Alere SAN Diego, Inc
FDA adverse event reports naming Alere SAN Diego, Inc as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: ALERE SAN DIEGO, INC.
- Total reports
- 1,590
- Deaths
- 6
- Injuries
- 239
- Malfunctions
- 1,328
Reports per year
Source: FDA adverse event reports by year received, 2010–2024. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Top device categories
- Test, Time, ProthrombinGJS1,347
- Visual, Pregnancy Hcg, Prescription UseJHI119
- Immunoassay Method, Troponin SubunitMMI74
- Test, Natriuretic PeptideNBC27
- Antigens, All Groups, Streptococcus Spp.GTY12
- Fibrinogen And Fibrin Split Products, Antigen, Antiserum, ControlDAP4
- Kit, Test, Pregnancy, Hcg, Over The CounterLCX2
- Enzymatic Esterase--Oxidase, CholesterolCHH1
- Powered Laser Surgical InstrumentGEX1
- Fibrin Split ProductsGHH1
Recent recalls
- 2026-09-25Open, ClassifiedDevice Design
McKesson CONSULT hCG Urine Cassette, Model/ Part: FHC-A102/5001; Alere hCG Cassette (20 mIU/mL) in Urine, Model/ Part: FHC-A102/92210; Fisher Scientific Sure-Vue Serum/UrinehCG-STAT, Model/ Part: FHC-A202/23900531, FHC-A202/23900530; McKesson CONSULT Diagnostics hCG Urine Cassette, Model/ Part: FHC-A202/5002; Alere hCG Combo Cassette (20/10 mIU/mL) for Urine or Serum, Model/ Part: FHC-A202/92215; Cardinal Health hCG Combo Rapid Test Serum or Urine, Model/ Part: FHC-A202/B-1077-23; Medline hCG Combo Pregnancy Test Cassette, Model/ Part: FHC-A202/MPH22025HCG;
hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result.
- 2026-09-25Open, ClassifiedDevice Design
Fisher Scientific Sure-Vue Serum/Urine hCG, Model/ Part: FHC-202/23900529, FHC-202/23900528; Beckman Coulter Inc's ICON 25 hCG, Model/ Part: FHC-202/43025; Henry Schein One Step+ hCG Combo test in Serum or Urine, Model/ Part: FHC-202/9004075; Alere hCG Combo Cassette (25mIU/mL) in Urine (CLIA Waived) or Serum, Model/ Part: FHC-202/92225; Pro Advantage By NDC Urine/Serum Pregnancy Cassette Device, Model/ Part: FHC-202/P080021
hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result.
- 2026-09-25Open, ClassifiedDevice Design
Fisher Scientific Sure-Vue Urine hCG Strip, Model/ Part: FHC-101/23900527; McKesson CONSULT hCG Urine Test Dipstick CLIA Waived, Model/ Part:FHC-101/5000; Henry Schein One Step+ hCG Urine Strip Test, Model/ Part: FHC-101/9004072; Alere hCG Dipstick (25mIU/mL) in Urine, Model/ Part: FHC-101/92211; Cardinal Health hCG Dipstick Rapid Test CLIA Waived, Model/ Part: FHC-101/B-1077-21; Medline hCG Pregnancy Test Strip, Model/ Part: FHC-101/MPH11025HCG; Pro Advantage by NDC Urine Pregnancy Test Strip, Model/ Part: FHC-101/P080023;
hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result.
- 2026-09-25Open, ClassifiedDevice Design
Fisher Scientific Sure-Vue Urine hCG, Model/ Part: FHC-102/23900526; Henry Schein, Inc. One Step + Urine Cassette Test, Model/ Part: FHC-102/9004074, FHC-102/9004073; Alere hCG Cassette (25mIU/mL) in Urine, Model/ Part: FHC-102/92217; Cardinal Health hCG Cassette Rapid test CLIA Waived, Model/ Part: FHC-102/B1077-22; Medline HCG Pregnancy Test Cassette, Model/ Part: FHC-102/MPH12025HCG; Pro Advantage By NDC Urine Pregnancy Cassette Device, Model/ Part: FHC-102/P080063;
hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result.
- 2024-04-30Open, ClassifiedLabeling design
Abbott Cholestech LDX Battery Kit, Model: 3LP656194-PCM-LD, PN: 55170
There is a potential that the Cholestech LDX battery (an external battery pack available as an optional accessory to the Cholestech LDX Analyzer) may swell which could cause minor burns, cuts or abrasions, or low current electrical shock to the user.
- 2016-11-16TerminatedEmployee error
Alere Triage Total 5 Control Level 1, Model No. 88753 Description of the product: The Triage Total Controls 5 are quality control materials that are used to assist the laboratory in monitoring test performance throughout the measurable range. They are not calibrators and are not used to calibrate the Triage tests. The results of quality control testing do not impact direct patient care and should not influence clinical decision making process the physician uses to make clinical diagnosis for the patient. Indications for Use: The Alere Triage Total 5 Controls are assayed materials to be used with the Alere Triage Cardiac Panel, Alere Triage CardioProfiler Panel, Alere Triage Profiler SOB Panel, Alere Triage BNP Test, Alere Triage D-Dimer Test, and Alere Triage Meters to assist the laboratory in monitoring test performance.
Encoded with incorrect ranges that are not consistent with the ranges printed on the Expected Value Card.
- 2016-08-19TerminatedOther
Alere INRatio PT/INR System Professional Model Numbers: 0100004 Product Usage: Alere INRatio 2 PT/INR Monitoring System (Professional Use): The Alere INRatio 2 PT/INR Monitoring System (Professional Use), consisting of the INRatio 2 Monitor and INRatio 2 PT/INR test strip, is used for quantitative determination of international normalized ratio (INR) in fresh capillary whole blood to monitor the effect of warfarin on clotting time by health care professionals. The Alere INRatio 2 PT/INR Monitoring System (Professional Use) is intended for use outside of the body (in vitro diagnostic use). The Alere INRatio 2 PT/INR Monitoring System (Professional Use) is not intended to be used for screening purposes.
Alere San Diego is initiating a voluntary recall for the Alere INRatio/INRatio2 PT/INR Monitoring System (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio/INRatio2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.
- 2016-08-19TerminatedOther
INRatio Prothrombin Time (PT) Monitoring System Model Number: 0100007 Product Usage: The INRatio Prothrombin Time Monitoring System uses a modified version of the one-stage PT test. After a drop of blood is applied to the test strip, it is drawn into the test area and mixed with reagents that cause coagulation to begin. The test strip area on the test strip is separated into three channels, each of which contains electrodes for detection of the blood clot. One channel contains the reagents to perform the Prothrombin time test on the blood sample. The other two channels contain the reagents. to perfom the Prothrombin time test on the blood sample. The other two channels contain the reagents to run control tests. No external quality control test are required for the INRatio System. The meter monitors the reactions, and calculates the PT and INR for the blood sample, which are reported on the display. If the control results are not within a set range, it indicates a problem with the test and an error message is reported on the display instead of a result.
Alere San Diego is initiating a voluntary recall for the Alere INRatio/INRatio2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio/INRatio2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.
- 2016-08-19TerminatedOther
Alere INRatio Replacement Monitor Model Number: 0100137 Usage: The INRatio 2 PT Monitoring system is used for the quantitative measurement of Prothrombin Time (PT) in fresh, capillary whole blood. The INRatio 2 PT Monitoring system is intended for use outside the body (in vitro diagnostic use) by people taking warfarin and other oral anticoagulant (blood thinning) therapy who need to monitor the clotting time of their blood. The INRatio 2 PT Monitoring System is not intended to be used for screening purpose.
Alere San Diego is initiating a voluntary recall for the Alere INRatio/INRatio2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio/INRatio2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.
- 2016-08-19TerminatedOther
Alere INRatio2 PT/INR Professional Testing System Model Number: 0200431 Product Usage: The INRatio 2 PT Monitoring system is used for the quantitative measurement of Prothrombin Time (PT) in fresh, capillary whole blood. The INRatio 2 PT Monitoring system is intended for use outside the body (in vitro diagnostic use) by people taking warfarin and other oral anticoagulant (blood thinning) therapy who need to monitor the clotting time of their blood. The INRatio 2 PT Monitoring System is not intended to be used for screening purpose.
Alere San Diego is initiating a voluntary recall for the Alere INRatio/INRatio2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio/INRatio2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.
Monitor Alere SAN Diego, Inc automatically