Skip to main content

Volcano Corporation

FDA adverse event reports naming Volcano Corporation as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: VOLCANO CORPORATION.

Total reports
1,383
Deaths
13
Injuries
229
Malfunctions
1,069

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Top device categories

Recent recalls

  • 2019-12-12TerminatedEmployee error

    Philips Volcano FFR software used in the following systems connected to the McKesson, Schwarzer, and some Philips Xper Hemodynamic Systems (HMS): (1) Volcano Imaging System s5i, Part #807400001; (2) s5ix Imaging System with s5iz PC Workstation Refurbished, Part#807400001 R and 807401200; (3) s5i Imaging System, Part#806300003; (4) s5i Imaging System, Part#5322466; (5) s5x Imaging System, Part#807300001; (6) s5x Imaging System Refurbished, Pat#807300001 R; (7) s5 Imaging System Refurbished, Part#804200001 R; (8) CORE Mobile Imaging System (120V), Part #400-0100.01; (9) CORE Mobile Imaging System Refurbished, Part #400-0100.01-R; (10) CORE Mobile Imaging System (240V), Part #400-0100.07; (11) CORE Mobile Imaging System Refurbished, Part #400-0100.07-R; (12) CORE Mobile Imaging System (100V), Part #400-0100.08; (13) CORE Mobile Imaging System Refurbished, Part #400-0100.08-R; and (14) CORE Imaging System, Part #400-0100.02; (15) CORE Integrated Imaging System Refurbished, Part#400 0100.02 R.

    Interoperability issue that affects certain systems that use the recalled software when it is connected to specific Hemodynamic Systems (HMS).

  • 2018-05-17TerminatedOther

    Philips Volcano CORE M2 Vascular System, Part # 400-0100.17, Rx Only. Used for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vessels of the peripheral vasculature.

    The incorrect resistor has been used in affected CORE M2 Imaging Systems, which causes noise and results in an overly bright IVUS Image.

  • 2017-04-10TerminatedUnder Investigation by firm

    Refinity Rotational Intravascular Ultrasound (IVUS) Catheter; Product number: 400-0200.289; Catalog number 89800. Cardiovascular: The REFINITY (or REFINITY ST) Rotational IVUS Catheter (REFINITY or REFINITY ST) is intended for the intravascular ultrasound examination of coronary arteries. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal interventional procedures. REFINITY (or REFINITY ST) catheters are designed for use as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures.

    High complaint rates of kinking near the junction of the proximal shaft to telescope section of the catheter. Kinking could result in disruption of normal clinical workflow or loss of imaging.

  • 2017-04-10TerminatedUnder Investigation by firm

    Refinity Short Tip Rotational Intravascular Ultrasound (IVUS) Catheter; Product number: 400-0200.273; Catalog number 89900. Cardiovascular: The REFINITY (or REFINITY ST) Rotational IVUS Catheter (REFINITY or REFINITY ST) is intended for the intravascular ultrasound examination of coronary arteries. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal interventional procedures. REFINITY (or REFINITY ST) catheters are designed for use as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures.

    High complaint rates of kinking near the junction of the proximal shaft to telescope section of the catheter. Kinking could result in disruption of normal clinical workflow or loss of imaging.

  • 2016-03-03TerminatedLabeling False and Misleading

    Volcano R100 Sterile Equipment Cover; Sterile; General Hospital: These Sterile Equipment covers are provided to cover equipment that is mean to be used in the sterile surgical field but cannot be sterilized. The sterile drapes are put over this equipment in case any sterile items may come into contact with them.

    Volcano Corporation has identified that certain lot numbers of Volcano R-100 Sterile Equipment Covers may have the incorrect expiry date on the outer shipping box; individually wrapped inner product bears the correct expiry date.

  • 2016-03-03TerminatedLabeling False and Misleading

    Volcano SpinVision Sterile Equip cover II; Sterile; General Hospital: These Sterile Equipment covers are provided to cover equipment that is mean to be used in the sterile surgical field but cannot be sterilized. The sterile drapes are put over this equipment in case any sterile items may come into contact with them.

    Volcano Corporation has identified that certain lot numbers of Volcano R-100 Sterile Equipment Covers may have the incorrect expiry date on the outer shipping box; individually wrapped inner product bears the correct expiry date.

  • 2015-09-02TerminatedUnder Investigation by firm

    Volcano Pioneer Plus Re-Entry Catheter; Product Code PPLUS20, Catheter for Crossing Total Occlusions. Cardiovascular: The Volcano Pioneer Plus Intravascular Ultrasound Re-Entry Catheter (Pioneer Plus) is intended to facilitate the placement and positioning of catheters within the peripheral vasculature.

    Due to a manufacturing defect, there is the possibility that a small wire could extend through the catheter shaft of the Pioneer Plus intravascular Ultrasound Guided Re-Entry Catheters.

  • 2015-07-20TerminatedSoftware design

    Volcano s5/s5i/CORE Intravascular Ultrasound Systems with software versions 3.3 and 3.4 (with and without iFR Scout Technology) and v.3.4 software kits; Model s5. Radiology: The Volcano s5/s5i and CORE Series Intravascular Imaging and Pressure System is used for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vessels of the peripheral vasculature

    A software issue was discovered where an inaccurate FFR/iFR value could be calculated under certain circumstances.

  • 2015-07-20TerminatedSoftware design

    Volcano s5/s5i/CORE Intravascular Ultrasound Systems software versions 3.3 and 3.4 (with and without iFR Scout Technology) and v.3.4 software kits; Model s5x. Radiology: The Volcano s5/s5i and CORE Series Intravascular Imaging and Pressure System is used for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vessels of the peripheral vasculature

    A software issue was discovered where an inaccurate FFR/iFR value could be calculated under certain circumstances.

  • 2015-07-20TerminatedSoftware design

    Volcano s5/s5i/CORE Intravascular Ultrasound Systems software versions 3.3 and 3.4 (with and without iFR Scout Technology) and v.3.4 software kits; Model CORE Mobile. Radiology: The Volcano s5/s5i and CORE Series Intravascular Imaging and Pressure System is used for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vessels of the peripheral vasculature

    A software issue was discovered where an inaccurate FFR/iFR value could be calculated under certain circumstances.

Monitor Volcano Corporation automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.