Skip to main content
NHLClass 3

Stimulator, Electrical, Implanted, For Parkinsonian Symptoms

FDA adverse event reports and recall history for stimulator, electrical, implanted, for parkinsonian symptoms (product code NHL), from the public MAUDE corpus and the live FDA recall database.

Total reports
1,837
Deaths
42
Injuries
1,611
Malfunctions
182

Reports per year

03
12
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2003–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2025-09-12Open, ClassifiedLabeling design

    Vercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09

    Wire break(s) have occurred in rechargeable deep brain stimulation implantable pulse generators, implanted in the submuscular pectoral location. Wire breaks prevent successful stimulation, thus requiring device removal/replacement and users may experience high monopolar impedances, undesired sensation, loss of therapy, return of symptoms and/or Bluetooth connectivity challenges.

  • 2025-08-19Open, ClassifiedComponent change control

    Burr Hole Cover Kit, REF: M365DB4600C0; Burr Hole Cover Spares Kit, REF: M365DB4605C0. Components of Vercise Deep Stimulation System used with the Surgical Implant Manual

    Resistance and difficulties could occur while closing the retaining clip locking mechanism, which is part of a burr hole cover kit, which is part of a Deep Brain Stimulation (DBS) system, which may lead to procedural delay, or lead-migration-which may require additional surgery.

  • 2024-05-24Open, ClassifiedDevice Design

    Vercise Genus Deep Brain Stimulation (DBS) Implantable Pulse Generator IPG: R16 IPG KIT, REF: M365DB12160, and R32 IPG KIT, REF: M365DB12320

    Deep Brain Stimulation (DBS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device reset. The device reset could lead transient loss of stimulation; patients may experience undesired sensations, transient worsening of movement disorder symptoms, which may lead patient to request surgical intervention for replacement or revision.

  • TerminatedLabeling Change Control

    Fifty-five (55) cm 8 Contact Extension Kit, part number M365NM3138550 - Product Usage: The Deep Brain Stimulation (DBS) System is indicated for use in bilateral stimulation of the subthalamic nucleus (STN) as an adjunctive therapy in reducing some of the symptoms of moderate to advanced levodopa-responsive Parkinson s disease (PD) that are not adequately controlled with medication.

    The firm is initiating a voluntary removal of certain spinal cord stimulation leads, lead extensions, adapters, and deep brain stimulation lead extensions due to the "use by date" displayed on the label which is later than the correct "use by date". No adverse health consequence is expected to occur from this issue.

Monitor stimulator, electrical, implanted, for parkinsonian symptoms automatically