Merge Healthcare

FDA adverse event reports naming Merge Healthcare as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: MERGE HEALTHCARE.

Total reports
1,134
Deaths
4
Injuries
16
Malfunctions
1,109

Reports per year

14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 20142026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Top device categories

Recent recalls

  • 2026-08-21Open, ClassifiedSoftware Design Change

    Merge PACS and Merge OrthoPACS, medical imaging diagnostic workstation

    Software issue may cause measurements to be displayed inaccurately.

  • 2026-08-04Open, ClassifiedSoftware Design Change

    Merge PACS, medical imaging diagnostic workstation

    Merge Healthcare has identified a potential issue that may cause thumbnails from a previously opened patient study to be displayed in the Thumbnail Bar while a different patient study is open.

  • 2026-08-04Open, ClassifiedSoftware Design Change

    Merge OrthoPACS, medical imaging diagnostic workstation

    Merge Healthcare has identified a potential issue that may cause thumbnails from a previously opened patient study to be displayed in the Thumbnail Bar while a different patient study is open.

  • 2025-09-04Open, ClassifiedProcess control

    Merge Hemo, Model RCSV2, Model/Catalog Number 99-01172-00, software versions 11.1 and 11.1.1 Product Description: The Merge Hemo, Model RCSV2 device is a hemodynamics recording computer system that monitors, measures, displays, records and stores various physiologic and blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices. The system is used for documenting and monitoring patients undergoing diagnostic and interventional catheter-based procedures. The procedures can be performed in the clinical areas of invasive cardiology, interventional radiology, and cardiac electrophysiology.

    Merge Healthcare has been informed by Schiller AG (Baar, Switzerland) of an issue with the Schiller ARGUS PB-3000 Patient Data Module (PDM) where certain units trigger an error message on the Merge Hemo Client (Pressure Sensor Defective) during non-invasive blood pressure (NIBP) measurements, specifically when the initial inflation pressure is set in the high-pressure range (>240 mmHg). The ARGUS PB-3000 is a vital signs monitoring instrument intended to be integrated into the Merge Hemo System. No other components of the Merge Hemo System are affected.

  • 2025-03-20Open, ClassifiedSoftware design

    VERICIS, Merge Cardio, Model/Catalog Number: Version 9.x, Software Version: 9.0.6 and 9.0.8; VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.

    When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.

  • 2025-03-20Open, ClassifiedSoftware design

    VERICIS, Merge Cardio, Model/Catalog Number: Version 10.x, Software Version: 10.0, 10.1.1.1, 10.1.2, 10.2.0, 10.2 P2 , 10.3; The VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.

    When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.

  • 2025-03-20Open, ClassifiedSoftware design

    VERICIS, Merge Cardio, Model/Catalog Number: Version 11.x, Software Version: 11.0.2, 11.0.3, 11.0.4, 11.1, 11.1.1; The VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.

    When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.

  • 2025-03-20Open, ClassifiedSoftware design

    VERICIS, Merge Cardio, Model/Catalog Number: Version 12.X, Software Version: 12.0.1, 12.0.2; The VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.

    When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.

  • 2025-03-20Open, ClassifiedSoftware design

    Merge Cardio, Model/Catalog Number: Version 12.X, Software Version: 12.1, 12.1.1, 12.2, 12.3, 12.3.1, 12.3.2, 12.3.3, 12.4, 12.4.1, 12.4.2; Merge Cardio is a software medical device that is an integrated multi-modality image cardiovascular information management system.

    When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.

  • 2019-02-25TerminatedDevice Design

    The WinStation Retinal Imager, also known as Merge Eye Station. DCF-1024 & DCF-512 Digital Imaging System Product Usage: The WinStation Retinal Imager has the same intended use as other fundus cameras. It is specifically used to produce color or black and white images of the retina (fundus imaging) and anterior segement (slit lamp imaging), and fluorescein and ICG angiographic images in a non-invasive manner.

    Certain configurations of Merge Eye Station hardware may result in electrical output that exceeds the allowable limits for medical devices. Leakage current was found to exceed allowable limits in earth leakage current, touch current in a single fault condition (of open ground) and patient leakage current in a single fault condition (of open ground).

Monitor Merge Healthcare automatically