System, Imaging, Pulsed Echo, Ultrasonic

FDA adverse event reports and recall history for system, imaging, pulsed echo, ultrasonic (product code IYO), from the public MAUDE/AEMS corpus and the live FDA recall database.

Total reports
6,496
Deaths
17
Injuries
239
Malfunctions
6,189

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 20122026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.

Recent recalls

  • 2026-04-01Open, ClassifiedSoftware change control

    eTRAX Needle System Starter Kit 12G (for Aurora Trackers), Part Number 667-149

    There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.

  • 2026-04-01Open, ClassifiedSoftware change control

    eTRAX Needle System Starter Kit 14G (for Aurora Trackers), Part Number 667-150

    There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.

  • 2026-04-01Open, ClassifiedSoftware change control

    eTRAX Needle System Starter Kit 16G(for Aurora Trackers), Part Number 667-151

    There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.

  • 2026-04-01Open, ClassifiedSoftware change control

    eTRAX Needle System Starter Kit 18G (for Aurora Trackers), Part Number 667-152

    There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.

  • 2026-04-01Open, ClassifiedSoftware change control

    eTRAX Needle Sensor - 12G (for Aurora Trackers), Part Number 667-156

    There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.

  • 2026-04-01Open, ClassifiedSoftware change control

    eTRAX Needle Sensor - 14G(for Aurora Trackers), Part Number 667-157

    There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.

  • 2026-04-01Open, ClassifiedSoftware change control

    eTRAX Needle Sensor - 16G(for Aurora Trackers), Part Number 667-158

    There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.

  • 2026-04-01Open, ClassifiedSoftware change control

    eTRAX Needle Sensor - 18G (for Aurora Trackers), Part Number 667-159

    There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.

  • 2022-01-24Open, ClassifiedDevice Design

    GE Vscan Extend, being sold as "Vscan Extend Console"

    The firm identified that the Vscan Extend device will not power on because of depleted battery, due to the device not being turned off prior to battery removal or when being connected to a computer without AC power.

  • 2022-01-24Open, ClassifiedDevice Design

    GE Vscan Extend, being sold as "Vscan Extend Dual DICOM"

    The firm identified that the Vscan Extend device will not power on because of depleted battery, due to the device not being turned off prior to battery removal or when being connected to a computer without AC power.

Monitor system, imaging, pulsed echo, ultrasonic automatically