Pulmonetic Systems, Inc.
FDA adverse event reports naming Pulmonetic Systems, Inc. as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: PULMONETIC SYSTEMS, INC..
- Total reports
- 1,847
- Deaths
- 50
- Injuries
- 45
- Malfunctions
- 1,741
Reports per year
Source: FDA adverse event reports by year received, 2000–2012. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.
Top device categories
Recent recalls
- 2005-11-23TerminatedOther
Lap Top Ventilator (LTV) Series continuous ventilator, model LTV-1000. The LTV1000 ventilator is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The LTV-1000 is a prescription medical device suitable for use in institutional, home and transport settings.
Pulmonetic Systems has identified the potential of a printed circuit board malfunction, which can result in a ventilator malfunctio (e.g. Vent INOP, HW Fault, Xducer Fault) and possibly resulting in failure of the ventilator to breathe for the patient. This malfunction may occur without an accompanying audible alarm.
- 2005-11-23TerminatedOther
Lap Top Ventilator (LTV) Series continuous ventilator, model LTV-950. The LTV-950 ventilator is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The LTV-950 is a prescription medical device suitable for use in institutional, home and transport settings.
Pulmonetic Systems has identified the potential of a printed circuit board malfunction, which can result in a ventilator malfunctio (e.g. Vent INOP, HW Fault, Xducer Fault) and possibly resulting in failure of the ventilator to breathe for the patient. This malfunction may occur without an accompanying audible alarm.
- 2005-11-23TerminatedOther
Lap Top Ventilator (LTV) Series continuous ventilator, model LTV-900. The LTV-900 ventilator is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The LTV-900 is a prescription medical device suitable for use in institutional, home and transport settings.
Pulmonetic Systems has identified the potential of a printed circuit board malfunction, which can result in a ventilator malfunctio (e.g. Vent INOP, HW Fault, Xducer Fault) and possibly resulting in failure of the ventilator to breathe for the patient. This malfunction may occur without an accompanying audible alarm.
- 2004-12-08TerminatedOther
Universal Cable Adaptor (Part #17765-001 and labeled on outer package as part no. 17820-001) distributed for Field Installation on certain LTV series Ventilator, Model 800 with either part # 10134 or 11511 power boards affected by the Pulmonetic Systems, Inc., November 1, 2004 Class I recall (FDA Z-1485-04).
Reports have been received indicating that the Universal Cable Adaptor, intended to correct an earlier Class I recall of LTV Series Ventilators due to their inability to continue operating when external power source is switched to internal battery, is not functioning as intended and/or the cable is not securely attaching to the connection on the ventilator.
- 2004-12-08TerminatedOther
Universal Cable Adaptor (Part #17765-001and labeled on outer package as part no. 17820-001) distributed for Field Installation on certain LTV Series Ventilator, Model 950 with either part # 10134 or 11511 power boards affected by the Pulmonetic Systems, Inc., November 1, 2004 Class I recall. (FDA Z-1485-04)
Reports have been received indicating that the Universal Cable Adaptor, intended to correct an earlier Class I recall of LTV Series Ventilators due to their inability to continue operating when external power source is switched to internal battery, is not functioning as intended and/or the cable is not securely attaching to the connection on the ventilator.
- 2004-12-08TerminatedOther
Universal Cable Adaptor (Part #17765-001and labeled on outer package as part no. 17820-001) distributed for Field Installation on certain LTV Series Ventilator, Model 900 with either part # 10134 or 11511 power boards affected by the Pulmonetic Systems, Inc., November 1, 2004 Class I recall. (FDA Z-1485-04)
Reports have been received indicating that the Universal Cable Adaptor, intended to correct an earlier Class I recall of LTV Series Ventilators due to their inability to continue operating when external power source is switched to internal battery, is not functioning as intended and/or the cable is not securely attaching to the connection on the ventilator.
- 2004-12-08TerminatedOther
Universal Cable Adaptor (Part #17765-001and labeled on outer package as part no. 17820-001) distributed for Field Installation on certain LTV Series Ventilator, Model 1000 with either part # 10134 or 11511 power boards affected by the Pulmonetic Systems, Inc., November 1, 2004 Class I recall. (FDA Z-1485-04)
Reports have been received indicating that the Universal Cable Adaptor, intended to correct an earlier Class I recall of LTV Series Ventilators due to their inability to continue operating when external power source is switched to internal battery, is not functioning as intended and/or the cable is not securely attaching to the connection on the ventilator.
Monitor Pulmonetic Systems, Inc. automatically