Microport Orthopedics Inc.
FDA adverse event reports naming Microport Orthopedics Inc. as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: MICROPORT ORTHOPEDICS INC..
- Total reports
- 13,014
- Deaths
- 14
- Injuries
- 12,844
- Malfunctions
- 110
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Top device categories
- Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component)KWA3,409
- Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component)JDL1,957
- Prosthesis, Hip, Semi-Constrained, Metal/Polymer, UncementedLWJ1,359
- Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, UncementedLZO1,354
- Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/PolymerJWH1,296
- Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous UncementedLPH920
- Prosthesis, Knee, Femorotibial, Semi-Constrained, Cemented, Metal/PolymerHRY862
- Prosthesis, Hip, Semi-Constrained, Metal/Polymer, CementedJDI485
- Prosthesis, Hip, Femoral, ResurfacingKXA242
- Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic/Metal, Cemented Or UncementedMRA217
Recent recalls
- 2026-01-20Open, ClassifiedNo Marketing Application
Brand Name: MicroPort Product Name: HIPTURN FEM HEAD TRIAL 36MM REUSABLE Model/Catalog Number: HTHT0036 Software Version: N/A Product Description: HIPTURN FEM HEAD TRIAL 36MM REUSABLE
Due to products not having FDA Premarket authorization to be distributed within the United States.
- 2024-12-18Open, ClassifiedLabeling Change Control
EVOLUTION MP FEMORAL COMPONENT CS/CR POROUS SIZE 5 RIGHT PRIMARY, REF: EFSRP5PR; EVOLUTION MP FEMORAL COMPONENT CS/CR NON-POROUS SIZE 6 LEFT PRIMARY, REF: EFSRN6PL Knee prosthesis component
Femoral implant porous size 5 right primary and femoral implant non-porous size 6 left primary have correct markings on devices, but may have incorrect packaging.
- 2024-07-01Open, ClassifiedUnder Investigation by firm
EVOLUTION MP Femoral CS/CR Non-Porous Non-porous MicroPort total knee replacement implants are for cemented use only.
One confirmed incident has been received that EFSRN4PR, lot number MP1982170, Size 4 Evolution MP Non-Porous Femoral Right was opened during surgery and contained part number EFSRN3PL, lot number MP1982097, Size 3 Evolution MP Non-Porous Femoral Left in the packaging.
- 2022-11-22TerminatedNonconforming Material/Component
EVOLUTION MP Tibial Bases, Size 2+ LEFT, REF ETPKN2PL, knee tibial base
One confirmed incident has been received that ETPKN2PL lot 1916559, size 2 Evolution MP Tibial Base, was opened during surgery and contained ETPKN7SL lot 1916715, size 7 Evolution MP Tibial Base, in the packaging.
- 2022-11-22TerminatedNonconforming Material/Component
EVOLUTION MP Tibial Bases, Size 7 LEFT, REF ETPKN7SL, knee tibial base
One confirmed incident has been received that ETPKN2PL lot 1916559, size 2 Evolution MP Tibial Base, was opened during surgery and contained ETPKN7SL lot 1916715, size 7 Evolution MP Tibial Base, in the packaging.
- 2020-09-18TerminatedDevice Design
PROFEMUR Neck Long Titanium, Catalog Numbers: PHA01204, PHA01214, PHA01224, PHA01234, PHA01244, PHA01254, PHA01264. Hip prosthesis component.
MicroPort Orthopedics Inc. states that there have been reports of fractures of the long and extra-long Titanium modular femoral neck component after implantation. Worldwide there were 680 Long and X-Long PROFEMUR Titanium modular neck fractures among 123,840 global sales representing is a cumulative rate of modular neck device fracture of 0.55% from 2002 to 2019. Within the United States there were 349 Long and X-Long PROFEMUR Titanium modular neck fractures among 15,786 United States sales representing a cumulative rate of modular neck device fracture with a fracture rate in the United States of 2.21% from 2002 to 2019. There is an average time of 5.4 years to device fracture after implantation.
- 2020-09-18TerminatedDevice Design
PROFEMUR Neck Extra Long Titanium, Catalog Numbers: PHA01206, PHA01236, PHA01256. Hip prosthesis component
MicroPort Orthopedics Inc. states that there have been reports of fractures of the long and extra-long Titanium modular femoral neck component after implantation. Worldwide there were 680 Long and X-Long PROFEMUR Titanium modular neck fractures among 123,840 global sales representing is a cumulative rate of modular neck device fracture of 0.55% from 2002 to 2019. Within the United States there were 349 Long and X-Long PROFEMUR Titanium modular neck fractures among 15,786 United States sales representing a cumulative rate of modular neck device fracture with a fracture rate in the United States of 2.21% from 2002 to 2019. There is an average time of 5.4 years to device fracture after implantation.
- 2020-09-09Open, ClassifiedLabeling Change Control
PROFEMUR Titanium and Cobalt Chrome modular necks, Item Numbers and descriptions: Item Number Description Description 2 PHAC1202 PROFEMUR NECK NEUTRAL SHORT COBALT CHROME PHAC1204 PROFEMUR NECK NEUTRAL LONG COBALT CHROME PHAC1212 PROFEMUR NECK A/R VAR/VAL 2 SHORT COBALT CHROME PHAC1214 PROFEMUR NECK A/R VAR/VAL 2 LONG COBALT CHROME PHAC1222 PROFEMUR NECK A/R VAR/VAL 1 SHORT COBALT CHROME PHAC1224 PROFEMUR NECK A/R VAR/VAL 1 LONG COBALT CHROME PHAC1232 PROFEMUR NECK A/R 8DG SHORT COBALT CHROME PHAC1234 PROFEMUR NECK A/R 8DG LONG COBALT CHROME PHAC1242 PROFEMUR NECK A/R 15DG SHORT COBALT CHROME PHAC1244 PROFEMUR NECK A/R 15DG LONG COBALT CHROME PHAC1252 PROFEMUR NECK VAR/VAL 8DG SHORT COBALT CHROME PHA01202 PROFEMUR NECK NEUTRALSHORT SHORT PHA01204 PROFEMUR NECK NEUTRAL LONG PHA01206 PROFEMUR NECK NEUTRAL X LONG PHA01212 PROFEMUR NECK A/R VAR/VAL 2 SHORT PHA01214 PROFEMUR NECK A/R VAR/VAL 2 LONG PHA01222 PROFEMUR NECK A/R VAR/VAL 1 SHORT PHA01224 PROFEMUR NECK A/R VAR/VAL 1 LONG PHA01232 PROFEMUR NECK 8DG A/R SHORT PHA01234 PROFEMUR NECK 8DG A/R LONG PHA01236 PROFEMUR NECK 8DG A/R X LONG PHA01242 PROFEMUR NECK 15DG A/R SHORT PHA01244 PROFEMUR NECK 15DG A/R LONG PHA01252 PROFEMUR NECK 8DG VAR/VAL SHORT PHA01254 PROFEMUR NECK 8DG VAR/VAL LONG PHA01256 PROFEMUR NECK 8DG VAR/VAL X LONG PHA01262 PROFEMUR NECK 15DG VAR/VAL SHORT PHA01264 PROFEMUR NECK 15DG VAR/VAL LONG Intended Use: MPO Total Hip Systems are indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients.
MicroPort Orthopedics Inc. is voluntarily recalling any existing inventory of PROFEMUR Titanium and Cobalt Chrome modular necks with previous package insert versions to replace the PROFEMUR Hip System Package Insert with the most recent revision (150803-8).
- 2017-01-03TerminatedProcess control
PROFEMUR(R) R RASP HANDLE, REF PPW38078, 1 EACH, NON-STERILE, MicroPort Orthopedics, Arlington, TN Instruments support the implantation of implants. Broach handles mate to the broach to allow for preparation of the femoral canal for the stem component to be implanted. This
The recall is being initiated because all broach handles in the affected lot had the same non-conformity with the connection feature. The broach handles may have machining issues that do not allow the handle to properly mate with an associated broach.
- 2016-12-01TerminatedDevice Design
Lineage Liners: REF 70102846, REF 70202852, REF 70202852, REF 70302858, REF 70303258, REF 70402864, REF 70403264, REF 70403664 For use with the conserve metal head
Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for suspected tissue reaction to metal debris over one percent for modular head CONSERVE(R) systems and DYNASTY(R) metal acetabular liner systems.