ST. Jude Medical, Inc.

FDA adverse event reports naming ST. Jude Medical, Inc. as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: ST. JUDE MEDICAL, INC..

Total reports
39,272
Deaths
526
Injuries
27,231
Malfunctions
11,475

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 20122026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.

Top device categories

Recent recalls

  • 2019-08-02TerminatedProcess control

    ELLIPSE VR, REF: CD1377-36C Implantable cardioverter defibrillators (ICDs)

    Electrical failures were identified in cardioverter defibrillators (ICDs) due to damaged aluminum wires. ICDs may contain electrical wire connections which may not be completely insulated. The potential patient impact could be the inability to deliver high voltage therapy. There is no available option to verify the vulnerability status for implanted devices.

  • 2019-08-02TerminatedProcess control

    ELLIPSE VR, REF: CD1377-36QC Implantable cardioverter defibrillators (ICDs)

    Electrical failures were identified in cardioverter defibrillators (ICDs) due to damaged aluminum wires. ICDs may contain electrical wire connections which may not be completely insulated. The potential patient impact could be the inability to deliver high voltage therapy. There is no available option to verify the vulnerability status for implanted devices.

  • 2019-08-02TerminatedProcess control

    ELLIPSE VR, Tiered-therapy cardioverter/defibrillator, REF: CD1411-36Q, UDI: 05414734507738

    Electrical failures were identified in cardioverter defibrillators (ICDs) due to damaged aluminum wires. ICDs may contain electrical wire connections which may not be completely insulated. The potential patient impact could be the inability to deliver high voltage therapy. There is no available option to verify the vulnerability status for implanted devices.

  • 2019-08-02TerminatedProcess control

    ELLIPSE DR, REF: CD2377-36QC Implantable cardioverter defibrillators (ICDs)

    Electrical failures were identified in cardioverter defibrillators (ICDs) due to damaged aluminum wires. ICDs may contain electrical wire connections which may not be completely insulated. The potential patient impact could be the inability to deliver high voltage therapy. There is no available option to verify the vulnerability status for implanted devices.

  • 2019-08-02TerminatedProcess control

    ELLIPSE DR, Tiered-therapy cardioverter/defibrillator, REF: CD2411-36C, UDI: 05414734507585

    Electrical failures were identified in cardioverter defibrillators (ICDs) due to damaged aluminum wires. ICDs may contain electrical wire connections which may not be completely insulated. The potential patient impact could be the inability to deliver high voltage therapy. There is no available option to verify the vulnerability status for implanted devices.

  • 2019-08-02TerminatedProcess control

    ELLIPSE DR, Tiered-therapy cardioverter/defibrillator, REF: CD2411-36Q, UDI: 05414734507615

    Electrical failures were identified in cardioverter defibrillators (ICDs) due to damaged aluminum wires. ICDs may contain electrical wire connections which may not be completely insulated. The potential patient impact could be the inability to deliver high voltage therapy. There is no available option to verify the vulnerability status for implanted devices.

  • 2018-03-23TerminatedSoftware change control

    Proclaim DRG Implantable Pulse Generator, Model Number 3664

    The firm received complaints of error messages that occurred during routine impedance checks on Proclaim DRG IPGs, model 3664. Some complaints were also associated with transient over stimulation which created discomfort for the patients.

  • 2017-10-10TerminatedDevice Design

    Fortify VR - Model Numbers: CD1231-40 and CD1231-40Q; Fortify ST VR, Model Numbers: CD1241-40 and CD1241-40Q; Fortify Assura VR, Model Numbers: CD1257-40, CD1257-40Q, CD1357-40C, CD1357-40Q; Fortify Assura ST VR, Model Numbers: CD1263-40, CD1263-40Q, CD1363-40C, CD1363-40Q; Fortify DR, Model Numbers: CD2231-40 and CD2231-40Q; Fortify ST DR, Model Numbers: CD2241-40 and CD2241-40Q; Fortify Assura DR, Model Numbers: CD2257-40, CD2257-40Q, CD2357-40C, CD2357-40Q; Fortify Assura ST DR, Model Numbers: CD2263-40, CD2263-40Q, CD2363-40C, CD2363-40Q Product Usage Fortify and Fortify Assura implantable cardioverter defibrillators (ICDS) are intended to provided ventricular anti-tachycardia pacing and ventricular defibrillation for automated treatment of life threatening ventricular arrhythmias. The devices may also be programmed to provide single or dual chamber bradycardia pacing

    The device may exhibit premature battery depletion.

  • 2017-10-10TerminatedDevice Design

    Unify, Model Numbers: CD3231-40 and CD3231-40Q; Unify Quadra, Model Numbers: CD3249-40 and CD3249-40Q; Unify Assura, Model Numbers: CD3257-40, CD3257-40Q, CD3357-40C, CD3357-40Q; Quadra Assura, Model Numbers: CD3265-40, CD3265-40Q, CD3365-40C, CD3365-40Q; Quadra Assura MP, Model Numbers: CD3269-40, CD3269-40Q, CD3369-40C, CD3369-40Q Unify and Unify Assura cardiac resynchronization therapy defibrillators (CRTDs) are intended to provide ventricular anti-tachycardia pacing and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias. The devices may also be programmed to provide single or dual chamber bradycardia pacing, in additional to cardiac resynchronization therapy via left ventricular pacing

    The device may exhibit premature battery depletion.

  • 2016-10-21TerminatedDevice Design

    Fortify VR, Model No. CD1231-40, CD1231-40Q. Fortify ST VR, Model No. CD1241-40, CD1241-40Q. Fortify Assura VR, Model No. CD1257-40, CD1257-40Q, CD1357-40C. Fortify Assura ST VR, Model No. CD1263-40, CD1263-40Q, CD1363-40, CD1363-40Q. Fortify DR, Model No. CD2231-40, CD2231-40Q. Fortify ST DR, Model No. CD2241-40, CD-2241-40Q, CD2263-40, CD2363-40C, CD2363-40Q. Fortify Assura DR, Model No. CD2257-40, CD2257-40Q, CD2357-40C, CD2357-40Q. Fortify and Fortify Assura implantable cardioverter defibrillators (ICDS) are intended to provided ventricular anti-tachycardia pacing and ventricular defibrillation for automated treatment of life threatening ventricular arrhythmias. The devices may also be programmed to provide single or dual chamber bradycardia pacing. Unify and Unify Assura cardiac resynchronization therapy defibrillators (CRTDs) are intended to provide ventricular anti-tachycardia pacing and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias. The devices may also be programmed to provide single or dual chamber bradycardia pacing, in additional to cardiac resynchronization therapy via left ventricular pacing

    St Jude Medical is recalling implantable cardiac defibrillator (ICD) and cardiac resynchronization therapy defibrillator (CRT-D) devices because they are subject to premature battery depletion due to lithium cluster with the battery causing a short circuit.