Skip to main content

Tandem Diabetes Care

FDA adverse event reports naming Tandem Diabetes Care as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: TANDEM DIABETES CARE.

Total reports
1,330,642
Deaths
284
Injuries
155,520
Malfunctions
1,174,836

Reports per year

14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2014–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Top device categories

Recent recalls

  • 2026-09-29Open, ClassifiedSoftware Design Change

    Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS Model/Catalog Number: iOS software version 2.13 Software Version: iOS software version 2.13 UDI-DI: 00389152470740 Component: No

    A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a software issue preventing successful pairing of the t:slim application with Abbott FreeStyle Libre 3 Plus (FSL3P) continuous glucose monitoring (CGM) sensors, resulting in the insulin pump being unable to establish a near-field communication (NFC) connection with the CGM and requiring FSL3P users to use their t:slim insulin pumps in manual mode. FSL3P users are unable to access their CGM's glucose values from within the t:slim app and cannot use their pumps in automated control (closed-loop) mode, which could lead to users experiencing hypoglycemia or hyperglycemia.

  • 2026-03-25Open, ClassifiedProcess change control

    Tandem pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utilize a motor-driven mechanism to deliver insulin from within a disposable cartridge, through an infusion set, into a patient s subcutaneous tissue. As with current insulin infusion pumps on the U.S. market, the desired timing and quantity of the insulin delivery is programmed by the user (i.e., the patient).

    The Czech language user guide contained multiple translation errors. The most significant error was that it incorrectly instructs users to verify that the infusion set is connected to the body, when it should instead direct users to confirm that the infusion set is not connected to the body. Following this instruction could result in unintended insulin delivery, potentially leading to a hypoglycemic event.

  • 2025-11-05Open, ClassifiedSoftware design

    Tandem Mobi Insulin Pump: Description/Catalog: Tandem insulin pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utilize a motor-driven mechanism to deliver insulin from within a disposable cartridge, through an infusion set, into a patient s subcutaneous tissue. As with current insulin infusion pumps on the U.S. market, the desired timing and quantity of the insulin delivery is programmed by the user (i.e., the patient). Pump, Tandem Mobi, Control-IQ/1010750, Pump, Tandem Mobi, Control-IQ, Replacement/1012719, Pump, Tandem Mobi, Control-IQ 7.6.0.3, Medicare/1013501, Pump, Tandem Mobi, Control-IQ 7.7/1013655, Pump, Tandem Mobi, Control-IQ, 7.7, Replacement/1013656, Pump, Tandem Mobi, Control-IQ 7.7, Medicare/1013700, Pharmacy Kit, Starter Pack, Tandem Mobi 7.7/1014081

    Insulin pump includes a vibration motor that gives tactile feedback for any alerts, alarms, or malfunctions. Pump may exhibit false vibration motor failure due to a software issue causing Malfunction 12: "Pump cannot operate, the mobile app can no longer receive data from the pump. Insulin delivery and any active CGM Sessions have been stopped", which could result in hyperglycemia.

  • 2025-10-23Open, ClassifiedSoftware change control

    t:slim X2 insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014070, 1014071, 1014072, 1014073, and User Guide when using Dexcom G7 sensor; Tandem Mobi insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014801, 1014811, 1014813, and User Guide when using Dexcom G7 sensor

    A software defect in versions 7.9.0.1 and 7.10.1 of the pump software for Tandem t:slim X2 and version 7.9.0.1 of the Tandem Mobi pump with Control-IQ+ technology that are paired with a G7 sensor may result in an unexpected automatic insulin correction bolus (autobolus ) which could result in hypoglycemia.

  • 2025-09-24Open, ClassifiedSoftware design

    t:slim X2 Insulin Pump with Interoperable Technology (with Tandem t:slim mobile app)

    an app defect that occurs when the phone is set to a right-to-left language, which causes app-pump pairing issues and graphical defects, which can potentially lead to incorrect therapy decisions and subsequent hypoglycemia or hyperglycemia.

  • 2025-08-27Open, ClassifiedNonconforming Material/Component

    Battery Operated infusion pumps. t:slim X2 G5, Replacement, Refurbished, REF: 1000911; t:slim X2 GS Classic, Replacement, Refurbished, REF: 1003808; t: slim X2, Basal-IQ, mg/dl, REF: 1004219; Pump, t:slim X2, Clinical Use Only, REF: 1004484; Pump, t:slim X2, Control-IQ, REF: 1005611; Pump, t:slim X2, Control-IQ, mmoUL, Refurbished Replacement, REF: 1005613; Pump, t:slim X2, Control-IQ, mg/dl, REF: 1005615; Pump, t:slim X2, Control-IQ, mg/dl, Refurbished Replacement, REF: 1005617; Pump, t:slim X2, Basal-IQ 6.4, Refurbished Replacement, REF: 1006379; Pump, t:slim X2, Control-IQ 7.4, REF: 1006402; Pump, t:slim X2, Control-IQ 7.4, Replacement, REF: 1006406; Pump, t:slim X2, Control-IQ 7.4, REF: 1006408; Pump, t:slim X2, Basal-IQ, , REF: 1006419; Pump, t:slim X2, Basal-IQ, mg/dL, REF: 1006420; Pump, t:slim X2, Basal-IQ, mmoUL, Refurbished Replacement, REF: 1006535; Pump, t:slim X2, Basal-IQ, mg/dL, Refurbished Replacement, REF: 1006537; Pump, t:slim X2, Control-IQ, mg/dl , Replacement, REF: 1008255; Pump, t:slim X2, Control-IQ, mmoUL, Replacement, REF: 1008256; Pump, t:slim X2, Basal-IQ, mmoUL, Replacement, REF: 1008257; Pump, t:slim X2, Basal-IQ, mg/dl ,Replacement, REF: 1008258; Pump, t:slim X2, Basal-IQ, REF: 1010004; Pump, t:slim X2, Basal-IQ, Replacement, REF: 1010005; Pump, t:slim X2, Basal-IQ, Medicare, REF: 1010006; Pump, t:slim X2, Basal-IQ, Medicare, REF: 1010007; Pump, t:slim X2, Control-IQ, REF: 1010009; Pump, t:slim X2, Control-IQ Replacement, REF: 1010010; Pump, t:slim X2, Control-IQ, Medicare, REF: 1010011; Pump, t:slim X2, Control-IQ, Refurbished Replacement, REF: 1010012; Pump, t:slim X2, Control-IQ, REF: 1012867; Pump, t:slim X2, Control-IQ, Medicare, REF: 1012868; Pump, t:slim X2, Control-IQ, Replacement, REF: 1012869; Pump, t:slim X2, Control-IQ 7.8, REF: 1013195; Pump, t:slim X2, Control-IQ 7.8, Medicare, REF: 1013196; Pump, t :slim X2, Control-IQ 7.8, Replacement, REF: 1013197; Pump, t:slim X2, Control-IQ 7.8, Refurbished Replacement, REF: 1013198

    Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result in a hyperglycemic event, which may lead to injury and/or hospitalization.

  • 2025-04-11Open, ClassifiedSoftware Design Change

    t:slim X2 Insulin Pump with Interoperable Technology

    A software defect in Version 7.9 of the pump software for Tandem t:slim X2 and Tandem Mobi pumps, when used with Control IQ+ technology, will cause the pump to incorrectly interpolate glucose trends when the Estimated Glucose Value (EGV) is above 255 mg/dL at the start or end of a gap in data collection due to a lapse in connection from a paired continuous glucose monitor (CGM) sensor, which can lead to under-delivery or over-delivery of insulin based on inaccurate result leading to severe cases of hypoglycemia or hyperglycemia.

  • 2025-04-11Open, ClassifiedSoftware Design Change

    Tandem Mobi Insulin Pump with Interoperable Technology

    A software defect in Version 7.9 of the pump software for Tandem t:slim X2 and Tandem Mobi pumps, when used with Control IQ+ technology, will cause the pump to incorrectly interpolate glucose trends when the Estimated Glucose Value (EGV) is above 255 mg/dL at the start or end of a gap in data collection due to a lapse in connection from a paired continuous glucose monitor (CGM) sensor, which can lead to under-delivery or over-delivery of insulin based on inaccurate result leading to severe cases of hypoglycemia or hyperglycemia.

  • 2024-05-06Open, ClassifiedSoftware Design Change

    t:connect mobile app used in conjunction with t:slim X2 insulin pump with Control-IQ technology

    During normal use, the mobile app version 2.7 may crash and be automatically relaunched by the iOS operating system. This cycle intermittently repeats, which leads to excessive Bluetooth communication that may result in pump battery drain and may lead to the pump shutting down sooner than typically expected. Pump shutdown will cause insulin delivery to suspend, which could lead to an under-delivery of insulin and may result in hyperglycemia, including severe hyperglycemia.

  • 2022-09-26Open, ClassifiedDevice Design

    t:slim X2 insulin pump; t:slim X2 insulin pump with Dexcom G5; t:slim X2 insulin pump with Basal-IQ technology; t: slim x2 insulin pump with the Control-IQ Technology

    Insulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On, 4) Unexpected Open Loop, that can be mitigated with a software update. Issues could result in hypoglycemia, hyperglycemia or diabetic ketoacidosis.

Monitor Tandem Diabetes Care automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.