Boston Scientific Neuromodulation Corporation
FDA adverse event reports naming Boston Scientific Neuromodulation Corporation as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: BOSTON SCIENTIFIC NEUROMODULATION CORPORATION.
- Total reports
- 21,973
- Deaths
- 226
- Injuries
- 20,092
- Malfunctions
- 1,651
Reports per year
Source: FDA adverse event reports by year received, 2011–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Top device categories
- Stimulator, Spinal-Cord, Totally Implanted For Pain ReliefLGW19,769
- Prosthesis, Spinous Process Spacer/PlateNQO841
- Stimulator, Electrical, Implanted, For Parkinsonian TremorMHY779
- Stimulator, Electrical, Implanted, For Parkinsonian SymptomsNHL389
- Stimulator, Electrical, Implanted, For Essential TremorPJS159
- Probe, Radiofrequency LesionGXI20
- Generator, Lesion, RadiofrequencyGXD10
- Stimulator, Spinal-Cord, Totally Implanted For Relief Of Pain Due To Diabetic NeuropathyQRB3
- Suture, Nonabsorbable, Synthetic, PolyethyleneGAT2
- Stimulator, Spinal-Cord, Implanted (Pain Relief)GZB1
Recent recalls
- 2026-08-05Open, ClassifiedDevice Design
INFINION CX LEAD KIT, 50CM, Model: SC-2317-50, Material: M365SC2317500; INFINION CX LEAD KIT, 70CM, Model: SC-2317-70, Material: M365SC2317700. part of the Boston Scientific Spinal Cord Stimulator (SCS) Systems.
Spinal cord stimulation system lead body may be exposed to stress at the anchor site. This may result in high impedance measurements or lead fractures, which may lead to inadequate stimulation, additional intervention, including lead explant and replacement.
- 2026-07-08Open, ClassifiedPackaging process control
Intracept Access Instruments, Description/REF: INTRACEPT ACCESS INSTRUMENTS- DRILL/RLV0035, INTRACEPT ACCESS INSTRUMENTS- DRILL EU/RLV0035-EU, INTRACEPT ACCESS INSTRUMENTS- 2VB/RLV0045, INTRACEPT ACCESS INSTRUMENTS- 2VB EU/RLV0045-EU, INTRACEPT ACCESS INSTRUMENTS- 3VB/RLV0055, INTRACEPT ACCESS INSTRUMENTS- 3VB EU/RLV0055-EU, INTRACEPT ACCESS INSTRUMENTS 2 VB/RLV0065, INTRACEPT ACCESS INSTRMNTS ADDITIONAL VB/RLV0075, INTRACEPT ACCESS INSTRMNTS ADDTL VB EU/RLV0075-EU
Intraosseous nerve ablation system instruments may have sterile barrier breach of the outer tray seal, which may introduce a risk of sterility breach, and if breached could lead to severe infection.
- 2026-06-11Open, ClassifiedUnder Investigation by firm
Boston Scientific CSK Electrodes: REF: CSK-TC10, CSK Electrode, STAINLESS STEEL, 10 cm x 28 ga (0.36 mm) REF: CSK-TC10-3M,CSK Electrode, STAINLESS STEEL, 3M Cable, 10 cm x 28 ga (0.36 mm); REF: CSK-TC15, CSK Electrode, STAINLESS STEEL, 15 cm x 28 ga (0.36 mm); REF: CSK-TC15-3M, CSK Electrode, STAINLESS STEEL, 3M Cable, 15 cm x 28 ga (0.36 mm) REF: CSK-TC20, CSK Electrode, STAINLESS STEEL, 20 cm x 28 ga (0.36 mm) REF: CSK-TC5,CSK Electrode, STAINLESS STEEL, 5 cm x 28 ga (0.36 mm); REF: CSK-TC5-3M, CSK Electrode, STAINLESS STEEL, 3M Cable, 5 cm x 28 ga (0.36 mm)
Reusable electrodes may not meet expected performance levels.
- 2026-06-11Open, ClassifiedUnder Investigation by firm
Boston Scientific TCN Electrodes: REF: TCN-5, TCN Electrode, NITINOL, 5 cm x 0.45 mm (0.0175 in); REF:TCN-10, TCN Electrode, NITINOL, 10 cm x 0.45 mm (0.0175 in) REF:TCN-15, TCN Electrode, NITINOL, 15 cm x 0.45 mm (0.0175 in); REF:TCN-20, TCN Electrode, NITINOL, 20 cm x 0.45 mm (0.0175 in) REF: TCN-5-3m, TCN Electrode, NITINOL, 3M CABLE, 5 cm x 0.45 mm (0.0175 in) REF: TCN-10-3M, TCN Electrode, NITINOL, 3M CABLE, 10 cm x 0.45 mm (0.0175 in); REF: TCN-15-3M, TCN Electrode, NITINOL, 3M CABLE, 15 cm x 0.45 mm (0.0175 in) REF: TCN-20-3M, TCN Electrode, NITINOL, 3M CABLE, 20 cm x 0.45 mm (0.0175 in)
Reusable electrodes may not meet expected performance levels.
- 2026-06-11Open, ClassifiedPackaging process control
Boston Scientific Model SC-4318, Clik X Anchor REF: M365SC43180
Sterile anchors my lack sterility assurance.
- 2025-09-12Open, ClassifiedLabeling design
Vercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09
Wire break(s) have occurred in rechargeable deep brain stimulation implantable pulse generators, implanted in the submuscular pectoral location. Wire breaks prevent successful stimulation, thus requiring device removal/replacement and users may experience high monopolar impedances, undesired sensation, loss of therapy, return of symptoms and/or Bluetooth connectivity challenges.
- 2025-09-02Open, ClassifiedDevice Design
Superion IDS Kit Superion Indirect Decompression System 8MM, VF Implant 101-9808 00884662000529 Superion Indirect Decompression System 10MM, VF Implant 101-9810 00884662000536 Superion Indirect Decompression System 12MM, VF Implant 101-9812 00884662000543 Superion Indirect Decompression System 14MM, VF Implant 101-9814 00884662000550 Superion Indirect Decompression System 16MM, VF Implant 101-9816 00884662000567 Superion Indirect Decompression System Kit, VF Instrument 102-9800 00884662000574 The Vertiflex Procedure Instrument Platform, VF Instrument 140-9800 00884662000611 The Superion IDS Kit includes a set of proprietary instruments to deliver the Superion Implant in a minimally invasive manner. Instruments specifically designed for implanting the Superion Implant are sterile, single-use disposable instruments, consisting of a Dilator Assembly, Cannula Assembly, Interspinous Gauge, Inserter, Reamer, and Driver.
As a result of a comprehensive product performance review it was determine that the device and instruments are not consistently meeting expected performance levels. Issues include breakage of the instruments and implant/device.
- 2025-08-19Open, ClassifiedComponent change control
Burr Hole Cover Kit, REF: M365DB4600C0; Burr Hole Cover Spares Kit, REF: M365DB4605C0. Components of Vercise Deep Stimulation System used with the Surgical Implant Manual
Resistance and difficulties could occur while closing the retaining clip locking mechanism, which is part of a burr hole cover kit, which is part of a Deep Brain Stimulation (DBS) system, which may lead to procedural delay, or lead-migration-which may require additional surgery.
- 2024-08-14Open, ClassifiedSoftware design
WAVEWRITER ALPHA: Model: SC-1216, 16 IPG KIT, REF: M365SC12160; Model: SC-1232, 32 IPG KIT, REF: M365SC12320
Spinal Cord Stimulation (SCS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device reset, which may lead to transient loss of stimulation; patients may experience undesired sensations when therapy turns off for approx. 10-15 seconds and then back on, which may lead patient to request surgical intervention for replacement or revision.
- 2024-05-24Open, ClassifiedDevice Design
Vercise Genus Deep Brain Stimulation (DBS) Implantable Pulse Generator IPG: R16 IPG KIT, REF: M365DB12160, and R32 IPG KIT, REF: M365DB12320
Deep Brain Stimulation (DBS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device reset. The device reset could lead transient loss of stimulation; patients may experience undesired sensations, transient worsening of movement disorder symptoms, which may lead patient to request surgical intervention for replacement or revision.
Monitor Boston Scientific Neuromodulation Corporation automatically
DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.