Skip to main content

Syncardia Systems, Inc.

FDA adverse event reports naming Syncardia Systems, Inc. as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: SYNCARDIA SYSTEMS, INC..

Total reports
1,429
Deaths
16
Injuries
40
Malfunctions
1,356

Reports per year

05
06
07
08
09
10
11
12
13
14
15
16
17

Source: FDA adverse event reports by year received, 20052017. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.

Top device categories

Recent recalls

  • 2015-09-17TerminatedComponent design/selection

    Temporary Total Artificial Heart (TAH-t) with Freedom Driver System REF 595000-001 Product Usage: Freedom Drivers are components of the SynCardia temporary Total Artificial Heart (TAH-t) System, which is indicated for use in transplant-eligible candidates at risk of imminent death from biventricular failure. Freedom Drivers are intended for use in and out of the hospital.

    SynCardia Systems, Inc. is recalling Freedom Drivers due to a specific component of the drive mechanism may fail and cause the drive mechanism to stop pumping; resulting in the loss of life-sustaining function.

  • 2012-12-28TerminatedProcess control

    Companion External Battery, Part Number: 293001-001 The Companion External Battery is a component of the Companion 2 Driver System and the Companion Driver System. Both Driver Systems, with the temporary Total Artificial Heart (TAH-t) are indicated for use as a bridge to transplantation in cardiac transplant-eligible candidates at risk of imminent death from biventricular failure.

    Syncardia Systems is recalling the Companion External Batteries used with the Companion 2 Driver System because they may potentially be at risk of thermal events and/or other physical damage such as melting, charring and/or burning.

  • TerminatedProcess control

    5000 Series Freedom Drivers. Freedom Driver System for Temporary Total Artificial Heart (TAH-t). Part number 595000-001

    The Main Printed Circuit Board Assembly (PCBA) of the affected Freedom Drivers may fail and cause the Freedom Driver to stop functioning without visual or audible alarms, resulting in the loss of life-sustaining function.

Monitor Syncardia Systems, Inc. automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.