021135 37613203021143 BD Alaris Pump Infusion Set, 1.2 Micron Filter, PE Lined Tubing, SmartSite Y-site 11532269 10885403232343 50885403232341 BD Alaris Pump Infusion Set, 0.2 Micron Filter, Back Check Valve, PE Lined Tubing, 2 SmartSite Y-sites 2202-0007* 10885403274039 50885403274037 BD Alaris Pump Infusion Set, 1.2 Micron Filter 2432-0007* 10885403232329 50885403232327 BD Alaris Pump Infusion Set, 0.2 Micron Filter, Back Check Valve, 3 SmartSite Y-sites 10010454 7613203015806 37613203015814 BD Alaris Pump Infusion Set, 0.2 Micron Filter, PE Lined Tubing, SmartSite Y-site 2465-0007 10885403221941 50885403221949 BD Alaris Pump Infusion Set, SmartSite Bag Access Non-Vented, 0.2 Micron Filter, PE Lined Tubing, SmartSite Y-site 10015012 10885403233951 50885403233959 BD Alaris Pump Infusion Burette Set, 0.2 Micron Filter, Smallbore Tubing, SmartSite Port (Burette), SmartSite Y-site 2441-0007 10885403235047 50885403235045 BD Alaris Pump Infusion Burette Set, Smallbore Tubing, SmartSite Port (Burette), 3 SmartSite Y-sites 11419365 10885403234736 50885403234734 BD Alaris Pump Infusion Set, Back Check Valve, Manifold, 3-Way Stopcock, 5 SmartSite Y-sites 2423-0007 7613203020985 37613203020979 BD Alaris Pump Infusion Set, Back Check Valve, 2 Ganged 4-Way Stopcocks, 4 SmartSite Y-sites 10813621 10885403234514 50885403234512 BD Alaris Pump Infusion Set, Back Check Valve, 3 Ganged 4-Way Stopcocks, 4 SmartSite Y-sites 10010483 10885403232602 50885403232600 BD Alaris Pump Infusion Set, Vented Syringe Adapter, Smallbore Tubing 10013072 7613203021210 37613203021228 BD Alaris Pump Infusion Set, PE Lined Tubing, Back Check Valve, 2 SmartSite Y-sites 10013361 10885403221965 50885403221963 BD Alaris Pump Infusion Set, SmartSite Bag Access Non-Vented, Back Check Valve, 2 SmartSite Y-sites 10013890 10885403232268 50885403232266 BD Alaris Pump Infusion Set, PE Lined Tubin

    Due to discrepancies in specific attributes in the performance data for infusion sets, which may lead to inappropriate pump performance.

  • 2026-08-13Open, ClassifiedProcess control

    60 In Non-DEHP Minibore Extension Set RES ME2010 UDI-DI code: 10885403236235 MaxPlus Ext Set w/Needleless and Y-Site REF MPX5305-C UDI-DI code: 10885403274619 9 IN Extension Set w/Max Y REF MPX5300-C UDI-DI code: 10885403237218 The Intravascular Administration Set and Extension Set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into the patient's artery or vein. The lntravascular Administration Set and Extension Set may incorporate componentry that aid in the prevention of accidental needle sticks.

    Due to mechanical issue with internal sterilization equipment resulting in potential underdosing of product during processing, which may have resulted in products not being sterilized correctly.

  • 2026-07-02Open, ClassifiedUnder Investigation by firm

    MEDLINE 111 GRAVITY SET W/3 Needle Free Y-Sites REF DYNDTN0555 Catalog Number: SAM2461 UDI code: N/A Engineering Sample/Not for Human Use

    Due to engineering sample labeled "Not for Human Use" being inadvertently shipped it the U.S. market.

  • 2026-06-18Open, ClassifiedDevice Design

    BD Connecta" BD Luer-Lok" 360 REF 394910 UDI-DI code: 00382903949106 BD Connecta" Stopcocks are used to administer fluids or medication through one or two different ports via an IV cannula or extension tube. The ports can be used to sample blood or for hemodynamic monitoring. When used with IV lipid nutritional products, the stopcock device can be used for up to 24 hours.

    Due to stopcock issues when using the plastic luer-lok syringe, the connection at the Port C may over thread or continue to spin rather than tightening securely. This behavior has led some clinical users to perceive the connection as unstable.

  • 2026-03-27Open, ClassifiedSoftware Manufacturing/Software Deployment

    BD Kiestra" ReadA; Catalog No.: 446948.

    In certain situations, following a system reboot, one or more modules may have become unreachable via the remote connectivity interface. When this occurred, there was potential for delays in plate retrieval from the incubator while connectivity was restored. This issue was intermittent and did not affect test results, performance of the assay, or sample integrity.

  • 2025-08-27Open, ClassifiedError in labeling

    BD PhaSeal Injector Luer (N30C), REF: 515001 BD PhaSeal Injector Luer Lock (N35), REF: 515003 BD PhaSeal Injector Luer Lock (N35C), REF: 515004 BD PhaSeal Injector Lu63er Lock (N35C Multi), REF: 515005 BD PhaSeal Protector (P14), REF: 515100 BD PhaSeal Protector (P21), REF: 515102 BD PhaSeal Protector (P21 Multi), REF: 515103 BD PhaSeal Protector (P28), REF: 515104 BD PhaSeal Protector (P50), REF: 515105 BD PhaSeal Protector (P50 Multi), REF: 515106 BD PhaSeal Protector (P53), REF: 515107 BD PhaSeal Protector (P55), REF: 515117 BD PhaSeal Connector Luer Lock (C35), REF: 515200 BD PhaSeal Connector Luer Lock (C45), REF: 515202 BD PhaSeal Y-site Connector (C80), REF: 515304 BD PhaSeal L Connector (C90), REF: 515305 BD PhaSeal Infusion Adapter (C100), REF: 515306 BD PhaSeal Infusion Adapter (C100 Multi), REF: 515307 BD PhaSeal Cap for Injector (M50), REF: 515404 BD PhaSeal Connector Cap (M70), REF: 515406 BD PhaSeal Optima Injector (N35-O), REF: 515052 BD PhaSeal Optima Injector (N35-O Multi), REF: 515053 BD PhaSeal Injector Locking (N40-O), REF: 515056 BD PhaSeal Injector Locking (N40-O Multi), REF: 515057 BD PhaSeal Optima Protector (P13-O), REF: 515060 BD PhaSeal Optima Protector (P13-O Multi), REF: 515062 BD PhaSeal Optima Protector (P28-O), REF: 515063 BD PhaSeal Optima Protector (P20-O), REF: 515064 BD PhaSeal Optima Protector (P20-O Multi), REF: 515065 BD PhaSeal Optima Connector (C35-O), REF: 515070 BD PhaSeal Optima Infusion Adapter (C100-O), 515078 BD PhaSeal Optima Infusion Adapter (C100-O Multi), REF: 515079

    Closed system drug transfer devices were shipped to the U.S. market without a U.S. version of the instructions for use, which could be an issue if a healthcare professional is prompted to review it.

  • 2025-08-01Open, ClassifiedLabeling design

    BD Trucount Controls, 30 Test - REF: 664343, RxOnly, IVD

    Control kits contain incorrect standard deviation (SD) values for the low and high control beads which compromises the accurate calculation of bead counts.

  • 2025-08-01Open, ClassifiedLabeling design

    BD Trucount Controls, 30 Test - REF: 340335

    Control kits contain incorrect standard deviation (SD) values for the low and high control beads which compromises the accurate calculation of bead counts.

  • 2025-07-03Open, ClassifiedProcess control

    REF: 335775 , BD FACS, 7-Color Setup Beads, 25 setups per kit, Rx Only, IVD. Used in flow cytometry for fluorescence compensation.

    Due to customer complaints, errors in spectral overlap factors table for 7-color setup beads may affect accurate setup of the flow cytometry instrument prior to analyzing samples.

  • 2025-07-03Open, ClassifiedProcess control

    REF: 666289 , BD FACS, 7-Color Setup Beads, 25 Runs, CE, IVD. Used in flow cytometry for fluorescence compensation.

    Due to customer complaints, errors in spectral overlap factors table for 7-color setup beads may affect accurate setup of the flow cytometry instrument prior to analyzing samples.

Monitor BD automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.