Skip to main content
FMGClass 2

Stopcock, I.V. Set

FDA adverse event reports and recall history for stopcock, i.v. set (product code FMG), from the public MAUDE corpus and the live FDA recall database.

Total reports
6,139
Deaths
25
Injuries
365
Malfunctions
5,699

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2026-09-11Open, ClassifiedProcess control

    IV Non-Dedicated Gravity Administration Sets with drip chambers

    Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall that does not enter the fluid path, which if seen by a healthcare provider may lead to replacement of product before use, which could result in a delay in therapy.

  • 2026-09-11Open, ClassifiedProcess control

    IV Sets with drip chambers

    Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall. The presence of discoloration and particulate matter in the fluid path has not been ruled out.

  • 2026-06-18Open, ClassifiedDevice Design

    BD Connecta" BD Luer-Lok" 360 REF 394910 UDI-DI code: 00382903949106 BD Connecta" Stopcocks are used to administer fluids or medication through one or two different ports via an IV cannula or extension tube. The ports can be used to sample blood or for hemodynamic monitoring. When used with IV lipid nutritional products, the stopcock device can be used for up to 24 hours.

    Due to stopcock issues when using the plastic luer-lok syringe, the connection at the Port C may over thread or continue to spin rather than tightening securely. This behavior has led some clinical users to perceive the connection as unstable.

  • 2025-11-26Open, ClassifiedProcess control

    Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/15 DROPS/ML, 108 IN; Catalog Number: 490189. 2. IV ADMIN SET 137 IN.; Catalog Number: 490226.

    Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

  • 2025-11-26Open, ClassifiedProcess control

    Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/15 DROPS/ML, 132 IN; Catalog Number: 490469. 2. IV ADMIN SET, 60D 2 CARESITE VLV, 132 IN; Catalog Number: 490470.

    Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

  • 2025-11-26Open, ClassifiedProcess control

    Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV AD. SET,15 DROPS/ML, W/EXT SET; Catalog Number: 490467.

    Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

  • 2025-11-26Open, ClassifiedProcess control

    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/ 10 DROPS/ML, 134 in.; Catalog Number: 490357.

    Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

  • 2025-11-26Open, ClassifiedProcess control

    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers: 1. IV ADMIN SET, 10 DROPS/ML, 144 IN; Catalog Number: 490385. 2. IV ADMIN SET, 3 CARESITES, 120 IN.; Catalog Number: 490401.

    Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

  • 2025-11-26Open, ClassifiedProcess control

    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. NF1270A SAFELINE ANESTHESIA IV SET; Anesthesia IV Administration Set with 3 CARESITE Injection Sites and Manifold; Catalog Number: NF1270A.

    Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

  • 2025-11-26Open, ClassifiedProcess control

    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. US1295 ULTRASITE IV ADMIN SET 60DROP; Catalog Number: 352394.

    Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

Monitor stopcock, i.v. set automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.