Calculator/Data Processing Module, For Clinical Use
FDA adverse event reports and recall history for calculator/data processing module, for clinical use (product code JQP), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 2,167
- Deaths
- 5
- Injuries
- 106
- Malfunctions
- 1,984
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2026-07-31Open, ClassifiedUnder Investigation by firm
REMISOL Advance with Software Version 2.5 UDI-DI code: 15099590802073 The Normand Info REMISOL Advance Data Manager collects and manages data and workflow for 128 connections
Due to software issues where does not handle combo assay results. The software may replace the final combo test result with one of the sub-results resulting in incorrect reporting of test outcomes.
- 2026-07-31Open, ClassifiedUnder Investigation by firm
Brand Name: WASPLab Product Name: WASPLab Model/Catalog Number: W087 Software Version: WebApp version 2.33.1, 2.33.3 and 2.34.2 Product Description: WASPLab is an automated device intended to be used for the incubation and the digital imaging of agar culture plates streaked with microbiological specimens derived by the human body for microbiological investigations. WebApp software is the user interface of WASPLab device.
Due to a software issue within the WebApp that potentially prevents the user to overwrite the results.
- 2025-10-23Open, ClassifiedSoftware design
BD EpiCenter Microbiology Data Management System, labeled as the following: 1. Computer System Epicenter F/G, Catalog Number: 441002. 2. COMPUTER OPTIPLEX EPICENTER SPARE, Catalog Number: 441421. 3. Generic NUC 2, Catalog Number: 445398. 4. Nuc 5i5 Os Image Bow 2.X, Catalog Number: 443971.
Product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
- 2025-09-16Open, ClassifiedSoftware design
MEDITECH Expanse Laboratory (LAB), Microbiology (MIC), Anatomical Pathology (PTH), Genetics (GEN). Calculator/data processing module for clinical use.
Entering multiple keys that trigger input simultaneously may remove data from first field of screens and/or questionnaires with more than one field.
- 2025-06-25Open, ClassifiedSoftware design
REMISOL Advance UDI-DI code: 13700962601874 All Software Versions: B92487; B92488; C24317; C28652; C37500; D04164; C44703; C57017; C69412; C69413; C73942; C73941; C88470; C88471 The Normand Info REMISOL Advance Data Manager collects and manages data and workflow for 128 connections (Beckman Coulter Instruments, automations, LIS&). It can also be connected to a host system. It provides data management capabilities such as automatic results validation, delta checking, reflex testing, quality control, results editing, archiving and restoration of patient results.
Due to a software issue, the software may incorrectly process the date of birth (DoB) received from the Laboratory Information System (LIS) and when the DoB is set to 01/01/1900 or left empty, REMISOL Advance may incorrectly identify a patient as a newborn, affecting the execution of validation rules based on incorrect reference ranges. The test results are not changed by this software defect and are accurately displayed by REMISOL Advance; however, the evaluation flagging of the results may be affected. This issue may lead to wrong reference range calculations and can result in erroneous flagged results being auto validated and uploaded to LIS.
- 2024-12-03Open, ClassifiedSoftware design
BD EpiCenter Single User Software Version or Model: 441007 Catalog Number: 441007
BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
- 2024-12-03Open, ClassifiedSoftware design
BD EpiCenter Data Management System Version or Model: 440887 Catalog Number: 440887
BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
- 2024-12-03Open, ClassifiedSoftware design
BD EpiCenter Data Management System, Multi User Software Version or Model: 440981 Catalog Number: 440981
BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
- 2024-12-03Open, ClassifiedSoftware design
BD Synapsys Informatics Solution (Laboratory Solution) Version or Model: 444150 Catalog Number: 444150
BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
- 2024-10-08Open, ClassifiedUnder Investigation by firm
BD Synapsys Informatics Solution with software versions: 4.20, 4.30, 5.10, 5.20 and 5.30- IVD standalone laboratory software solution providing data management and workflow management functionality across locations where clinical diagnostic activities take place. Catalog Number: 444150
Synapsys software versions 4.20, 4.30, 5.10, 5.20, or 5.30 and utilizing the in-line breakpoint feature, an error is occurring where BD Synapsys will incorrectly display breakpoints for organisms that do not have defined breakpoints in CLSI or EUCAST. This occurs when the identified organism is included in the same interpretive group as a species for which there is a specific breakpoint. May result in either a false susceptible or false resistant result. A false susceptible result can cause inadequate or incomplete antibiotic treatment due to the use of an ineffective antibiotic, and a false resistant result may limit access to a preferred antibiotic regimen with possible elongation of therapy
Monitor calculator/data processing module, for clinical use automatically
DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.