Skip to main content

Remote Diagnostic Technologies Ltd.

FDA adverse event reports naming Remote Diagnostic Technologies Ltd. as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: REMOTE DIAGNOSTIC TECHNOLOGIES LTD..

Total reports
2,643
Deaths
12
Injuries
25
Malfunctions
2,606

Reports per year

21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2021–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Top device categories

Recent recalls

  • 2026-01-08Open, ClassifiedDevice Design

    Philips Tempus Pro Patient Monitor, REF: 00-1004, 00-1007, 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R. USB 2-Channel IBP Module, REF: 01-2017. Used with User/Operator Manual

    Patient monitor measures Intracranial Pressure (ICP) and Bladder Pressure (BDR) but ICP and BDR measurements have not been tested or validated for accuracy. If healthcare providers base clinical judgment on unvalidated measurements, this could potentially lead to incorrect clinical decisions.

  • 2025-07-25Open, ClassifiedSoftware change control

    Inseego USB8 4G Dongle Kit, Part Number: 01-2298, which contains USB Flash Drive, Part Number: 43-2011. The kit is an accessory to the Tempus Pro Monitor, REF: 00-1007-R, 00-1024-R, 00-1026-R

    Kit USB flash drive contains outdated software, which could result in reverting patient monitor to outdated software and reintroducing the video laryngoscope issue (related to an August 2023 recall), which could lead to delay in diagnosis, delayed treatment, hypoxia due to unexpected loss of video laryngoscopy and all other monitor measurements during system restart.

  • 2025-05-07Open, ClassifiedSoftware design

    Philips Tempus Pro Patient Monitor, REF: 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R, 00-1004, 00-1007

    Patient Monitor software update to address issues:1)May freeze at startup with MPM Application error and Attention Restart Required messages,2)Monitor may freeze and reboot at start of 12-lead ECG or after, 3)May reboot after intermittent connection with Smart Mount, 4) 12-lead ECG measurements with Louvain algorithm and software before vx.36 could produce inaccurate rhythm/morphology statements.

  • 2025-04-24Open, ClassifiedDevice Design

    Tempus Pro Patient Monitor, REFs: 00-1004-R (Tempus Pro non-printer, IBP, without headset dock), 00-1007-R and 00-1007 (Tempus Pro non-printer, IBP, with headset dock), 00-1024-R (Tempus Pro printer), 00-1026-R (Tempus Pro printer, IBP), 00-1007 [(Tempus Pro, BT Dock (IB Only)], 00-1004 00-1004 [Tempus Pro (IB Only)]

    software issue that causes an error screen on a portable vital signs monitor when attaching video Laryngoscope.

  • 2024-11-26Open, ClassifiedLabeling Change Control

    Tempus LS-Manual Defibrillator Model 00-3020 The Tempus LS-Manual is a defibrillator that can deliver a shock in manual or Pacemaker mode

    Due to incorrect labels applied to devices during servicing. Labels contain the incorrect product name and CAT number.

  • 2024-06-06Open, ClassifiedLabeling design

    Philips Tempus LS-Manual Defibrillator, Part Number: 00-3020 when using Tempus LS-Manual User/Operator Manual

    Simultaneous ECG measurements with multiple devices, including a manual defibrillator, on the same patient may cause electrical noise or interference, which may lead to distortions of the ECG signal; distortions may lead to no or ineffective pacing, or no or inappropriate defibrillation pulse.

  • 2023-09-21Open, ClassifiedDevice Design

    Philips Tempus Pro Patient Monitor, REF: 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R, 00-1004, and 00-1007, when Used With Tempus Pro Video Laryngoscope, REF 01-2044

    Full screen error message may occur on patient monitors with affected hardware version, either during or after laryngoscopes are unplugged from monitors, informing users a 60-100 second shutdown/restart are required. Error prevents: viewing graphical patient vital signs, sounding of pulse audio tone, visualization of patient airway used during intubation; may lead to treatment delay or hypoxia.

  • 2023-06-12Open, ClassifiedDevice Design

    Philips Tempus LS-Manual Defibrillator, Part Number: 00-3020

    Defibrillator/pacemaker module (DPM) may encounter a problem causing communication from the Mainboard to the DPM to fail, leading to no or ineffective pacing with displayed error message "HW failure DPM".

  • 2023-04-01Open, ClassifiedUnder Investigation by firm

    Tempus Pro Patient Monitor is a portable vital signs monitor intended to be used by clinicians and medically qualified personnel, for the attended or unattended monitoring of single or multiple vital signs in clinical and pre hospital care applications. The affected power supplies are related to the following Tempus Pro Patient Monitoring units: Model Numbers: 00-1004 00-1004-R 00-1007 00-1007-R 00-1024-R 00-1026-R Model Numbers: 00-1004 00-1004-R 00-1007 00-1007-R 00-1024-R 00-1026-R

    During internal testing, Regulatory Compliance issues were identified regarding the Monitor and Power Supplies for fluid ingress and basic safety issues.

  • 2020-12-11TerminatedSoftware design

    Tempus LS - Manual Defibrillator, Model Number: 00-3020 Software Version: 1.3.4 - Product Usage: The device is intended to be used in the following environmental: - Primarily in pre-hospital care and transport applications (including ambulances, fixed and rotary wing aircraft).

    A software error was detected within software version 1.3.4 for the Tempus LS device. When exiting pacer back into manual mode the pacer pulse wave may put the ECG calibration out of calibration. This software error may result in a delay in therapy, as the clinician is not provided an ECG waveform on which a shockable rhythm decision can be made. The device is still able to deliver a manual defibrillation shock.

Monitor Remote Diagnostic Technologies Ltd. automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.