Skip to main content
GESClass 1

Blade, Scalpel

FDA adverse event reports and recall history for blade, scalpel (product code GES), from the public MAUDE corpus and the live FDA recall database.

Total reports
994
Deaths
0
Injuries
238
Malfunctions
697

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2022-11-23Open, ClassifiedUnder Investigation by firm

    Southmedic SAFETY CARTRIDGE WITH SOUTHMEDIC PLUS SURGICAL BLADE, REF 73-8020, STERILE EO

    Upon review of its batch record, Southmedic identified that lot # W92761 was not sent out for sterilization, and was instead transferred directly to distribution where it was sold and distributed to customers as non-sterile product labelled as sterile.

  • 2016-09-08TerminatedMixed-up of materials/components

    Intermediate carton label reads in part as: "Bard-Parker***Stainless Steel Surgical Blades***#12B***QTY 150 (3 x 50)***REF 371712***LOT 0088464." Intended for tissue separation and other procedures that require a sharp surgical blade to puncture or cut.

    After receiving a complaint the firm disclosed that Conventional Stainless Steel Surgical Blade catalog number 371712 lot 0088464 was incorrectly identified in the single unit foil pack as 12 instead of 12B.

  • 2012-11-05TerminatedProcess control

    "***Sharpoint IQ Geometry Slit Knife Angled, Bevel Up***Surgical Specialties Corporation Reading, PA***REF 72-2740***" Ophthalmic surgery

    Potential for bent tips.

  • 2012-11-05TerminatedProcess control

    "***Sharpoint IQ Geometry Slit Knife Angled, Bevel Up***Surgical Specialties Corporation Reading, PA***REF 72-2840***" Ophthalmic surgery

    Potential for bent tips.

  • 2012-11-05TerminatedProcess control

    "***Sharpoint IQ Geometry Slit Knife Angled, Bevel Up***Surgical Specialties Corporation Reading, PA***REF 72-3040***" Ophthalmic surgery

    Potential for bent tips.

  • 2012-11-05TerminatedProcess control

    "***Sharpoint IQ Geometry Slit Knife Angled, Bevel Up***Surgical Specialties Corporation Reading, PA***REF 72-3240***" Ophthalmic surgery

    Potential for bent tips.

  • 2009-12-14TerminatedOther

    Merit Medical Systems, Inc. Futura Safety Scalpels, Futura Safety Scalpel REF/CAT No SMS210 Used for cutting tissue in surgical and laboratory procedures.

    Potential sterility breach caused by excessive shipment damage.

  • 2009-02-20TerminatedProcess control

    Stainless Steel Surgical Blades Part no 4-312B. The product is supplied sterile individually packaged in boxes containing 100 blades.

    Blade may break: Over milling of the blade resulted in tip more narrow than specified.

  • 2007-10-02TerminatedPackaging change control

    Shelf Pack: Bard-Parker Rib-Back Carbon Steel no.11, surgical blade. Unit Package: Contents of unopened or undamaged package guaranteed sterile; After use place product in a puncture resistant container and dispose properly. Becton Dickinson Becton Dickinson Acute Care Franklin Lakes, NJ 07417

    Complaints received of rust on the blades of BD Bard-Parker Surgical blades.

  • 2007-10-02TerminatedPackaging change control

    Shelf Pack: Bard-Parker Rib-Back Stainless Steel no.11, surgical blade. Unit Package: Contents of unopened or undamaged package guaranteed sterile; After use place product in a puncture resistant container and dispose properly. Becton Dickinson Becton Dickinson Acute Care Franklin Lakes, NJ 07417

    Complaints received of rust on the blades of BD Bard-Parker Surgical blades.

Monitor blade, scalpel automatically