Skip to main content

Arrow International LLC

FDA adverse event reports naming Arrow International LLC as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: ARROW INTERNATIONAL LLC.

Total reports
13,417
Deaths
44
Injuries
691
Malfunctions
12,682

Reports per year

20
22
23
24
25
26

Source: FDA adverse event reports by year received, 20202026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.

Top device categories

Recent recalls

  • 2026-06-13Open, ClassifiedNonconforming Material/Component

    PARACENTESIS KIT, REF: AK-00376, ASK-00376-VCU

    Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

  • 2026-06-13Open, ClassifiedNonconforming Material/Component

    Description/REF: ACCESS TRAY/ASK-04001-SUH-S; ACCESS TRAY/ASK-04001-VM1; STIMUCATH SFTY KIT: 19G CTH, 17Gx4CM NDL/AB-19604-KS; STIMUCATH SFTY KIT: 19G CTH, 17Gx8CM NDL/AB-19608-KS; STIMUCATH SFTY KIT: 20G CTH, 18Gx8CM NDL/AB-20608-KS; STIMUCATH KIT: 19G CATH, 17G x 9CM NDL/AB-05060-PK; FLEXBLOCK ULTRASOUND KIT: 17Gx9CM NS NDL/ASK-19609-VCU; CONTINUOUS NERVE BLOCK KIT/ASK-19600-CC2; CONTINUOUS NERVE BLOCK KIT 19 GA X 60/ASK-19608-CC; CONTINUOUS NERVE BLOCK KIT 19GA X 60CM/ASK-19608-KP; CONTINUOUS NERVE BLOCK KIT 19GA X 60CM/ASK-19608-SHC; FLEXBLOCK KIT:ECHO CATH, 17Gx8CM STM NDL/FB-19608-K; FLEXBLOCK ULTRASOUND KIT: 17Gx8CM STM ND/FB-19608-KU; FLEXBLOCK KIT:ECHO CATH, 17Gx9CM NS NDL/FB-19609-K; FLEXBLOCK ULTRASOUND KIT: 17Gx9CM NS NDL/FB-19609-KU; CONTINUOUS PERIPHERAL NERVE BLOCK KIT/ASK-19608-US1; EPIDURAL CATHETERIZATION KIT/ASK-05001-SLR1; EPIDURAL CATHETERIZATION KIT/MH-05001; EPIDURAL CATHETERIZATION KIT/WJ-05401; SPINAL ANES/EPIDURAL CATH 19GA X 90CM/ASK-05501-USC1; EPIDURAL NEEDLE KIT/LG-03000; EPIDURAL ANESTHESIA KIT/MTO-03000-VM; EPIDURAL ANESTHESIA KIT/SM-03001; Anesthesia conduction kit/FB-19611-K

    Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

  • 2026-06-13Open, ClassifiedNonconforming Material/Component

    Description/REF: BIOPSY SYSTEM COMPREHENSIVE KIT/ASK-09458-AH, ASK-09458-COC, ASK-09458-EUH, ASK-09458-SD, ASK-09458-VU

    Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

  • 2026-06-13Open, ClassifiedNonconforming Material/Component

    Description/REF: PACING PSI KIT: 6 FR/AI-06210-KS, PACING PSI KIT: 5 FR 2-L/AI-07155-KS, PACING PSI KIT: 5 FR 2L/ASK-07155-UPM

    Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

  • 2026-06-13Open, ClassifiedNonconforming Material/Component

    Description/REF: AGBA PICC G4 Stylet: 3L 6FR x 55CM BP/DLX-45563-VPSB; AGBA PICC G4 Stylet: 3L 6FR x 55CM TCG/DLX-45563-VPSC; ARROW(R) VPS(R) ACCESS KIT/ASK-04001-UCHD1; ARROW(R) VPS(R) ACCESS KIT/ASK-04001-UCHS1; ARROW(R) VPS(R) ACCESS KIT/ASK-04001-DU13; ARROW(R) VPS(R) ACCESS KIT/ASK-04001-UNHS1

    Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

  • 2026-06-13Open, ClassifiedNonconforming Material/Component

    Description/REF: RA CATH KIT: 20 GA X 1 3/4 IN/TJ-04020-1; RA CATH KIT: 20 GA X 1 1/2IN/NA-04220-S1A; ARTERIAL LINE KIT/AK-04510-S; ARTERIAL CATH KIT: 18 GA X 6IN (16 CM)/ASK-04018-HUMC; ARTERIAL LINE KIT: 18 GA X 6 IN (16 CM)/ASK-04018-OH1; RA CATH KIT: 20 GA X 1 3/4IN (4.45 CM)/ASK-04020-BID; RA CATH KIT: 20 GA X 1-3/4IN/ASK-04020-MAR; ARTERIAL LINE KIT: 20 GA X 5 IN/ASK-04020-PM4; RA CATH KIT: 20 GA X 1 1/2IN (3.81 CM)/ASK-04120-HF3; RA CATH SET: 20 GA X 2 1/2 IN (6.35 CM)/ASK-04120-UR; ARTERIAL CATH KIT: 20 GA X 3 1/16 IN/ASK-04125-MGH; ARTERIAL CATH KIT: 20 GA X 5 IN/ASK-04500-HFH4; ARTERIAL LINE KIT: 20GA X 5IN/ASK-04510-BID2; ARTERIAL LINE KIT: 20GA X 5IN (12 CM)/ASK-04510-EHP1; ARTERIAL LINE KIT/ASK-04510-FEM1; ARTERIAL LINE KIT: 20 GA X 5 IN (12 CM)/ASK-04510-HFH; ARTERIAL CATH KIT: 20 GA X 5 IN/ASK-04510-HFH4; ARTERIAL LINE KIT/ASK-04510-SE; ARTERIAL LINE KIT: 20 GA X 5 IN (12CM)/ASK-04510-TGH2; RA CATH KIT: 20 GA X 12 CM/NA-04550-1A; RA CATH KIT: 20 GA X 12 CM/NA-04550-X1A; RA CATH KIT: 18 GA X 1-3/4IN/AK-04018; RA CATH KIT: 20 GA X 1-3/4IN/AK-04020; RA CATH KIT: 20 GA X 1-3/4IN/AK-04020-C; RA CATH KIT: 20 GA X 1-1/2IN/AK-04120; RA CATH KIT: 22 GA X 1-3/8IN/AK-04122; RA CATH KIT: 20 GA X 1-1/2IN/AK-04220; RA CATH KIT: 20 GA X 1 3/4 IN (4.45 CM)/ASK-04020-AH; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-ATH; RA CATH KIT: 20 GA X 1 3/4IN (4.45CM)/ASK-04020-AU2; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-BHS; ARTERIAL CATHETERIZATION KIT/ASK-04020-BIDB2; RA CATH KIT: 20 GA X 4.45 CM/ASK-04020-CCD1; ARTERIAL CATHETERIZATION KIT/ASK-04020-CR; ART LINE KIT: 20 GA X 1 3/4 IN (4.45 CM)/ASK-04020-EUH2; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-FAM1; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-FMC; RA CATH KIT: 20 GA X 1-3/4IN/ASK-04020-KSF2; RA CATH KIT: 20 GA X 1 3/4 IN (4.5 CM)/ASK-04020-LHC2; RA CATH KIT: 20 GA X 1 3/4 IN (4.5 CM)/ASK-04020-LHC3; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-LM; RA CATH KIT: 20 GA X 1 3/4IN/ASK-04020-MCH1; RA CATH KIT: 20 GA X 1-3/4IN (4.45 CM)/ASK-04020-MIH1; RA CATH KIT: 20 GA X 4.45 CM/ASK-04020-NWMX; RA CATH KIT: 20 GA X 1-3/4IN/ASK-04020-PHP; RA CATH KIT: 20 GA X 1-3/4 IN (4.45CM)/ASK-04020-RIHM; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-RMC1; RA CATH KIT: 20 GA X 1-3/4IN/ASK-04020-RU; RA CATH KIT: 20 GA X 1 3/4 IN (4.45 CM)/ASK-04020-SE; RA CATH KIT: 20 GA X 1-3/4IN/ASK-04020-SHM; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-SRMC1; RA CATH KIT: 20 GA X 1 3/4 IN (4.45 CM)/ASK-04020-UCL1; RA CATH KIT: 20 GA X 4.45 CM/ASK-04020-UMI; RA CATH KIT: 20 GA X 1-3/4 IN (4.45CM)/ASK-04020-UN; RA CATH KIT: 20 GA X 1-3/4IN (4.45 CM)/ASK-04020-UNC; RA CATH KIT: 20 GA X 1-3/4IN (4.45 CM)/ASK-04020-UNC1; RA CATH SET: 20 GA X 1-1/2IN (3.81CM)/ASK-04220-PSU; RA CATH KIT: 20 GA X 1-1/2 IN (3.81 CM)/ASK-04220-UCL1; RA CATH KIT: 20 GA X 1-3/4IN/NA-04020-S1A; CVC KIT: 20 GA X 5 IN (12CM)/ASK-04550-JMC1; ARTERIAL ACCESS TRAY/ASK-04001-BW2; ARTERIAL ACCESS TRAY WITH .025 IN SWG/ASK-04001-WUH1; ARTERIAL LINE KIT/ASK-04510-HUM

    Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

  • 2026-06-13Open, ClassifiedNonconforming Material/Component

    Description/REF: PSI KIT: 7 FR X 4IN ANTIMICROBIAL/ASK-09701-UMH; PSI KIT: 8.5 FR X 4 1/8 IN (10 CM)/ASK-09803-OSU4; PSI/CVC KIT: 9 FR/7 FR X 6IN (16CM)/ASK-09903-AMC1; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-09903-CMC1; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-09903-OSU4; PSI KIT: 9 FR X 4 1/8 IN (10CM)/ASK-09903-PCMH1; PSI KIT: 8.5 FR X 10 CM/CDC-09803-1A; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-09903-ECL3; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-11242-PHP2; MULTI LUMEN/PSI KIT/ASK-11142-NSU; MULTI-LUMEN/PSI KIT/AK-21242-X; MAC KIT: 2L 9 FR DISTAL X 10 CM AGB/ASK-21142-DMC; MAC KIT: 2L 9 FR DISTAL X 10 CM AGB/ASK-21142-GMC1; MULTILUMEN/PSI KIT: 9 FR X 4 IN (10 CM)/ASK-21142-HF5; MAC KIT: 2L 9 FR DISTAL X 10 CM AGB/ASK-21142-LMDT; 2L MAC KIT: 9 FR DISTAL X 4IN (10 CM)/ASK-21142-UCL1; 2-L MAC KIT: 9 FR DISTAL X 4 IN (10 CM)/ASK-21142-UPMC2; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-SL1; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UMH; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UN1; MAC KIT: 2-L 9 FR DISTAL X 10 CM AGB/CDC-21142-XCN1A; MAC KIT: 2-L 9 FR DISTAL X 10 CM AGB/CDC-21142-XN1A; MAC KIT: 2-L 9 FR DISTAL X 11.5 CM AGB/CDC-21242-XCN1A; MAC KIT: 2-L 9 FR DISTAL X 11.5 CM AGB/CDC-21242-XN1A; 2L MAC KIT: 9 FR DISTAL X 4 IN/ASK-21142-CO1; 2L MAC KIT: 9 FR DISTAL X 4 IN (10 CM)/ASK-21142-FH3; 2L MAC KIT: 9 FR DISTAL X 4 IN (10 CM)/ASK-21142-OHSU3; MULTILUMEN/PSI KIT: 9 FR X 4 IN (10 CM)/ASK-21142-SG2; MULTILUMEN/MAC KIT: 9 FR X 4 IN (10 CM)/ASK-21142-UOI1; 2 LUMEN/PSI KIT: 9 FR X 4 IN (10 CM)/ASK-21142-WHC2; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-BID1; AGB MAC KIT: 2L 9 FR DISTAL/ASK-21242-BSWH2; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-DU1; MULTILUMEN/PSI KIT 9 FR X 11.5 CM/ASK-21242-PCMH1; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UCH2; AGB MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UHC2; PSI KIT: 8.5 FR X 10 CM ANTIMICROBIAL/ASK-29803-BSWH2; PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-29803-HHC; PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-29803-MCY1; PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-29803-UPM1; PSI KIT: 8.5 FR X 4 1/8 IN (10 CM)/ASK-29804-VCU; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/ASK-29903-MCY1; PSI KIT: 9 FR/ASK-29903-MHI; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-29903-OHU4; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-29903-UCH2; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/ASK-29903-UCL1; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/ASK-29903-UHC4; PSI KIT: 9FR X 4IN (10 CM) ANTIMICROBIAL/ASK-29903-UPM1; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-29907-LGH; PSI KIT: 8.5 FR X 10 CM ANTIMICROBIAL/CDC-29803-1A; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/CDC-29903-1A; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/CDC-29903-X1A; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/CDC-29903-XCN1A; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/CDC-29903-XN1A; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/ASK-29903-AHH; PSI KIT: 8.5 FR X 10 CM ANTIMICROBIAL/CDC-29803-XN1A; PSI ACCESS TRAY/ASK-04001-MM2; MAC KIT: 2L 9 FR DISTAL X 10 CM AGB/ASK-21142-GMC; MULTI LUMEN/PSI KIT: 9 FR DISTAL X 4 1/2/ASK-21242-OSU3

    Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

  • 2026-06-13Open, ClassifiedNonconforming Material/Component

    Description/REF: PI AGBA PICC KIT: 3L 6 FR X 40 CM W/VPS/ASK-44063-NS AGBA PICC/DELTA: 2L 5.5 FR X 50 CM/ASK-45052-NS AGBA PICC/DELTA KIT: 1L 4.5 FR X 55 CM/ASK-45541-MHBV PICC KIT: 2L 5.5 FR X 55 CM CG+ VPS/ASK-45552-RH AGBA PICC NaviCurve: 2L 5.5FR x 55CM BP/DLX-45552-CURVB AGBA PICC NaviCurve: 2L 5.5FR x 55CM TCG/DLX-45552-CURVC AGBA PICC: 2L 5.5FR x 55CM w Biopatch/DLX-45552-HPKB AGBA PICC NaviCurve: 3L 6FR x 55CM BP/DLX-45563-CURVB PI PICC KIT: 4 FR X 50 CM/ASK-35041-NH PI PICC KIT: 2-L 5 FR X 50 CM/ASK-35052-NH PI PICC G4 Stylet: 1L 4FR x 55CM w CHG/DLX-35541-VPSC PI PICC NaviCurve: 2L 5FR x 55CM w BP/DLX-35552-CURVB PI PICC NaviCurve: 3L 6FR x 55CM w BP/DLX-35563-CURVB PI PICC G4 Stylet: 3L 6FR x 55CM w BP/DLX-35563-VPSB AGBA PI PICC KIT: 2L 5.5 FR X 55CM/CDC-45552-HPK1A PI AGBA PICC KIT: 3-L 6 FR X 55 CM/CDC-45563-HPK1A PI AGBA PICC KIT: 1L 4.5 FR X 55 CM VPS/ASK-45541-UOI1 PICC KIT: 1L 4.5 FR X 55 CM VPS/ASK-45541-VFH1 PI AGBA PICC KIT: 2L 5.5 FR X 55 CM VPS/ASK-45552-UOI1 PICC KIT: 2L 5.5 FR X 55 CM VPS/ASK-45552-VFH1 PI AGBA PICC KIT: 3L 6 FR X 55 CM W/VPS/ASK-45563-UWH1 AGBA PICC NaviCurve: 1L 4.5FR x 55CM BP/DLX-45541-CURVB AGBA PICC NaviCurve: 1L 4.5FR x 55CM TCG/DLX-45541-CURVC AGBA PICC G4 Stylet: 1L 4.5FR x 55CM TCG/DLX-45541-VPSC AGBA PICC G4 Stylet: 2L 5.5FR x 55CM BP/DLX-45552-VPSB AGBA PICC G4 Stylet: 2L 5.5FR x 55CM TCG/DLX-45552-VPSC PI PICC:1L 4FR X 50CM PRELOAD VPS STYLET/CDC-35041-VPS PICC KIT: 3 FR X 55 CM/CDC-05531-PK1A

    Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

  • 2026-06-13Open, ClassifiedNonconforming Material/Component

    Description/REF: EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05502-VM1; EPIDURAL CATHETERIZATION KIT: 19 GA/AK-05501; EPIDURAL CATHETERIZATION KIT/AK-05502; EPIDURAL CATHETERIZATION KIT/AK-05502D; EPIDURAL CATHETERIZATION KIT/AK-05503; EPIDURAL CATHETERIZATION KIT/AK-05503-L; SPINAL ANES/EPIDURAL CATH KIT/AK-05560; SPINAL ANES/EPIDURAL CATH KIT/ASK-02520-SBMC1; EPIDURAL CATHETERIZATION KIT/ASK-05400-BW1; EPIDURAL CATHETERIZATION KIT/ASK-05400-GH; SPINAL ANES/EPIDURAL CATH KIT/ASK-05401-LMC; EPIDURAL CATHETERIZATION KIT/ASK-05401-NM; SPINAL ANES/EPIDURAL CATH KIT/ASK-05500-BID1; EPIDURAL CATHETERIZATION KIT: 19 GA/ASK-05500-JHH1; EPIDURAL CATHETERIZATION KIT: 19 GA/ASK-05500-MCV1; EPIDURAL CATHETER KIT: 19GA X 90CM ASK-05500-MVH; EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05500-UMC; 19 GA X 35.5 IN EPIDURAL CATHETER KIT/ASK-05502-MLH; SPINAL ANES/EPIDURAL CATH KIT/ASK-05560-JMH1; SPINAL ANES/EPIDURAL CATH KIT/ASK-05560-MKD2; SPINAL ANES/EPIDURAL CATH KIT/ASK-05560-OM1; EPIDURAL CATHETERIZATION KIT/ASK-17019-SM; EPIDURAL CATHETERIZATION KIT/BP-05501; EPIDURAL CATHETERIZATION KIT/CK-05401; EPIDURAL CATHETERIZATION KIT/HS-05501; EPIDURAL CATHETERIZATION KIT/MM-05501; EPIDURAL CATHETERIZATION KIT: 19 GA/MP-17019-TK; EPIDURAL CATHETERIZATION KIT: 19 GA/MP-17019-TKD; EPIDURAL CATHETERIZATION KIT: 19 GA/MP-17019-TKL; EPIDURAL CATHETERIZATION KIT/MSO-05401-JMH; EPIDURAL CATHETERIZATION KIT/NM-05401; EPIDURAL CATHETERIZATION KIT/SJ-05501; EPIDURAL CATHETERIZATION KIT/SM-05401; EPIDURAL CATHETER KIT: 19 GA X 90 CM/TM-05401; EPIDURAL CATHETERIZATION KIT/TM-05502; SPINAL ANES/EPIDURAL CATH KIT/UI-05502; SPINAL ANES/EPIDURAL CATH KIT: 19 GA/ASK-05400-SHM; EPIDURAL CATH KIT: 19 GA/ASK-05400-UOM; EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05500-MR; EPIDURAL CATHETERIZATION KIT: 19 GA/ASK-05500-TM; EPIDURAL CATHETERIZATION KIT/ASK-05501-BHL1; EPIDURAL CATHETERIZATION KIT/ASK-05501-GH; EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05501-SM; EPIDURAL CATHETERIZATION KIT/ASK-05502-BSM; EPIDURAL CATHETERIZATION KIT/ASK-05502-MGH1; EPIDURAL CATHETERIZATION KIT/ASK-05503-BWB1; EPIDURAL CATHETERIZATION KIT/ASK-05503-HMH; EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05503-SPH; EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05503-UPM1; SPINAL ANES/EPIDURAL CATH KIT/ASK-05560-UCI2; EPIDURAL CATHETERIZATION KIT: 19 GA/ASK-17019-UMP; EPIDURAL CATHETERIZATION KIT/MH-05503-1; EPIDURAL CATHETERIZATION KIT/ASK-05401-VB2; Epidural Catheterization Kit: 19GA/ASK-05500-SC1; Epidural Catheterization Kit: 19GA/ASK-05500-SC3; EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05501-AFH1; EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05501-BMC; EPIDURAL CATHETERIZATION KIT/ASK-05502-NY; EPIDURAL CATHETERIZATION KIT/ASK-17019-RMH

    Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

  • 2026-06-13Open, ClassifiedNonconforming Material/Component

    Description/REF: HEMODIALYSIS KIT: 2L 14 FR X 20 CM AGB CDC-25142-X1A

    Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

Monitor Arrow International LLC automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.