Fibrinogen And Fibrin Split Products, Antigen, Antiserum, Control
FDA adverse event reports and recall history for fibrinogen and fibrin split products, antigen, antiserum, control (product code DAP), from the public MAUDE/AEMS corpus and the live FDA recall database.
- Total reports
- 225
- Deaths
- 3
- Injuries
- 12
- Malfunctions
- 196
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.
Recent recalls
- 2026-03-19Open, ClassifiedProcess change control
Brand Name: STA Liatest D-Di Product Name: STA Liatest D-Di Model/Catalog Number: REF 00515 Product Description: in vitro diagnostic. The STA - Liatest D-Di kit is an assay for the quantitative determination of D-dimer in venous plasma for use on STA-R, STA Compact and STA Satellite analyzers by professional laboratory personnel.
After receiving customer complaints, investigations confirmed the presence of a positive bias in D-Dimer measurements throughout the entire analytical range.
- 2025-12-01Open, ClassifiedNo Marketing Application
D-Dimer Test
Distribution without premarket approval/clearance.
- 2025-10-16Open, ClassifiedUnder Investigation by firm
Product: STA Liatest D-Di; REF: 00515;
A potential risk of underestimation of D-Dimer (D-Di) levels with the two listed lots.
- 2016-01-07TerminatedLabeling design
Siemens Acute Care DDMR TestPak Catalog Number: CDDMRE SMN: 10701511 in vitro diagnostic
Stratus CS Cartridge Barcode labels may not scan on the Stratus CS systems
- 2014-08-14TerminatedUnder Investigation by firm
VIDAS D-Dimer Exclusion II, an automated quantitative test for use on instruments of the VIDAS family for the immunoenzymatic determination of fibrin degradation products in human plasma using the Enzyme Linked Fluorescent Assay technique. Also indicated for use in conjunction with clinical pretest probability assessment model to exclude DVT and PE disease.
Customer complaints about not repeatable results below the detection limit (<45 ng/mL). Therefore false negatives are being reported for results around the medical decision point of 500 ng/mL that initially tested at <45 ng/mL.
- 2012-08-28TerminatedNonconforming Material/Component
Siemens Stratus(R) CS Acute Care(TM) D-dimer Calibrator (DDM CalPak) lots 502058002 and 501318002 The Stratus(R) CS Acute Care(TM) D-dimer Calibrator (DDMR CalPak), Catalog No. CDDMR-C is an in-vitro diagnostic product intended to be used for the calibration of the Stratus(R) CS Acute Care(TM) D-dimer )DDMR) method
Siemens has confirmed that the Stratus(R) CS Acute Care(TM) D-Dimer CalPak calibrator lots 502058002 and 501318002 may exhibit SLOPE or CALCV error messages (instrument errors), which indicate calibration failures.
- 2012-07-06TerminatedProcess design
Alere Triage Profiler SOB Panel PN 97300 Used for the determination of creatine kinase MB, myoglobin, troponin I, B-type natriuretic peptide, and cross-linked fibrin degradation products containing D-dimer in EDTA anticoagulated whole blood and plasma specimens.
Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.
- 2006-05-19TerminatedOther
Stratus CS STAT Fluorometric Analyzer DDMR TestPak (CDDMR) in vitro diagnostic.
Foil sealing of the TestPak may occlude the pipet tip resulting in insufficient aspiration of fluids. Erroneous test results may be generated without an associated error code. Test results may be falsely elevated or depressed, and the magnitude of the inaccuracy may vary based on the degree of occlusion.
- 2006-05-19TerminatedOther
Stratus CS STAT Fluorometric Analyzer DDMR DilPak (CDDMR-D) in vitro diagnostic.
Foil sealing of the DilPak may occlude the pipet tip resulting in insufficient aspiration of fluids. Erroneous test results may be generated without an associated error code. Test results may be falsely elevated or depressed, and the magnitude of the inaccuracy may vary based on the degree of occlusion.
- 2005-11-18TerminatedOther
Advanced D-Dimer used with the Sysmex CA-1500, CA-7000, CA-500 Systems when using the AKIMA method.
Reports of false negative results on Sysmex Coagulation System.
Monitor fibrinogen and fibrin split products, antigen, antiserum, control automatically