Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component)
FDA adverse event reports and recall history for prosthesis, hip, semi-constrained (metal cemented acetabular component) (product code JDL), from the public MAUDE/AEMS corpus and the live FDA recall database.
- Total reports
- 3,877
- Deaths
- 7
- Injuries
- 3,803
- Malfunctions
- 47
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.
Recent recalls
- 2010-08-12TerminatedSoftware design (manufacturing process)
DYNASTY A-CLASS POLY LINER, REF DLXP-LD36, 1 EACH, Rx ONLY, STERILE EO, GROUP D, I.D. 36 mm, LIP 15o, LINER STD, Wright Medical Technology, 4677 Airline Road, Arlington, TN 38002. Product is used in total hip arthroplasty.
The inner and outer product packaging was missing the following information: shelf life, translations, manufacturing date, CE marking and also listed the incorrect sterilization method.
- 2004-07-20TerminatedOther
Crossfire Acetabular II Insert. Catalog No: 2041C-2642.
This device is not approved for sale in the US.
- TerminatedPackaging process control
CONSERVE, TOTAL A-CLASS HEAD W/BFH TECH., REF 38AM-3604, O.D. Size: 36mm, Neck Length: -3.5mm, Taper: SLT, Sterile R, 1 each, Rx only, Wright Medical Technology, Inc., 5677 Airline, Road, Arlington, TN 38002. The CONSERVE Total Femoral Head is indicated for use in total hip arthroplasty for reduction of relief of pain and/or improved hip function in skeletally mature patients.
Incorrect packaging, swapped sizes
- TerminatedPackaging process control
CONSERVE, TOTAL A-CLASS HEAD W/BFH TECH., REF 38AM-3635, O.D. Size: 36mm, Neck Length: +3.5mm, Taper: SLT, Sterile R, 1 each, Rx only, Wright Medical Technology, Inc., 5677 Airline, Road, Arlington, TN 38002. The CONSERVE Total Femoral Head is indicated for use in total hip arthroplasty for reduction of relief of pain and/or improved hip function in skeletally mature patients.
Incorrect packaging, swapped sizes
Monitor prosthesis, hip, semi-constrained (metal cemented acetabular component) automatically
DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.