FDA adverse event reports naming Moog Medical Devices Group as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: MOOG MEDICAL DEVICES GROUP.
- Total reports
- 9,204
- Deaths
- 23
- Injuries
- 76
- Malfunctions
- 8,967
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Top device categories
- Pump, InfusionFRN5,621
- Pump, Infusion, EnteralLZH3,154
- Enteral Specific Transition ConnectorsPIO366
- Tubes, Gastrointestinal (And Accessories)KNT20
- Pump, Infusion, ElastomericMEB20
- Set, Administration, IntravascularFPA10
- Pump, Infusion, PcaMEA7
- Gastrointestinal Tubes With Enteral Specific ConnectorsPIF5
- Syringe, PistonFMF1
- Accessories, Pump, InfusionMRZ1
Recent recalls
- 2019-07-23TerminatedNonconforming Material/Component
MOOG Curlin Infusion Administration Set, REF 340-4130, Non-DEHP Tubing with Non-Vented Bag Spike and Vented 0.2 Micron Filter, packaged 20 sets/carton, sterile. Product Usage: Administration set to be used with the Moog Curlin Infusion Pump. The pump can be used for intravenous, intra-arterial, epidural, subcutaneous, or enteral infusions.
Administration sets leaked at the filter.
- 2019-07-23TerminatedNonconforming Material/Component
MOOG Curlin Infusion Administration Set, REF 340-4130-V, Non-DEHP Tubing with Non-Vented Bag Spike and Vented 0.2 Micron Filter with add-on check valve, packaged 20 sets/carton, sterile. The firm name on the label is ZEVEX, Inc., Salt Lake City, UT.
Administration sets leaked at the filter.
- 2016-02-11TerminatedProcess control
Moog Curlin Infusion, Ambulatory Infusion Pump The Curlin Ambulatory Infusion System includes Curlin Ambulatory multi-therapy and PainSmart pumps. Curlin 4000, catalog number 340; Curlin 6000, 6000 CMS, and PainSmart IOD, catalog number 360; and Curlin Pump Calibration Set, catalog number 340-4038.
Moog is recalling Curlin 4000, 6000, and PainSmart Infusion Pumps manufactured or serviced by MMDG between March 18, 2015 and November 6, 2015 because the master pumps used to calibrate them were out of tolerance and may cause an over delivery of medications and solutions at a rate of 1.2 - 6.8% faster than programmed.
- 2013-01-31TerminatedNonconforming Material/Component
Infinity Enteralite Ambulatory Feeding Pumps, List number: INFKIT2 - Addendum to Operator's Manual Product usage: provide a means for the volumetric delivery of fluids used in Enteral applications.
The firm is adding an addendum to the operator's manual for Infinity Enteralite Ambulatory Feeding Pumps pertaining to the Infinite Dose mode and the inability of the Pump to detect air in line.
- 2012-11-15TerminatedComponent design/selection
EnteraLite Infinity Enteral Feeding Pump identified with the following list numbers: INFKIT2 and INFORNG. The device is used with disposable PVC administration tubing sets to provide a means for the volumetric delivery of fluids used in enteral applications. The EnteraLite Infinity enteral feeding pump with disposable PVC administration tubing sets and user accessories provide a means for the volumetric delivery of fluids used in enteral applications.
MOOG Medical Devices Group is conducting a voluntary for a limited set of serial numbers of the EnteraLite Infinity and Infinity Orange Small Volume Enteral Feeding Pumps because certain diodes on the PCB assembly may result in insufficient voltage to a portion of the pump resulting in an error 13 message and shutting the pump down entirely.
- 2012-10-17TerminatedProcess control
***This recall is being conducted due to an incomplete recall for RES 57829*** Curlin Infusion Administration Set, REF 340-4165, Sterile EO, The Curlin Ambulatory Infusion Pump System with disposable PVC Administration Tubing Sets and user accessories, provide a means for the volumetric delivery of fluids used in Parenteral, Enteral, Epidural, Subcutaneous and Intravenous applications.
This recall is the same issue for the FDA processed recall initiated by Moog Medical Devices on 1/25/11 which was related to customers reporting unclearable Air-In-Line alarm with specific lot numbers of Curlin Administration Sets used with Curlin Ambulatory Infusion Pumps. The lots associated with this recall were inadvertently omitted from the original recall which has since been terminated by
- 2012-05-30TerminatedProcess design
Non-DEHP Microbore Administration Set with Non-Vented Spike Packaged Assembly, product code: 340-4114, Method of Sterilization using Ethylene Oxide. The Curlin Ambulatory Infusion Pump System with disposable PVC Administration Tubing Sets and user accessories, provide a means for the volumetric delivery of fluids used in Parenteral, Enteral, Epidural, Subcutaneous and Intravenous applications.
Moog Medical Device Group became aware on March 23, 2012 via customer complaint, of a reversed pump segment on an Administration Set. This malfunction could reverse the flow of fluid or medication backwards from what was intended.
- 2012-05-30TerminatedProcess design
Non-DEHP Microbore Administrative Set with Female Luer Adapter Packaged Assembly, product code: 340-4115, Method of Sterilization using Ethylene Oxide. The Curlin Ambulatory Infusion Pump System with disposable PVC Administration Tubing Sets and user accessories, provide a means for the volumetric delivery of fluids used in Parenteral, Enteral, Epidural, Subcutaneous and Intravenous applications.
Moog Medical Device Group became aware on March 23, 2012 via customer complaint, of a reversed pump segment on an Administration Set. This malfunction could reverse the flow of fluid or medication backwards from what was intended.
- 2012-05-30TerminatedProcess design
Epidural Admin Set with Non-Vented Bag Spike and 0.2 Micron Filter Package Assembly, product code: 340-4126, Method of Sterilization using Ethylene Oxide. The Curlin Ambulatory Infusion Pump System with disposable PVC Administration Tubing Sets and user accessories, provide a means for the volumetric delivery of fluids used in Parenteral, Enteral, Epidural, Subcutaneous and Intravenous applications.
Moog Medical Device Group became aware on March 23, 2012 via customer complaint, of a reversed pump segment on an Administration Set. This malfunction could reverse the flow of fluid or medication backwards from what was intended.
- 2012-05-30TerminatedProcess design
Non-DEHP Admin Set with Non-Vented Spike and 1.2 Micron Filter Packaged Assembly, product code: 340-4128, Method of Sterilization using Ethylene Oxide. The Curlin Ambulatory Infusion Pump System with disposable PVC Administration Tubing Sets and user accessories, provide a means for the volumetric delivery of fluids used in Parenteral, Enteral, Epidural, Subcutaneous and Intravenous applications.
Moog Medical Device Group became aware on March 23, 2012 via customer complaint, of a reversed pump segment on an Administration Set. This malfunction could reverse the flow of fluid or medication backwards from what was intended.