Merge Healthcare
FDA adverse event reports naming Merge Healthcare as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: MERGE HEALTHCARE.
- Total reports
- 1,133
- Deaths
- 4
- Injuries
- 16
- Malfunctions
- 1,108
Reports per year
Source: FDA adverse event reports by year received, 2014–2026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.
Top device categories
- Computer, Diagnostic, ProgrammableDQK491
- System, Image Processing, RadiologicalLLZ443
- Camera, Ophthalmic, Ac-PoweredHKI86
- Calculator/Data Processing Module, For Clinical UseJQP46
- System, Image Management, OphthalmicNFJ32
- Device, Digital Image Storage, RadiologicalLMB26
- Medical Device Data SystemOUG7
- Gas-Machine, AnesthesiaBSZ1
- Digitizer, Images, OphthalmicNFH1
Recent recalls
- 2025-09-04Open, ClassifiedProcess control
Merge Hemo, Model RCSV2, Model/Catalog Number 99-01172-00, software versions 11.1 and 11.1.1 Product Description: The Merge Hemo, Model RCSV2 device is a hemodynamics recording computer system that monitors, measures, displays, records and stores various physiologic and blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices. The system is used for documenting and monitoring patients undergoing diagnostic and interventional catheter-based procedures. The procedures can be performed in the clinical areas of invasive cardiology, interventional radiology, and cardiac electrophysiology.
Merge Healthcare has been informed by Schiller AG (Baar, Switzerland) of an issue with the Schiller ARGUS PB-3000 Patient Data Module (PDM) where certain units trigger an error message on the Merge Hemo Client (Pressure Sensor Defective) during non-invasive blood pressure (NIBP) measurements, specifically when the initial inflation pressure is set in the high-pressure range (>240 mmHg). The ARGUS PB-3000 is a vital signs monitoring instrument intended to be integrated into the Merge Hemo System. No other components of the Merge Hemo System are affected.
- 2025-03-20Open, ClassifiedSoftware design
VERICIS, Merge Cardio, Model/Catalog Number: Version 9.x, Software Version: 9.0.6 and 9.0.8; VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.
When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.
- 2025-03-20Open, ClassifiedSoftware design
VERICIS, Merge Cardio, Model/Catalog Number: Version 10.x, Software Version: 10.0, 10.1.1.1, 10.1.2, 10.2.0, 10.2 P2 , 10.3; The VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.
When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.
- 2025-03-20Open, ClassifiedSoftware design
VERICIS, Merge Cardio, Model/Catalog Number: Version 11.x, Software Version: 11.0.2, 11.0.3, 11.0.4, 11.1, 11.1.1; The VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.
When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.
- 2025-03-20Open, ClassifiedSoftware design
VERICIS, Merge Cardio, Model/Catalog Number: Version 12.X, Software Version: 12.0.1, 12.0.2; The VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.
When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.
- 2025-03-20Open, ClassifiedSoftware design
Merge Cardio, Model/Catalog Number: Version 12.X, Software Version: 12.1, 12.1.1, 12.2, 12.3, 12.3.1, 12.3.2, 12.3.3, 12.4, 12.4.1, 12.4.2; Merge Cardio is a software medical device that is an integrated multi-modality image cardiovascular information management system.
When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.
- 2019-02-25TerminatedDevice Design
The WinStation Retinal Imager, also known as Merge Eye Station. DCF-1024 & DCF-512 Digital Imaging System Product Usage: The WinStation Retinal Imager has the same intended use as other fundus cameras. It is specifically used to produce color or black and white images of the retina (fundus imaging) and anterior segement (slit lamp imaging), and fluorescein and ICG angiographic images in a non-invasive manner.
Certain configurations of Merge Eye Station hardware may result in electrical output that exceeds the allowable limits for medical devices. Leakage current was found to exceed allowable limits in earth leakage current, touch current in a single fault condition (of open ground) and patient leakage current in a single fault condition (of open ground).
- 2017-08-18TerminatedProcess design
Merge Unity software, formerly known as DR Systems Unity PACS software. The firm name on the label is Merge Healthcare, Hartland, WI. Merge Unity is a medical image and information management system that allows viewing, selection, processing, printing, telecommunications, and media interchange of medical images from a variety of diagnostic imaging systems
The software is not identifying the patient as having atypical hyperplasia, resulting in an incorrect Gail Risk calculation.
- 2017-07-10TerminatedSoftware design
Merge Eye Care PACS Viewer 5.2 Merge Eye Care PACS is a software solution for the display, management, archive, interface and integration of ophthalmic device reports, images and data. Merge Eye Care PACS is a software solution using databases for patient demographics, server and other systems of intercommunication with hospital systems via HL7 and DICOM to provide to clinicians a single image viewing and management solution of images imported from various contributing devices throughout the clinical environment. Merge Eye Care PACS provides the ability to review data from any network-connected computer, and is protected by appropriate security login which permits only authorized user access. Symphony uses 256 bit or greater encryption via secure socket layer (SSL) to assure a network environment which is secure.
When the user has not set up any user preference on the sorting order to render the study images (OS/OD/etc.) and reports, the ECP may get an incorrect study index. If the user attempts to delete an image, ECP may then read that image as OS when it is, in fact, OD.
- 2017-05-25TerminatedSoftware design
Merge LIS software. The firm name on the label is Merge Healthcare. Merge LIS system is a complete system for ordering, managing and reporting a patient s laboratory work, from the time of order entry to the time the laboratory test results are reported.
There are potential issues with results reporting for certain run-based tests. Under certain conditions, the wrong results could inadvertently be verified.
Monitor Merge Healthcare automatically