Microvention, Inc.
FDA adverse event reports naming Microvention, Inc. as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: MICROVENTION, INC..
- Total reports
- 5,414
- Deaths
- 152
- Injuries
- 2,430
- Malfunctions
- 2,824
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Top device categories
- Device, Neurovascular EmbolizationHCG1,770
- Device, Vascular, For Promoting EmbolizationKRD1,494
- Intrasaccular Flow Disruption DeviceOPR679
- Intracranial Aneurysm Flow DiverterOUT517
- Intracranial Coil-Assist StentQCA268
- Catheter, PercutaneousDQY224
- Catheter, Intravascular Occluding, TemporaryMJN171
- Intracranial Neurovascular StentNJE104
- Catheter, Intravascular, DiagnosticDQO72
- Catheter, Thrombus RetrieverNRY38
Recent recalls
- 2026-08-07Open, ClassifiedUnder Investigation by firm
BOBBY Balloon Guide Catheter REF: BOB-895-CN BOB-895-EMEA BOB-895-EMEA-AL BOB-895-US MV-BGA01095 For use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neurovascular systems.
Due to an increase rate of balloon pinhole defects.
- 2026-05-14Open, ClassifiedMixed-up of materials/components
LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL): 214022-CAS/4.0 mm x 22 mm x 18 mm, 214518-CAS/4.5 mm x 18 mm x 14 mm, 213517-CAS/3.5 mm x 17 mm x 13 mm
Stent-assisted coiling device may due to commingling during manufacturing have dimensional nonconformance and/or incorrect device configuration. If device is oversized may lead to increased vessel wall stress, neointimal hyperplasia, vessel injury/rupture; if undersized may lead to incomplete expansion/malposition, acute stent thrombosis, distal embolization, ischemic neurological events
- 2025-10-10Open, ClassifiedDevice Design
Flow Re-Direction Endoluminal Devices: FRED 27, REF: FRED5019, FRED5029, FRED4528, FRED4026, FRED5526, FRED4508, FRED5514, FRED3536, FRED3511, FRED5014, FRED3516, MV-F451827; FRED X 27, REF: XFRED4017-MVE, XFRED4528-MVE, MV-F501427X, XFRED4518-MVE, XFRED5526-MVE, XFRED4539-MVE, FREDX5522-PMA, FREDX3522-PMA, FREDX4525-PMA, FREDX5015-PMA, FREDX5532-PMA, FREDX4013-PMA, FREDX3513-PMA, FREDX4515-PMA, FREDX3517-PMA, MV-F352427X
Flow diverter may have a manufacturing issue related to the tantalum length and/or tantalum attachment pattern, which may lead the device to be incompletely open or not properly apposed to the vessel wall, which may necessitate device recapture and withdrawal of the delivery system and microcatheter together. Insufficient apposition could lead to long-term health consequences.
- 2025-07-28Open, ClassifiedDevice Design
LIFEPEARL Drug Elutable Microspheres: 200 +/- 50 micrometers, REF: 8LP2S200; 400 +/- 50 micrometers, REF: 8LP2S400
Drug Elutable Microspheres have a smaller actual average diameter that is not within specification, which may lead to inability to reach the desired treatment location (non-targeted embolization), additional procedure/treatment required, blockage other than target vessel, increased procedure time, incomplete embolization, and/or inability to treat the patient.
- 2025-02-07Open, ClassifiedNonconforming Material/Component
MicoVention Terumo, LVIS Intraluminal Support Device, REF: 213522-CAS, od, 3.5 mm, TL 22 mm, WL 18 mm, Sterile R, CE0297. The intended purpose of the LVIS Device is to serve as a stent-assisted coiling (SAC) device providing support to the embolic coils used in the treatment of wide-necked aneurysms.
Potential for Polytetrafluoethylene (PTFE) material encasing the stent loop may prevent the stent from being pushed out of the introducer.
- 2025-01-17Open, ClassifiedPackaging process control
Microvention, Headway" 17 Advanced Straight Microcatheter with Hydrophilic Coating, REF: MC172150SX-CN, Sterile R, Rx Only, Lot# 0000629846
Sterility assurance; microcatheter inner packaging may contain incomplete seal.
- 2024-09-17Open, ClassifiedUnder Investigation by firm
ERIC Retrieval Device REF: ER173020 ER174030 ER176044 The ERIC Retrieval Device is a mechanical thrombectomy device designed to restore blood flow by effectively removing clots from occlusive vasculature in patients suffering from acute ischemic stroke.
Due to Retrieval Device product labels stating the incorrect expiration dates that (5 years instead of 3 years) which extends beyond their true expiration date.
- 2024-01-31Open, ClassifiedDevice Design
TERUMO HydroPearl Compressible Microspheres for Embolisation REF 8HP2S600
Due to, during the manufacturing process, the prescribed manufacturing and quality processes were not followed and the product was inadvertently distributed.
- 2023-10-27Open, ClassifiedProcess control
AZUR CX 35 Peripheral Coil System Detachable 13mm x 24 CM, REF: 45-751324
Peripheral coil system detachable has a potential of unsealed pouch packaging.
- 2023-07-27Open, ClassifiedProcess control
FRED Flow Re-Direction Endoluminal Device, REF: FRED3507-PMA, FRED3511, FRED3511-PMA, FRED3516-CA, FRED3516-PMA, FRED3524, FRED3524-CA, FRED3524-PMA, FRED3536-CA, FRED3536-PMA, FRED4007-PMA, FRED4012-CA, FRED4012-PMA, FRED4017-PMA, FRED4026, FRED4026-PMA, FRED4038-PMA, FRED4508, FRED4508-PMA, FRED4513-PMA, FRED4518-PMA, FRED4528, FRED4528-PMA, FRED4539, FRED4539-CA, FRED4539-PMA, FRED5009, FRED5009-PMA, FRED5014-PMA, FRED5019-PMA, FRED5029, FRED5029-CA, FRED5029-PMA, FRED5514-CA, FRED5514-PMA, FRED5526-CA, FRED5526-PMA, MV-F351127, MV-F352427, MV-F401227, MV-F401727, MV-F451327, MV-F500927, MV-F501427, MV-F501927, MV-F552627
Intracranial aneurysm flow diverter device may be manufactured with an incorrect inner stent length and may not perform as intended.